Agreement of total corneal power between 2 swept-source optical coherence tomography and Scheimpflug tomography in normal and keratoconic patients

被引:14
作者
Asawaworarit, Rosepon [1 ]
Satitpitakul, Vannarut [1 ]
Taweekitikul, Parichart [1 ]
Pongpirui, Krit [2 ]
机构
[1] Chulalongkorn Univ, King Chulalongkorn Mem Hosp, Excellence Ctr Cornea & Limbal Stem Cell Transpla, Fac Med,Dept Ophthalmol, Bangkok, Thailand
[2] Chulalongkorn Univ, Fac Med, Bangkok, Thailand
来源
PLOS ONE | 2022年 / 17卷 / 05期
关键词
REPEATABILITY; ASTIGMATISM; EYES; RELIABILITY; KERATOMETRY;
D O I
10.1371/journal.pone.0268856
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose To evaluate agreement of total corneal power (TCP) measured by swept-source anterior segment optical coherence tomography (CASIA 2), Scheimpflug tomography (Pentacam AXL), and swept-source optical biometer (IOLMaster 700) in normal and keratoconic patients. Methods This is a prospective observational study conducted at King Chulalongkorn Memorial Hospital, Bangkok, Thailand. Biometric values were measured by each device for three times by two operators to evaluate repeatability and reproducibility of TCP. The agreement of TCP and other parameters including total corneal astigmatism, anterior keratometry, anterior corneal astigmatism, posterior keratometry, posterior corneal astigmatism, anterior chamber depth, white-to-white corneal diameter (WTW), central corneal thickness, and intraocular power were also evaluated. Results This study enrolled 100 healthy participants and 34 patients with keratoconus. The repeatability and reproducibility of TCP were high in all devices (ICC > 0.9). The agreement of TCP was excellent among three devices in both groups (ICC > 0.9). However, the agreement of TCP between CASIA 2 and IOLMaster 700 was slightly lower in healthy participants (ICC = 0.867) and keratoconic patients (ICC = 0.852) with mean differences of more than 1.0 diopter is clinically significant. Wider 95% limit of agreement were found in keratoconic eyes. Most of other parameters showed good to excellent agreement except WTW which showed poor to moderate agreement in both groups. IOL power showed clinically significant differences in patients with keratoconus. Conclusions The agreement of TCP measured by three devices was excellent in normal and keratoconic patients. However, TCP cannot be used interchangeably between devices.
引用
收藏
页数:14
相关论文
共 50 条
  • [21] Agreement between swept-source optical biometry and Scheimpflug-based tomography in eyes with previous myopic refractive surgery
    Karmiris, Efthymios
    Gartaganis, Panos S.
    Ntravalias, Thomas
    Manousakis, Evangelos
    Giannakis, Ioannis
    Chalkiadaki, Evangelia
    [J]. SAUDI JOURNAL OF OPHTHALMOLOGY, 2022, 36 (02) : 229 - 236
  • [22] Agreement between the Swept-Source Optical Coherence Tomography and the Image-Guided System for Biometry Assessment in Cataract Surgery
    Moreno-Paramo, Edel
    Garcia-Arroyo, Santiago
    Ortiz-Ramirez, Grecia Y.
    Garza-Leon, Manuel
    Gonzalez-Salinas, Roberto
    [J]. SEMINARS IN OPHTHALMOLOGY, 2022, 37 (03) : 324 - 329
  • [23] Comparison of Anterior Segment Measurements between Scheimpflug-Placido Camera and New Swept-source Optical Coherence Tomography
    Choe, Gon Soo
    Kim, Kook Young
    [J]. JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY, 2022, 63 (01): : 10 - 19
  • [24] Swept-Source Optical Coherence Tomography Features of Choroidal Nevi
    Francis, Jasmine H.
    Pang, Claudine E.
    Abramson, David H.
    Milman, Tatyana
    Folberg, Robert
    Mrejen, Sarah
    Freund, K. Bailey
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2015, 159 (01) : 169 - 176
  • [25] SWEPT-SOURCE OPTICAL COHERENCE TOMOGRAPHY ANGIOGRAPHY IMAGING OF THE CHORIOCAPILLARIS
    Ledesma-Gil, Gerardo
    Fernandez-Avellaneda, Pedro
    Spaide, Richard F.
    [J]. RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2021, 41 (07): : 1373 - 1378
  • [26] Assessment of corneal thickness measurement using swept-source Fourier-domain anterior segment optical coherence tomography and Scheimpflug camera
    Huang, Jinhai
    Feng, Yifan
    Wang, Qinmei
    Pesudovs, Konrad
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (07) : 1305 - 1306
  • [27] Choroidal Imaging with Swept-Source Optical Coherence Tomography in Patients with Birdshot Chorioretinopathy
    Dastiridou, Anna I.
    Bousquet, Elodie
    Kuehlewein, Laura
    Tepelus, Tudor
    Monnet, Dominique
    Salah, Sawsen
    Brezin, Antoine
    Sadda, Srinivas R.
    [J]. OPHTHALMOLOGY, 2017, 124 (08) : 1186 - 1195
  • [28] Comparison of corneal measurements in keratoconus eyes with two swept-source-optical coherence tomography devices and a Scheimpflug device
    Bograd, Alexandra
    Himmel, Isabelle
    Pfister, Isabel B.
    Seiler, Theo G.
    Frueh, Beatrice E.
    Tappeiner, Christoph
    [J]. GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2023, 262 (3) : 891 - 901
  • [29] Repeatability of a fully automated swept-source optical coherence tomography biometer and agreement with a low coherence reflectometry biometer
    Dominguez-Vicent, Alberto
    Venkataraman, Abinaya Priya
    Dalin, Amanda
    Brautaset, Rune
    Montes-Mico, Robert
    [J]. EYE AND VISION, 2023, 10 (01)
  • [30] Repeatability of a fully automated swept-source optical coherence tomography biometer and agreement with a low coherence reflectometry biometer
    Alberto Domínguez-Vicent
    Abinaya Priya Venkataraman
    Amanda Dalin
    Rune Brautaset
    Robert Montés-Micó
    [J]. Eye and Vision, 10