Fast Energy Dependent Scatter Correction for List-Mode PET Data

被引:7
作者
Alvarez-Gomez, Juan Manuel [1 ]
Santos-Blasco, Joaquin [1 ]
Moliner Martinez, Laura [1 ]
Rodriguez-Alvarez, Maria Jose [1 ]
机构
[1] Univ Politecn Valencia, Inst Instrumentac Imagen Mol I3M, CSIC, Ctr Mixto, Valencia 46022, Spain
关键词
list-mode; positron emission tomography (PET); energy-based scatter correction; Monte Carlo simulation; 3D BRAIN PET; IMAGE-RECONSTRUCTION; COMPENSATION;
D O I
10.3390/jimaging7100199
中图分类号
TB8 [摄影技术];
学科分类号
0804 ;
摘要
Improvements in energy resolution of modern positron emission tomography (PET) detectors have created opportunities to implement energy-based scatter correction algorithms. Here, we use the energy information of auxiliary windows to estimate the scatter component. Our method is directly implemented in an iterative reconstruction algorithm, generating a scatter-corrected image without the need for sinograms. The purpose was to implement a fast energy-based scatter correction method on list-mode PET data, when it was not possible to use an attenuation map as a practical approach for the scatter degradation. The proposed method was evaluated using Monte Carlo simulations of various digital phantoms. It accurately estimated the scatter fraction distribution, and improved the image contrast in the simulated studied cases. We conclude that the proposed scatter correction method could effectively correct the scattered events, including multiple scatters and those originated in sources outside the field of view.
引用
收藏
页数:13
相关论文
共 44 条
  • [21] Efficient scatter modelling for incorporation in maximum likelihood reconstruction
    Hutton, BF
    Baccarne, V
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1998, 25 (12) : 1658 - 1665
  • [22] Validation of a simplified scatter correction method for 3D brain PET with 15O
    Ibaraki, Masanobu
    Matsubara, Keisuke
    Sato, Kaoru
    Mizuta, Tetsuro
    Kinoshita, Toshibumi
    [J]. ANNALS OF NUCLEAR MEDICINE, 2016, 30 (10) : 690 - 698
  • [23] GATE:: a simulation toolkit for PET and SPECT
    Jan, S
    Santin, G
    Strul, D
    Staelens, S
    Assié, K
    Autret, D
    Avner, S
    Barbier, R
    Bardiès, M
    Bloomfield, PM
    Brasse, D
    Breton, V
    Bruyndonckx, P
    Buvat, I
    Chatziioannou, AF
    Choi, Y
    Chung, YH
    Comtat, C
    Donnarieix, D
    Ferrer, L
    Glick, SJ
    Groiselle, CJ
    Guez, D
    Honore, PF
    Kerhoas-Cavata, S
    Kirov, AS
    Kohli, V
    Koole, M
    Krieguer, M
    van der Laan, DJ
    Lamare, F
    Largeron, G
    Lartizien, C
    Lazaro, D
    Maas, MC
    Maigne, L
    Mayet, F
    Melot, F
    Merheb, C
    Pennacchio, E
    Perez, J
    Pietrzyk, U
    Rannou, FR
    Rey, M
    Schaart, DR
    Schmidtlein, CR
    Simon, L
    Song, TY
    Vieira, JM
    Visvikis, D
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (19) : 4543 - 4561
  • [24] A MONTE-CARLO CORRECTION FOR THE EFFECT OF COMPTON-SCATTERING IN 3-D PET BRAIN IMAGING
    LEVIN, CS
    DAHLBOM, M
    HOFFMAN, EJ
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) : 1181 - 1185
  • [25] Evaluation of Scintillator Detection Materials for Application within Airborne Environmental Radiation Monitoring
    Lowdon, Matthew
    Martin, Peter G.
    Hubbard, M. W. J.
    Taggart, M. P.
    Connor, Dean T.
    Verbelen, Yannick
    Sellin, P. J.
    Scott, Thomas B.
    [J]. SENSORS, 2019, 19 (18)
  • [26] Implementation and analysis of list mode algorithm using tubes of response on a dedicated brain and breast PET
    Moliner, L.
    Correcher, C.
    Gonzalez, A. J.
    Conde, P.
    Hernandez, L.
    Orero, A.
    Rodriguez-Alvarez, M. J.
    Sanchez, F.
    Soriano, A.
    Vidal, L. F.
    Benlloch, J. M.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 702 : 129 - 132
  • [27] ESTIMATION OF A PROBABILITY DENSITY-FUNCTION AND MODE
    PARZEN, E
    [J]. ANNALS OF MATHEMATICAL STATISTICS, 1962, 33 (03): : 1065 - &
  • [28] Comparative evaluation of scatter correction in 3D PET using different scatter-level approximations
    Polycarpou, Irene
    Thielemans, Kris
    Manjeshwar, Ravindra
    Aguiar, Pablo
    Marsden, Paul K.
    Tsoumpas, Charalampos
    [J]. ANNALS OF NUCLEAR MEDICINE, 2011, 25 (09) : 643 - 649
  • [29] PET energy-based scatter estimation and image reconstruction with energy-dependent corrections
    Popescu, Lucretiu M.
    Lewitt, Robert M.
    Matej, Samuel
    Karp, Joel S.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (11) : 2919 - 2937
  • [30] Qian H, 2017, IEEE NUCL SCI CONF R