Diastolic function measurements and diagnostic consequences: a comparison of pulsed wave- and color-coded tissue Doppler imaging

被引:22
作者
Hummel, Y. M. [1 ]
Klip, I. J. T. [1 ]
de Jong, R. M. [1 ]
Pieper, P. G. [1 ]
van Veldhuisen, D. J. [1 ]
Voors, A. A. [1 ]
机构
[1] Univ Groningen, Dept Cardiol, Univ Med Ctr Groningen, NL-9700 RB Groningen, Netherlands
关键词
Tissue velocity; Diastolic function; Echocardiography; Tissue Doppler imaging; HEART-FAILURE; ECHOCARDIOGRAPHY; DYSFUNCTION;
D O I
10.1007/s00392-010-0141-y
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Tissue Doppler imaging (TDI) plays an important role in assessing diastolic function using echocardiography. However, two different methods [pulsed wave (PW-TDI) and color-coded (CC-TDI)] are currently used. We aimed to compare both measurements. We included 114 patients that were referred to our echocardiography department for evaluation of diastolic left ventricular function. In these patients, we sequentially measured tissue velocities of basal lateral and septal myocardium of the left ventricle in an apical four-chamber view with both PW-TDI and CC-TDI. Our cohort consisted of a heterogeneous group of patients with and without a history of cardiac disease. Mean age of the patients was 52 +/- A 16.7 years, and 62% were males. We found a strong correlation between PW-TDI- and CC-TDI-derived myocardial velocities (r = 0.93; p = 0.001). However, E' (mean of lateral and septal) velocities measured with PW-TDI were consistently higher compared to CC-TDI values [PW-TDI E' 10.3 +/- A 3.9 (SD) cm/s vs. CC-TDI E' 7.7 +/- A 3.1 cm/s; p < 0.001]. From these data, we calculated that the relation between E' measured with PW-TDI and CC-TDI can be described as: E' (PW-TDI) = 1.25 + 1.17 x E' (CC-TDI). Consequently, E/E' measured with PW-TDI was consistently lower compared with CC-TDI (9.1 +/- A 3.1 vs. 12.5 +/- A 5.7; p < 0.001) From these data, we calculated that the relation between E/E' measured with PW-TDI and CC-TDI can be described as: E/E' (PW-TDI) = 2.13 + 0.56 x E/E' (CC-TDI). Despite a strong correlation, tissue velocities measured with PW-TDI will yield higher values as compared with CC-TDI. This should be taken into account when defining cut-off values for the evaluation of diastolic function.
引用
收藏
页码:453 / 458
页数:6
相关论文
共 17 条
[1]   Characteristics of left ventricular diastolic dysfunction in the community: an echocardiographic survey [J].
Abhayaratna, W. P. ;
Marwick, T. H. ;
Smith, W. T. ;
Becker, N. G. .
HEART, 2006, 92 (09) :1259-1264
[2]   Isolated diastolic dysfunction - diagnostic value of tissue Doppler imaging, colour M-mode and N-terminal pro B-type natriuretic peptide [J].
Hettwer, S. ;
Panzner-Grote, B. ;
Witthaut, R. ;
Werdan, K. .
CLINICAL RESEARCH IN CARDIOLOGY, 2007, 96 (12) :874-882
[3]   Perception of symptoms is out of proportion to cardiac pathology in patients with "diastolic heart failure'' [J].
Ingle, L. ;
Cleland, J. G. F. ;
Clark, A. L. .
HEART, 2008, 94 (06) :748-753
[4]   A comparison of regional myocardial velocity information derived by pulsed and color Doppler techniques: An in vitro and in vivo study [J].
Kukulski, T ;
Voigt, JU ;
Wilkenshoff, UM ;
Strotmann, JM ;
Wranne, B ;
Hatle, L ;
Sutherland, GR .
ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2000, 17 (07) :639-651
[5]   Recommendations for chamber quantification [J].
Lang, Roberto M. ;
Bierig, Michelle ;
Devereux, Richard B. ;
Flachskampf, Frank A. ;
Foster, Elyse ;
Pellikka, Patricia A. ;
Picard, Michael H. ;
Roman, Mary J. ;
Seward, James ;
Shanewise, Jack ;
Solomon, Scott ;
Spencer, Kirk T. ;
Sutton, Martin St. John ;
Stewart, William .
EUROPEAN JOURNAL OF ECHOCARDIOGRAPHY, 2006, 7 (02) :79-108
[6]   Irbesartan in Patients with Heart Failure and Preserved Ejection Fraction. [J].
Massie, Barry M. ;
Carson, Peter E. ;
McMurray, John J. ;
Komajda, Michel ;
McKelvie, Robert ;
Zile, Michael R. ;
Anderson, Susan ;
Donovan, Mark ;
Iverson, Erik ;
Staiger, Christoph ;
Ptaszynska, Agata .
NEW ENGLAND JOURNAL OF MEDICINE, 2008, 359 (23) :2456-2467
[7]   Color tissue Doppler myocardial velocities consistently underestimate spectral tissue Doppler velocities: Impact on calculation peak transmitral pulsed Doppler velocity/early diastolic tissue Doppler velocity (E/Ea) [J].
McCulloch, Marti ;
Zoghbi, William A. ;
Davis, Robert ;
Thomas, Chris ;
Dokainish, Hisharn .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2006, 19 (06) :744-748
[8]   Doppler estimation of left ventricular filling pressure in sinus tachycardia -: A new application of tissue Doppler imaging [J].
Nagueh, SF ;
Mikati, I ;
Kopelen, HA ;
Middleton, KJ ;
Quiñones, MA ;
Zoghbi, WA .
CIRCULATION, 1998, 98 (16) :1644-1650
[9]   Doppler tissue imaging: A noninvasive technique for evaluation of left ventricular relaxation and estimation of filling pressures [J].
Nagueh, SF ;
Middleton, KJ ;
Kopelen, HA ;
Zoghbi, WA ;
Quinones, MA .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1997, 30 (06) :1527-1533