Novel Cryoballoon Ablation System for Single Shot Pulmonary Vein Isolation - The Prospective ICE-AGE-X Study -

被引:33
作者
Tilz, Roland Richard [1 ,2 ]
Meyer-Saraei, Roza [1 ]
Eitel, Charlotte [1 ]
Fink, Thomas [1 ]
Sciacca, Vanessa [1 ]
Lopez, Lisbeth Delgado [1 ]
Kirstein, Bettina [1 ]
Schluter, Michael [3 ]
Vogler, Julia [1 ]
Kuck, Karl-Heinz [1 ,3 ]
Heeger, Christian-H [1 ,2 ]
机构
[1] Univ Hosp Schleswig Holstein, Univ Heart Ctr Lubeck, Med Clin 2, Dept Cardiol Angiol & Intens Care Med,Div Electro, Lubeck, Germany
[2] Univ Hosp Schleswig Holstein, Lubeck, Germany
[3] German Ctr Cardiovasc Res DZHK, Partner Site Hamburg Kiel Lubeck, Lubeck, Germany
关键词
Acute efficacy; Atrial fibrillation; Cryoballoon; Pulmonary vein isolation; 2ND-GENERATION CRYOBALLOON; ATRIAL-FIBRILLATION; CLINICAL SUCCESS; FREEZE; LESSONS; LESIONS;
D O I
10.1253/circj.CJ-21-0094
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The arctic front cryoballoon (AF-CB) provides effective and durable pulmonary vein isolation (PVI) associated with encouraging clinical outcome. The POLARx cryoballoon incorporates unique features and design changes that may translate into improved efficacy, safety and further simplified balloon-based procedures. Efficacy and safety of the novel POLARx cryoballoon was compared to the fourth generation AF-CB (AF-CB4). Methods and Results: Twenty-five consecutive patients with paroxysmal or persistent atrial fibrillation were prospectively enrolled, underwent POLARx-based PVI (POLARx group) and were compared to 25 consecutive patients treated with the AF-CB4 (AF-CB4 group). All PVs were successfully isolated utilizing the POLARx and AF-CB4. A significant difference regarding the mean minimal cryoballoon temperatures reached using the AF-CB4 and POLARx (-50 +/- 6 degrees C vs. -57 +/- 7 degrees C, P=0.004) was observed. Real-time PVI was visualized in 81% of POLARx patients and 42% of AF-CB4 patients (P<0.001). Utilizing the POLARx, a trend towards shorter median procedure time (POLARx: 45 [39, 53] min vs. AF-CB4: 55 [50, 60] min; P=0.062) was found. No differences were observed between AF-CB4 and POLARx concerning catheter maneuverability, catheter stability and periprocedural complications. Conclusions: The novel POLARx showed similar safety and efficacy compared to the AF-CB4. A higher rate of real-time PV recordings and significantly lower minimal balloon temperatures were observed using the POLARx.
引用
收藏
页码:1296 / 1304
页数:9
相关论文
共 18 条
[1]   High rate of durable pulmonary vein isolation after second-generation cryoballoon ablation: analysis of repeat procedures [J].
Bordignon, Stefano ;
Fuernkranz, Alexander ;
Perrotta, Laura ;
Dugo, Daniela ;
Konstantinou, Athanasios ;
Nowak, Bernd ;
Schulte-Hahn, Britta ;
Schmidt, Boris ;
Chun, Kyoung Ryul Julian .
EUROPACE, 2015, 17 (05) :725-731
[2]   Phrenic nerve paralysis during cryoballoon ablation for atrial fibrillation: A comparison between the first- and second-generation balloon [J].
Casado-Arroyo, Ruben ;
Chierchia, Gian-Battista ;
Conte, Giulio ;
Levinstein, Moises ;
Sieira, Juan ;
Rodriguez-Manero, Moises ;
di Giovanni, Giacomo ;
Baltogiannis, Yannis ;
Wauters, Kristel ;
de Asmundis, Carlo ;
Sarkozy, Andrea ;
Brugada, Pedro .
HEART RHYTHM, 2013, 10 (09) :1318-1324
[3]   Single 3-minute freeze for second-generation cryoballoon ablation: One-year follow-up after pulmonary vein isolation [J].
Ciconte, Giuseppe ;
de Asmundis, Carlo ;
Sieira, Juan ;
Conte, Giulio ;
Di Giovanni, Giacomo ;
Mugnai, Giacomo ;
Saitoh, Yukio ;
Baltogiannis, Giannis ;
Irfan, Ghazala ;
Coutino-Moreno, Hugo Enrique ;
Hunuk, Burak ;
Velagic, Vedran ;
Brugada, Pedro ;
Chierchia, Gian-Battista .
HEART RHYTHM, 2015, 12 (04) :673-680
[4]   Pulmonary vein isolation as index procedure for persistent atrial fibrillation: One-year clinical outcome after ablation using the second-generation cryoballoon [J].
Ciconte, Giuseppe ;
Ottaviano, Luca ;
Deasmundis, Carlo ;
Baltogiannis, Giannis ;
Conte, Giulio ;
Sieira, Juan ;
Di Giovanni, Giacomo ;
Saitoh, Yukio ;
Irfan, Ghazala ;
Mugnai, Giacomo ;
Storti, Cesare ;
Montenero, Annibale Sandro ;
Chierchia, Gian-Battista ;
Brugada, Pedro .
HEART RHYTHM, 2015, 12 (01) :60-66
[5]   Reduced incidence of esophageal lesions by luminal esophageal temperature-guided second-generation cryoballoon ablation [J].
Fuernkranz, Alexander ;
Bordignon, Stefano ;
Boehmig, Michael ;
Konstantinou, Athanasios ;
Dugo, Daniela ;
Perrotta, Laura ;
Klopffleisch, Tom ;
Nowak, Bernd ;
Dignass, Axel U. ;
Schmidt, Boris ;
Chun, Julian K. R. .
HEART RHYTHM, 2015, 12 (02) :268-274
[6]   Improved 1-Year Clinical Success Rate of Pulmonary Vein Isolation with the Second-Generation Cryoballoon in Patients with Paroxysmal Atrial Fibrillation [J].
Fuernkranz, Alexander ;
Bordignon, Stefano ;
Dugo, Daniela ;
Perotta, Laura ;
Gunawardene, Melanie ;
Schulte-Hahn, Britta ;
Nowak, Bernd ;
Schmidt, Boris ;
Chun, Julian K. R. .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2014, 25 (08) :840-844
[7]   Second-generation cryoballoon-based pulmonary vein isolation: Lessons from a five-year follow-up [J].
Heeger, Christian-Hendrik ;
Subin, Behnam ;
Wissner, Erik ;
Fink, Thomas ;
Mathew, Shibu ;
Maurer, Tilman ;
Lemes, Christine ;
Rillig, Andreas ;
Wohlmuth, Peter ;
Reissmann, Bruno ;
Tilz, Roland Richard ;
Ouyang, Feifan ;
Kuck, Karl-Heinz ;
Metzner, Andreas .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2020, 312 :73-80
[8]   Reconduction After Second-Generation Cryoballoon-Based Pulmonary Vein Isolation - Impact of Different Ablation Strategies - [J].
Heeger, Christian-Hendrik ;
Rexha, Enida ;
Maack, Sabrina ;
Rottner, Laura ;
Fink, Thomas ;
Mathew, Shibu ;
Maurer, Tilman ;
Lemes, Christine ;
Rillig, Andreas ;
Wohlmuth, Peter ;
Reissmann, Bruno ;
Tilz, Roland Richard ;
Ouyang, Feifan ;
Kuck, Karl-Heinz ;
Metzner, Andreas .
CIRCULATION JOURNAL, 2020, 84 (06) :902-910
[9]   Efficacy and Safety of Cryoballoon Ablation in Patients With Heart Failure and Reduced Left Ventricular Ejection Fraction A Multicenter Study [J].
Heeger, Christian-Hendrik ;
Abdin, Amr ;
Mathew, Shibu ;
Reissmann, Bruno ;
Yalin, Kivanc ;
Liosis, Spyridon ;
Fink, Thomas ;
Proietti, Riccardo ;
Eitel, Charlotte ;
Vogler, Julia ;
Lemes, Christine ;
Maurer, Tilman ;
Rillig, Andreas ;
Meyer-Saraei, Roza ;
Graf, Tobias ;
Wohlmuth, Peter ;
Goldmann, Britta ;
Ouyang, Feifan ;
Kuck, Karl-Heinz ;
Metzner, Andreas ;
Tilz, Roland Richard .
CIRCULATION JOURNAL, 2019, 83 (08) :1653-1659
[10]   Once Isolated, Always Isolated? Incidence and Characteristics of Pulmonary Vein Reconduction After Second-Generation Cryoballoon-Based Pulmonary Vein Isolation [J].
Heeger, Christian-Hendrik ;
Wissner, Erik ;
Mathew, Shibu ;
Deiss, Sebastian ;
Lemes, Christine ;
Rillig, Andreas ;
Wohlmuth, Peter ;
Reissmann, Bruno ;
Tilz, Roland Richard ;
Ouyang, Feifan ;
Kuck, Karl-Heinz ;
Metzner, Andreas .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2015, 8 (05) :1088-1094