Implementation of Time-Resolved Step-Scan Fourier Transform Infrared (FT-IR) Spectroscopy Using a kHz Repetition Rate Pump Laser

被引:6
|
作者
Magana, Donny [1 ]
Parul, Dzmitry [1 ]
Dyer, R. Brian [1 ]
Shreve, Andrew P. [2 ]
机构
[1] Emory Univ, Dept Chem, Atlanta, GA 30322 USA
[2] Los Alamos Natl Lab, Mat Phys & Applicat Div, Ctr Integrated Nanotechnol MPA CINT, Los Alamos, NM 87545 USA
关键词
Spectroscopic instrumentation; Infrared spectroscopy; Time-resolved spectroscopy; Time-resolved step-scan Fourier transform infrared spectroscopy; FT-IR spectroscopy; Nanosecond phenomena; EXCITED-STATES; VIBRATIONAL SPECTROSCOPY; CARBON-MONOXIDE; TRANSITION; DYNAMICS; CO;
D O I
10.1366/10-06179
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Time-resolved step-scan Fourier transform infrared (FT-IR) spectroscopy has been shown to be invaluable for studying excited-state structures and dynamics in both biological and inorganic systems. Despite the established utility of this method, technical challenges continue to limit the data quality and more wide ranging applications. A critical problem has been the low laser repetition rate and interferometer stepping rate (both are typically 10 Hz) used for data acquisition. Here we demonstrate significant improvement in the quality of time-resolved spectra through the use of a kHz repetition rate laser to achieve kHz excitation and data collection rates while stepping the spectrometer at 200 Hz. We have studied the metal-to-ligand charge transfer excited state of Ru(bipyridine)(3)Cl-2 in deuterated acetonitrile to test and optimize high repetition rate data collection. Comparison of different interferometer stepping rates reveals an optimum rate of 200 Hz due to minimization of long-term baseline drift. With the improved collection efficiency and signal-to-noise ratio, better assignments of the MLCT excited-state bands can be made. Using optimized parameters, carbonmonoxy myoglobin in deuterated buffer is also studied by observing the infrared signatures of carbon monoxide photolysis upon excitation of the heme. We conclude from these studies that a substantial increase in performance of ss-FT-IR instrumentation is achieved by coupling commercial infrared benches with kHz repetition rate lasers.
引用
收藏
页码:535 / 542
页数:8
相关论文
共 50 条
  • [21] Time-resolved step-scan FT-IR spectroscopy of a nematic solution of 2-naphthaldehyde in an electric field
    Jordanov, B
    Okretic, S
    Siesler, HW
    APPLIED SPECTROSCOPY, 1997, 51 (03) : 447 - 449
  • [22] Step-scan FT-IR time-resolved spectroscopy (S-2(FT-IR) TRS) of the electro-reorientation of liquid crystalline materials
    Fuji, A
    Palmer, RA
    Chen, P
    Jiang, EY
    Chao, JL
    MIKROCHIMICA ACTA, 1997, : 599 - 602
  • [23] TIME-RESOLVED STEP-SCAN FOURIER-TRANSFORM INFRARED-SPECTROSCOPY APPLIED TO BIOLOGICAL-SYSTEMS
    WEIDLICH, O
    RODIG, C
    SIEBERT, F
    BIOPHYSICAL JOURNAL, 1994, 66 (02) : A251 - A251
  • [24] Application of time-resolved step-scan Fourier transform infrared spectroscopy to photochemical mechanistic investigations of alkyl phenylglyoxylates
    Fedorov, AV
    Danilov, EO
    Merzlikine, AG
    Rodgers, MAJ
    Neckers, AC
    JOURNAL OF PHYSICAL CHEMISTRY A, 2003, 107 (18): : 3208 - 3214
  • [26] APPLICATIONS OF TIME-RESOLVED STEP-SCAN AND RAPID-SCAN FT-IR SPECTROSCOPY - DYNAMICS FROM 10 SECONDS TO 10 NANOSECONDS
    JOHNSON, TJ
    SIMON, A
    WEIL, JM
    HARRIS, GW
    APPLIED SPECTROSCOPY, 1993, 47 (09) : 1376 - 1381
  • [27] Nanosecond time-resolved step-scan FT-IR spectroscopy of the photochemical ring-opening of 1,3-cyclohexadiene
    Kauffmann, E
    Frei, H
    Mathies, RA
    CHEMICAL PHYSICS LETTERS, 1997, 266 (5-6) : 554 - 559
  • [28] Investigation of Electronic Structures of Triplet States Using Step-Scan Time-Resolved Fourier-Transform Near-Infrared Spectroscopy
    Wu, Chia Chun
    Tsai, Yu-Xiang
    Chu, Li-Kang
    Chen, I-Chia
    JOURNAL OF PHYSICAL CHEMISTRY LETTERS, 2024, 15 (04): : 912 - 918
  • [29] Application of doubled-angle phase correction method to time-resolved step-scan FT-IR spectra
    Hutson, MS
    Braiman, MS
    VIBRATIONAL SPECTROSCOPY, 1999, 19 (02) : 379 - 383
  • [30] Characterization of alkoxycarbonyl radicals by step-scan time-resolved infrared spectroscopy
    Bucher, G
    Halupka, M
    Kolano, C
    Schade, O
    Sander, W
    EUROPEAN JOURNAL OF ORGANIC CHEMISTRY, 2001, 2001 (03) : 545 - 552