Finite Element Modeling of Magnitude and Location of Brain Micromotion Induced Strain for Intracortical Implants

被引:9
作者
Al Abed, Ali [1 ]
Amatoury, Jason [2 ]
Khraiche, Massoud [3 ]
机构
[1] Amer Univ Beirut, Dept Mech Engn, Beirut, Lebanon
[2] Amer Univ Beirut, Biomed Engn Program, Sleep & Upper Airway Res Grp, Beirut, Lebanon
[3] Amer Univ Beirut, Biomed Engn Program, Neural Engn & Nanobiosensors Grp, Beirut, Lebanon
关键词
intracortical; micromotion; FEM; implants; brain; neuron; strain; glia; NEURAL PROBES; MICROELECTRODES;
D O I
10.3389/fnins.2021.727715
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Micromotion-induced stress remains one of the main determinants of life of intracortical implants. This is due to high stress leading to tissue injury, which in turn leads to an immune response coupled with a significant reduction in the nearby neural population and subsequent isolation of the implant. In this work, we develop a finite element model of the intracortical probe-tissue interface to study the effect of implant micromotion, implant thickness, and material properties on the strain levels induced in brain tissue. Our results showed that for stiff implants, the strain magnitude is dependent on the magnitude of the motion, where a micromotion increase from 1 to 10 mu m induced an increase in the strain by an order of magnitude. For higher displacement over 10 mu m, the change in the strain was relatively smaller. We also showed that displacement magnitude has no impact on the location of maximum strain and addressed the conflicting results in the literature. Further, we explored the effect of different probe materials [i.e., silicon, polyimide (PI), and polyvinyl acetate nanocomposite (PVAc-NC)] on the magnitude, location, and distribution of strain. Finally, we showed that strain distribution across cortical implants was in line with published results on the size of the typical glial response to the neural probe, further reaffirming that strain can be a precursor to the glial response.
引用
收藏
页数:12
相关论文
共 37 条
[1]   Highly Flexible Single-Unit Resolution All Printed Neural Interface on a Bioresorbable Backbone [J].
Almasri, Reem M. ;
AlChamaa, Walid ;
Tehrani-Bagha, Ali Reza ;
Khraiche, Massoud L. .
ACS APPLIED BIO MATERIALS, 2020, 3 (10) :7040-7051
[2]  
Bilston LE, 2011, STUD MECHANOBIOL TIS, V3, P1, DOI 10.1007/978-3-642-13890-4
[3]   Chronically Implanted Intracranial Electrodes: Tissue Reaction and Electrical Changes [J].
Campbell, Andrew ;
Wu, Chengyuan .
MICROMACHINES, 2018, 9 (09)
[4]   Stimuli-responsive polymer nanocomposites inspired by the sea cucumber dermis [J].
Capadona, Jeffrey R. ;
Shanmuganathan, Kadhiravan ;
Tyler, Dustin J. ;
Rowan, Stuart J. ;
Weder, Christoph .
SCIENCE, 2008, 319 (5868) :1370-1374
[5]  
Chou ZN, 2015, IEEE ENG MED BIO, P3949, DOI 10.1109/EMBC.2015.7319258
[6]   Low-intensity ultrasound neuromodulation: An overview of mechanisms and emerging human applications [J].
Fomenko, Anton ;
Neudorfer, Clemens ;
Dallapiazza, Robert F. ;
Kalia, Suneil K. ;
Lozano, Andres M. .
BRAIN STIMULATION, 2018, 11 (06) :1209-1217
[7]   Brain micromotion around implants in the rodent somatosensory cortex [J].
Gilletti, Aaron ;
Muthuswamy, Jit .
JOURNAL OF NEURAL ENGINEERING, 2006, 3 (03) :189-195
[8]  
Helton Kristen L, 2011, J Diabetes Sci Technol, V5, P647
[9]  
Hong GS, 2019, NAT REV NEUROSCI, V20, P330, DOI 10.1038/s41583-019-0140-6
[10]  
Khraiche M.L., 2013, ULTRAHIGH PHOTOSENSI