Oscillometric sphygmomanometers: a critical appraisal of current technology

被引:30
作者
Amoore, John Nicholas [1 ]
机构
[1] Crosshouse Hosp, Dept Med Phys, Kilmarnock KA2 OBE, Scotland
关键词
accuracy; blood pressure measurement; oscillometric; sphygmomanometer; validation; BLOOD-PRESSURE-MEASUREMENT; HYPERTENSION INTERNATIONAL PROTOCOL; EUROPEAN-SOCIETY; MEASURING DEVICES; VALIDATION; MONITORS; CUFF; ACCURACY; STANDARD; PREVALENCE;
D O I
10.1097/MBP.0b013e32835026b0
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Aim Blood pressure (BP), a key vital sign, monitors general health. Oscillometric devices are increasingly used for measurement, although their accuracy continues to be critically debated. A functional block diagram is used to review the components that affect accuracy. Methods A block diagram is presented covering the components from cuff to algorithm. The oscillometric waveform is described, considering factors that can alter its shape. Methods used to assess accuracy, including the potential use of simulators, are described. Results and discussion The block diagram focuses attention on cuff, amplifier, signal processing and algorithm. The importance of correct cuff size is emphasized. Accuracy can be affected by the extraction of the oscillometric pulses and the interpolation to compensate for higher deflation rates. Modern electronic amplifiers are assumed to be stable and do not drift, an assumption largely untested. Crucial to accuracy is the algorithm, but there is no standard algorithm and limited theoretical basis, leading to significant measurement errors in groups of patients, even by approved devices. The causes are not well understood, but differences in oscillometric waveform shape between patient groups have been observed and may explain auscultatory-oscillometric differences. The ability of theoretical models to explain the effects of arterial stiffness on BP measurements is discussed. Validation remains statistical though steps have been taken to improve it. Conclusion The indirect nature of BP measurement poses particularly problems for ensuring accuracy. Critical assessment has done much to improve standards, but a solid theoretical understanding of the technique has not been formulated and further work is required. Blood Press Monit 17:80-88 (C) 2012 Wolters Kluwer Health | Lippincott Williams & Wilkins.
引用
收藏
页码:80 / 88
页数:9
相关论文
共 65 条
[21]   CHARACTERIZATION OF THE OSCILLOMETRIC METHOD FOR MEASURING INDIRECT BLOOD-PRESSURE [J].
GEDDES, LA ;
VOELZ, M ;
COMBS, C ;
REINER, D ;
BABBS, CF .
ANNALS OF BIOMEDICAL ENGINEERING, 1982, 10 (06) :271-280
[22]   Limitations of current validation protocols for home blood pressure monitors for individual patients [J].
Gerin, W ;
Schwartz, AR ;
Schwartz, JE ;
Pickering, TG ;
Davidson, KW ;
Bress, J ;
O'Brien, E ;
Atkins, N .
BLOOD PRESSURE MONITORING, 2002, 7 (06) :313-318
[23]   Prevalence of blood pressure cuff sizes in a referral practice of 430 consecutive adult hypertensives [J].
Graves, JW .
BLOOD PRESSURE MONITORING, 2001, 6 (01) :17-20
[24]   An arm phantom: A digital simulation system for testing sphygmomanometers [J].
Gross, H ;
Mieke, S ;
Ulbrich, M .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1996, 20 (02) :75-83
[25]   CARDIAC CYCLE PHASE UNCERTAINTY - ANOTHER SOURCE OF ERROR IN INDIRECT BLOOD-PRESSURE MEASUREMENT [J].
HATSELL, CP .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1992, 16 (04) :157-158
[26]  
Hill L, 1897, Br Med J, V2, P904
[27]  
*ISO, 2009, 81060 ISO 2
[28]  
*ISO, 2007, 81060 ISO 1
[29]   Validation of the Welch Allyn 'Vital Signs' oscillometric blood pressure monitor [J].
Jones, CR ;
Taylor, K ;
Poston, L ;
Shennan, AH .
JOURNAL OF HUMAN HYPERTENSION, 2001, 15 (03) :191-195
[30]   Oscillometric blood pressure measurements by different devices are not interchangeable [J].
Kaufmann, MA ;
Pargger, H ;
Drop, LJ .
ANESTHESIA AND ANALGESIA, 1996, 82 (02) :377-381