Accuracy of common IOL power formulas in 611 eyes based on axial length and corneal power ranges

被引:23
作者
Roeggla, Veronika [1 ]
Langenbucher, Achim [2 ]
Leydolt, Christina [1 ]
Schartmueller, Daniel [1 ]
Schwarzenbacher, Luca [1 ]
Abela-Formanek, Claudette [1 ]
Menapace, Rupert [1 ]
机构
[1] Med Univ Vienna, Dept Ophthalmol & Optometry, Vienna, Austria
[2] Saarland Univ, Inst Expt Ophthalmol, Saarbrucken, Germany
关键词
Clinical Trial; Optics and Refraction; Treatment Surgery; INTRAOCULAR-LENS POWER; ANTERIOR-CHAMBER DEPTH; HOFFER-Q; SRK/T; BIOMETRY;
D O I
10.1136/bjophthalmol-2020-315882
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Aims To provide clinical guidance on the use of intraocular lens (IOL) power calculation formulas according to the biometric parameters. Methods 611 eyes that underwent cataract surgery were retrospectively analysed in subgroups according to the axial length (AL) and corneal power (K). The predicted residual refractive error was calculated and compared to evaluate the accuracy of the following formulas: Haigis, Hoffer Q, Holladay 1 and SRK/T. Furthermore, the percentages of eyes with <=+/- 0.25, <=+/- 0.5 and 1 dioptres (D) of the prediction error were recorded. Results The Haigis formula showed the highest percentage of cases with <= 0.5 D in eyes with a short AL and steep K (90%), average AL and steep cornea (73.2%) but also in long eyes with a flat and average K (65% and 72.7%, respectively). The Hoffer Q formula delivered the lowest median absolute error (MedAE) in short eyes with an average K (0.30 D) and Holladay 1 in short eyes with a steep K (Holladay 1 0.24 D). SRK/T presented the highest percentage of cases with <= 0.5 D in average long eyes with a flat and average K (80.5% and 68.1%, respectively) and the lowest MedAE in long eyes with an average K (0.29 D). Conclusion Overall, the Haigis formula shows accurate results in most subgroups. However, attention must be paid to the axial eye length as well as the corneal power when choosing the appropriate formula to calculate an IOL power, especially in eyes with an unusual biometry.
引用
收藏
页码:1661 / 1665
页数:5
相关论文
共 33 条
[1]   Intraocular lens power calculation for eyes with an axial length greater than 26.0 mm: Comparison of formulas and methods [J].
Abulafia, Adi ;
Barrett, Graham D. ;
Rotenberg, Michael ;
Kleinmann, Guy ;
Levy, Adi ;
Reitblat, Olga ;
Koch, Douglas D. ;
Wang, Li ;
Assia, Ehud I. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (03) :548-556
[2]   Formula choice: Hoffer Q, Holladay 1, or SRK/T and refractive outcomes in 8108 eyes after cataract surgery with biometry by partial coherence interferometry [J].
Aristodemou, Petros ;
Cartwright, Nathaniel E. Knox ;
Sparrow, John M. ;
Johnston, Robert L. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (01) :63-71
[4]   Accuracy of Intraocular Lens Power Formulas Involving 148 Eyes with Long Axial Lengths: A Retrospective Chart-Review Study [J].
Chen, Chong ;
Xu, Xian ;
Miao, Yuyu ;
Zheng, Gaoxin ;
Sun, Yong ;
Xu, Xun .
JOURNAL OF OPHTHALMOLOGY, 2015, 2015
[5]   Assessment of the accuracy of new and updated intraocular lens power calculation formulas in 10 930 eyes from the UK National Health Service [J].
Darcy, Kieren ;
Gunn, David ;
Tavassoli, Shokufeh ;
Sparrow, John ;
Kane, Jack X. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2020, 46 (01) :2-7
[6]   Accuracy of intraocular lens power calculations in eyes with axial length &lt;22.00 mm [J].
Day, Alexander C. ;
Foster, Paul J. ;
Stevens, Julian D. .
CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2012, 40 (09) :855-862
[7]   Comparison of Hoffer Q and Haigis Formulae for Intraocular Lens Power Calculation According to the Anterior Chamber Depth in Short Eyes [J].
Eom, Youngsub ;
Kang, Su-Yeon ;
Song, Jong Suk ;
Kim, Yong Yeon ;
Kim, Hyo Myung .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2014, 157 (04) :818-824
[8]   Intraocular lens power calculation in short eyes [J].
Gavin, E. A. ;
Hammond, C. J. .
EYE, 2008, 22 (07) :935-938
[9]   Accuracy of 8 intraocular lens calculation formulas in relation to anterior chamber depth in patients with normal axial lengths [J].
Gokce, Sabite Emine ;
De Oca, Ildamaris Montes ;
Cooke, David L. ;
Wang, Li ;
Koch, Douglas D. ;
Al-Mohtaseb, Zaina .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2018, 44 (03) :362-368
[10]   Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis [J].
Haigis, W ;
Lege, B ;
Miller, N ;
Schneider, B .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2000, 238 (09) :765-773