Accuracy of a new photorefractometer in young and adult patients

被引:46
作者
Schimitzek, T [1 ]
Lagrèze, WA [1 ]
机构
[1] Schielbrhandlung Univ Augenklin, Sekt Neuroophthalmol Kinderophthalmol, D-79106 Freiburg, Germany
关键词
D O I
10.1007/s00417-004-1056-y
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background: Photorefraction can be carried out in both eyes simultaneously from a distance and is therefore suitable for examination of children. This study evaluated the accuracy of a commercially available photorefractometer (PowerRefractor, Plusoptix, Erlangen, Germany) and investigated whether the working distance relaxes the accommodation sufficiently without cycloplegia. Methods: Photorefractometer readings were compared to cycloplegic retinoscopy. Because of the limited working range the group of patients consisted of low and moderate ametropic eyes within a spherical power of -7.0 to +5.0 D. One hundred and ninety-two eyes from 104 patients (2-81 years) were photorefracted under cycloplegia. A subgroup of 83 eyes from 46 patients was additionally refracted without cycloplegia. Results: Under cycloplegia, the PowerRefractor measured the spherical equivalent slightly below that of cycloplegic retinoscopy (too much minus). The mean difference in spherical equivalent was -0.12 +/- 0.91 D (SD). The mean difference of cylindrical power was -0.17 +/- 0.73 D. The mean weighted axis difference was 0.61 +/- 0.71 D which is comparable to an axis deviation of 18 degrees at a cylinder power of 1.00 D. Without cycloplegia, the mean difference of the spherical equivalent was -0.73 +/- 1.25 D. The mean difference of cylindrical power was -0.20 +/- 0.65 D. The mean weighted axis difference was 0.44 +/- 0.58 D which is comparable to an axis deviation of 13 degrees at a cylinder power of 1.00 D. Conclusions: Without cycloplegia, the spherical equivalent of the PowerRefractor tends to be underestimated due to uncontrolled accommodation, especially in children. Cycloplegia improves the accuracy in evaluating the spherical equivalent, but decreases the accuracy of cylinder power and axis.
引用
收藏
页码:637 / 645
页数:9
相关论文
共 27 条
[21]   Measured inaccuracy of the videorefractometry at higher ametropia [J].
Schimitzek, T ;
Krzizok, T .
KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2001, 218 (06) :438-444
[22]   Accuracy and precision of the Tomey ViVA infrared photorefractor [J].
Thompson, AM ;
Li, T ;
Peck, LB ;
Howland, HC ;
Counts, R ;
Bobier, WR .
OPTOMETRY AND VISION SCIENCE, 1996, 73 (10) :644-652
[23]   Retinoscopy in infants using a near noncycloplegic technique, cycloplegia with tropicamide 1%, and cycloplegia with cyclopentolate 1% [J].
Twelker, JD ;
Mutti, DO .
OPTOMETRY AND VISION SCIENCE, 2001, 78 (04) :215-222
[24]   THEORY OF ECCENTRIC PHOTOREFRACTION (PHOTORETINOSCOPY) - ASTIGMATIC EYES [J].
WESEMANN, W ;
NORCIA, AM ;
ALLEN, D .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1991, 8 (12) :2038-2047
[25]  
WESEMANN W, 1987, AM J OPTOM PHYS OPT, V64, P627
[26]   Accuracy and accommodation capability of a handheld autorefractor [J].
Wesemann, W ;
Dick, B .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (01) :62-70
[27]  
WESEMANN W, 1994, STRABISMUS, V2, P147