Oscillometric sphygmomanometers: a critical appraisal of current technology

被引:30
作者
Amoore, John Nicholas [1 ]
机构
[1] Crosshouse Hosp, Dept Med Phys, Kilmarnock KA2 OBE, Scotland
关键词
accuracy; blood pressure measurement; oscillometric; sphygmomanometer; validation; BLOOD-PRESSURE-MEASUREMENT; HYPERTENSION INTERNATIONAL PROTOCOL; EUROPEAN-SOCIETY; MEASURING DEVICES; VALIDATION; MONITORS; CUFF; ACCURACY; STANDARD; PREVALENCE;
D O I
10.1097/MBP.0b013e32835026b0
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Aim Blood pressure (BP), a key vital sign, monitors general health. Oscillometric devices are increasingly used for measurement, although their accuracy continues to be critically debated. A functional block diagram is used to review the components that affect accuracy. Methods A block diagram is presented covering the components from cuff to algorithm. The oscillometric waveform is described, considering factors that can alter its shape. Methods used to assess accuracy, including the potential use of simulators, are described. Results and discussion The block diagram focuses attention on cuff, amplifier, signal processing and algorithm. The importance of correct cuff size is emphasized. Accuracy can be affected by the extraction of the oscillometric pulses and the interpolation to compensate for higher deflation rates. Modern electronic amplifiers are assumed to be stable and do not drift, an assumption largely untested. Crucial to accuracy is the algorithm, but there is no standard algorithm and limited theoretical basis, leading to significant measurement errors in groups of patients, even by approved devices. The causes are not well understood, but differences in oscillometric waveform shape between patient groups have been observed and may explain auscultatory-oscillometric differences. The ability of theoretical models to explain the effects of arterial stiffness on BP measurements is discussed. Validation remains statistical though steps have been taken to improve it. Conclusion The indirect nature of BP measurement poses particularly problems for ensuring accuracy. Critical assessment has done much to improve standards, but a solid theoretical understanding of the technique has not been formulated and further work is required. Blood Press Monit 17:80-88 (C) 2012 Wolters Kluwer Health | Lippincott Williams & Wilkins.
引用
收藏
页码:80 / 88
页数:9
相关论文
共 65 条
[1]   Obesity, hypertension and home sphygmomanometer cuffs [J].
Akpolat, Tekin .
EUROPEAN JOURNAL OF INTERNAL MEDICINE, 2010, 21 (04) :338-340
[2]  
Amoore J. N., 2006, Journal of Clinical Engineering, V31, P85
[3]   Auditing the technology used to measure blood pressure [J].
Amoore J.N. ;
Guehenec M. ;
Scordecchia R. ;
Scott D.H.T. .
Journal of Medical Engineering and Technology, 2010, 34 (03) :209-216
[4]  
Amoore JN, 2000, BLOOD PRESS MONIT, V5, P69
[5]   Evaluation of the Critikon 8100 and Spacelabs 90207 non-invasive blood pressure monitors using a test simulator [J].
Amoore, JN ;
Geake, WB .
JOURNAL OF HUMAN HYPERTENSION, 1997, 11 (03) :163-169
[6]  
Amoore JN, 2000, BLOOD PRESS MONIT, V5, P81
[7]   Can a simulator that regenerates physiological waveforms evaluate oscillometric non-invasive blood pressure devices? [J].
Amoore, JN ;
Vacher, E ;
Murray, IC ;
Mieke, S ;
King, ST ;
Smith, FE ;
Murray, A .
BLOOD PRESSURE MONITORING, 2006, 11 (02) :63-67
[8]   Do SpaceLabs ambulatory non-invasive blood pressure recorders measure blood pressure consistently over several years use? [J].
Amoore, JN ;
Dewar, D ;
Gough, K ;
Padfield, PL .
BLOOD PRESSURE MONITORING, 2005, 10 (01) :51-56
[9]   ASSESSMENT OF OSCILLOMETRIC NONINVASIVE BLOOD-PRESSURE MONITORS USING THE DYNATECH NEVADA CUFFLINK ANALYZER [J].
AMOORE, JN .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1993, 17 (01) :25-31
[10]   Automatic blood pressure measurement: the oscillometric waveform shape is a potential contributor to differences between oscillometric and auscultatory pressure measurements [J].
Amoore, John N. ;
Lemesre, Yann ;
Murray, Ian C. ;
Mieke, Stephan ;
King, Susan T. ;
Smith, Fiona E. ;
Murray, Alan .
JOURNAL OF HYPERTENSION, 2008, 26 (01) :35-43