Characterization of Interfacial Failure in Cemented Total Joint Replacements via Self-Sensing Bone Cement, Electrical Impedance Tomography, and Machine Learning

被引:3
作者
Ghaednia, H. [1 ]
Owens, C. [2 ]
Keiderling, L. E. [1 ]
Hart, J. [2 ]
Varadarajan, K. [1 ]
Schwab, J. [1 ]
Tallman, T. N. [3 ]
机构
[1] Massachusetts Gen Hosp, Dept Orthopaed Surg, Boston, MA 02114 USA
[2] MIT, Dept Mech Engn, Cambridge, MA 02139 USA
[3] Purdue Univ, Sch Aeronaut & Astronaut, W Lafayette, IN 47907 USA
来源
HEALTH MONITORING OF STRUCTURAL AND BIOLOGICAL SYSTEMS XV | 2021年 / 11593卷
关键词
self-sensing materials; medical devices; electrical impedance tomography; machine learning; TOTAL KNEE ARTHROPLASTY; DESIGN; SYSTEM; STRAIN; IMPACT; HIP;
D O I
10.1117/12.2581950
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
At an estimated cost of $8 billion annually in the United States, revision total joint replacement surgeries represent a substantial financial burden to the health care system. Fixation failures such as implant loosening and mechanical instability at the cement-to-bone interface are the main causes of long-term implant failure. Early and accurate diagnosis of cement-bone interface failure is critical to develop therapeutic strategies and reduce the risk of delayed revision surgery. Unfortunately, prevailing imaging modalities such as plain radiographs struggle to detect the precursors of implant failure and are often interpreted incorrectly. Our team has recently shown that modification of poly(methyl methacrylate) (PMMA) bone cement with low concentrations of conductive fillers makes it piezoresistive and therefore self-sensing. By electrical impedance tomography (EIT) it is possible to monitor load transfer across the PMMA using real-time electrical measurements, which are physiologically benign and would ultimately be cost-effective in a clinical setting. Herein, we demonstrate the use of machine learning with EIT in order to precisely characterize model failure events. In phantom tests, we introduce three defect types: loosening, vertical cracks, and horizontal cracks. We show that implant defect type can be classified with greater than 95% accuracy by combining principal component analysis (PCA) coupled with the k-nearest neighbor (KNN) algorithm and a two layer feed-forward neural network with three hidden neurons. These preliminary results show that the combination of smart materials, EIT, and machine learning may be a powerful tool for predicting and diagnosing failure in joint replacements.
引用
收藏
页数:8
相关论文
共 28 条
[1]   PROJECT OF HIP JOINT ENDOPROSTHESIS FOR AN INDIVIDUAL PATIENT WITH MATERIALS SELECTION [J].
Bartmanski, M. ;
Swieczko-Zurek, B. .
ADVANCES IN MATERIALS SCIENCE, 2015, 15 (01) :30-36
[2]   Clinical and Economic Burden of Revision Knee Arthroplasty [J].
Bhandari, Mohit ;
Smith, Jon ;
Miller, Larry E. ;
Block, Jon E. .
CLINICAL MEDICINE INSIGHTS-ARTHRITIS AND MUSCULOSKELETAL DISORDERS, 2012, 5 :89-94
[3]   Unusually High Rate of Early Failure of Tibial Component in ATTUNE Total Knee Arthroplasty System at Implant-Cement Interface [J].
Bonutti, Peter M. ;
Khlopas, Anton ;
Chughtai, Morad ;
Cole, Connor ;
Gwam, Chukwuweike U. ;
Harwin, Steven F. ;
Whited, Brent ;
Omiyi, Didi E. ;
Drumm, Joshua E. .
JOURNAL OF KNEE SURGERY, 2017, 30 (05) :435-439
[4]   Three-dimensional EIT imaging of breast tissues: System design and clinical testing [J].
Cherepenin, VA ;
Karpov, AY ;
Korjenevsky, AV ;
Kornienko, VN ;
Kultiasov, YS ;
Ochapkin, MB ;
Trochanova, OV ;
Meister, JD .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (06) :662-667
[5]   Solving electrical impedance tomography with deep learning [J].
Fan, Yuwei ;
Ying, Lexing .
JOURNAL OF COMPUTATIONAL PHYSICS, 2020, 404
[6]   Damage detection in 2.5D C/SiC composites using electrical resistance tomography [J].
Gao, Xiguang ;
Wei, Tingting ;
Dong, Hongnian ;
Song, Yingdong .
JOURNAL OF THE EUROPEAN CERAMIC SOCIETY, 2019, 39 (13) :3583-3593
[7]   Interfacial load monitoring and failure detection in total joint replacements via piezoresistive bone cement and electrical impedance tomography [J].
Ghaednia, H. ;
Owens, C. E. ;
Roberts, R. ;
Tallman, T. N. ;
Hart, A. J. ;
Varadarajan, K. M. .
SMART MATERIALS AND STRUCTURES, 2020, 29 (08)
[8]  
Ghaednia H., 2020, ORTHOPAEDIC P 102 S1, P130, DOI [10.1302/1358, DOI 10.1302/1358]
[9]   A Review of Elastic-Plastic Contact Mechanics [J].
Ghaednia, Hamid ;
Wang, Xianzhang ;
Saha, Swarna ;
Xu, Yang ;
Sharma, Aman ;
Jackson, Robert L. .
APPLIED MECHANICS REVIEWS, 2017, 69 (06)
[10]   A comprehensive study of the elasto-plastic contact of a sphere and a flat [J].
Ghaednia, Hamid ;
Pope, Sara A. ;
Jackson, Robert L. ;
Marghitu, Dan B. .
TRIBOLOGY INTERNATIONAL, 2016, 93 :78-90