Robot-Assisted Total Hip Arthroplasty for Arthrodesed Hips

被引:15
作者
Chai, Wei [1 ]
Kong, Xiangpeng [1 ]
Yang, Minzhi [1 ,2 ]
Puah, Ken Lee [3 ]
Tang, Peifu [1 ]
Chen, Jiying [1 ]
机构
[1] Chinese Peoples Liberat Army Gen Hosp, Dept Orthoped, Beijing, Peoples R China
[2] Nankai Univ, Med Coll, Tianjin, Peoples R China
[3] Singapore Gen Hosp, Dept Orthoped Surg, Singapore, Singapore
关键词
total hip arthroplasty; robot-assisted surgery; cup positioning; hip arthrodesis; ACETABULAR CUP PLACEMENT; ANKYLOSING-SPONDYLITIS; PLAIN ANTEROPOSTERIOR; COMPONENT; RELIABILITY; REPLACEMENT; PRECISION; VALIDITY; PATIENT;
D O I
10.2147/TCRM.S246565
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
Background: Conversion of arthrodesed hips to total hip arthroplasty (THA) remains technically demanding. This study aims to evaluate the safety and efficacy of robotassisted THA in arthrodesed hips. Methods: We retrospectively analyzed 45 ankylosing spondylitis patients with hip arthrodesis in the Chinese PLA General Hospital between August 2018 and August 2019. All surgeries were carried out by one single surgeon. The patients were followed at 3 months after surgery. Gender, body mass index, angle of hip arthrodesis, operating time, intraoperative fluoroscopic times, postoperative length of hospitalization, cup positioning, postoperative leg length discrepancy, offset discrepancy, intraoperative and postoperative complications, and postoperative Harris Hip Score were collected for all patients. Results: Twenty-two patients (35 hips) who underwent robot-assisted THA and 23 patients (37 hips) who underwent manual THA were enrolled in this study. There were no significant differences in demographics and arthrodesed angles between the two groups. The fluoroscopic times during manual THA were significantly higher than those during robot-assisted THA (2.16 +/- 1.61 vs 0.47 +/- 0.61, respectively, p=0.000). In the robotic group, the percentage of acetabular cups within the safe zone was significantly greater than in the manual group (94.29% vs 67.56%, respectively, p=0.042). For manual THA, the anteversions were significantly different between the left and right sides (21.14 +/- 7.86 vs 16.00 +/- 6.32, respectively, p=0.042); however, no such significant difference was found in robot-assisted THA. Conclusion: Compared with manual THA for arthrodesed hips, robot-assisted THA had significant advantages in improving the frequency of achieving cup positioning within the target zone with diminished radiation dose and no increase in operating time.
引用
收藏
页码:357 / 368
页数:12
相关论文
共 50 条
[31]   Robot-Assisted Unicompartmental Knee Arthroplasty [J].
Pearle, Andrew D. ;
O'Loughlin, Padhraig F. ;
Kendoff, Daniel O. .
JOURNAL OF ARTHROPLASTY, 2010, 25 (02) :230-237
[32]   Cementless total hip arthroplasty with subtrochanteric transverse shortening osteotomy for severely dysplastic or dislocated hips [J].
Yalcin, Nadir ;
Kilicarslan, Kasim ;
Karatas, Fuat ;
Mutlu, Tansel ;
Yildirim, Hasan .
HIP INTERNATIONAL, 2010, 20 (01) :87-93
[33]   Robot-Assisted Total Hip Arthroplasty Demonstrates Improved 90-Day Clinical and Patient-Reported Outcomes [J].
Prinos, Alana ;
Buehring, Weston ;
Di Gangi, Catherine ;
Meere, Patrick ;
Meftah, Morteza ;
Hepinstall, Matthew .
ARTHROPLASTY TODAY, 2024, 27
[34]   Conversion of hip fusion to total hip arthroplasty: clinical, radiological outcomes and complications in 40 hips [J].
Mustafa Celiktas ;
Ozkan Kose ;
Adil Turan ;
Ferhat Guler ;
Cagri Ors ;
Emre Togrul .
Archives of Orthopaedic and Trauma Surgery, 2017, 137 :119-127
[35]   Conversion of hip fusion to total hip arthroplasty: clinical, radiological outcomes and complications in 40 hips [J].
Celiktas, Mustafa ;
Kose, Ozkan ;
Turan, Adil ;
Guler, Ferhat ;
Ors, Cagri ;
Togrul, Emre .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2017, 137 (01) :119-127
[36]   A conservative approach to dislocation following total hip arthroplasty: a review of 8606 hips [J].
Ogonda, Luke ;
Cassidy, Roslyn S. ;
Beverland, David E. .
HIP INTERNATIONAL, 2022, 32 (03) :291-297
[37]   Total Hip Arthroplasty with Modular Stem for Dysplastic Hips in South Asian Population [J].
Gautam, Deepak ;
Malhotra, Rajesh .
ARCHIVES OF BONE AND JOINT SURGERY-ABJS, 2019, 7 (06) :506-513
[38]   Robot-assisted primary cementless total hip arthroplasty using surface registration techniques: a short-term clinical report [J].
Nakamura, Nobuo ;
Sugano, Nobuhiko ;
Nishii, Takashi ;
Miki, Hidenobu ;
Kakimoto, Akihiro ;
Yamamura, Mitsuyoshi .
INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2009, 4 (02) :157-162
[39]   Robot-assisted primary cementless total hip arthroplasty using surface registration techniques: a short-term clinical report [J].
Nobuo Nakamura ;
Nobuhiko Sugano ;
Takashi Nishii ;
Hidenobu Miki ;
Akihiro Kakimoto ;
Mitsuyoshi Yamamura .
International Journal of Computer Assisted Radiology and Surgery, 2009, 4 :157-162
[40]   Precision and accuracy of robot-assisted technology with simplified express femoral workflow in measuring leg length and offset in total hip arthroplasty [J].
Cozzi Lepri, Andrea ;
Villano, Marco ;
Innocenti, Matteo ;
Porciatti, Tommaso ;
Matassi, Fabrizio ;
Civinini, Roberto .
INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2020, 16 (05) :1-6