Intraocular pressure measurement with Corvis ST in comparison with applanation tonometry and Tomey non-contact tonometry

被引:10
作者
Luebke, Jan [1 ,2 ]
Bryniok, L. [3 ]
Neuburger, M. [4 ]
Jordan, J. F. [5 ]
Boehringer, D. [1 ,2 ]
Reinhard, T. [1 ,2 ]
Wecker, T. [1 ,2 ,6 ]
Anton, A. [1 ,2 ]
机构
[1] Univ Freiburg, Univ Hosp Freiburg, Eye Ctr, Med Ctr, Killianstr 5, D-79106 Freiburg, Germany
[2] Albert Ludwigs Univ Freiburg, Fac Med, Freiburg, Germany
[3] AurichDent, Aurich, Germany
[4] Augenarztpraxis Dres Neuburger Burau Schmidt, Achern, Germany
[5] Berufsausubungsgemeinschaft GbR Dr Med Michael Vo, Frankfurt, Germany
[6] Augenarztpraxis Dr Wecker, Heilbronn, Germany
关键词
Glaucoma; Intraocular pressure; Corvis ST; NCT; GAT; BIOMECHANICAL PROPERTIES; OCULAR HYPERTENSION; CORNEAL THICKNESS; ICARE;
D O I
10.1007/s10792-019-01098-5
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose Intraocular pressure (IOP) measurement can be performed with different methods. Newer methods have to be compared to the standard method, the Goldmann applanation tonometry (GAT). We herein compare two air-puff tonometers, the non-contact tonometer (Tomey NCT) and the Corvis ST (CST) with GAT in eyes with a broad spectrum of IOP. Methods Two hundred and forty-nine eyes of 249 patients (with diagnosis of either glaucoma or ocular hypertension) were included in this monocenter prospective cohort study. Each eye underwent IOP measurements via GAT, NCT and CST. Bland-Altman plots were calculated to compare the different methods in the three groups. Paired t tests were used for statistical comparison between the three measurement methods. The difference between the different methods was tested on correlation against central corneal thickness (CCT). Results Mean IOP in GAT was 17.6 mmHg (standard deviation (SD) 5.9), 16.3 mmHg (SD 5.6) in NCT and 18.0 mmHg (SD 5.5) in CST. Comparisons between GAT and CST vs. NCT showed significant differences (p < 0.001), while GAT vs. CST showed no significant difference (p = 0.1162). Mean CCT was 538.7 mu m (SD 35.1). Conclusions Mean values of GAT and CST show comparable results. However, both GAT and CST differ significantly from NCT. NCT shows lower IOP values compared to both other methods.
引用
收藏
页码:2517 / 2521
页数:5
相关论文
共 22 条
[1]  
Baneros-Rojas P, 2014, Arch Soc Esp Oftalmol, V89, P260, DOI 10.1016/j.oftal.2014.02.001
[2]   Systematic Review of the Agreement of Tonometers with Goldmann Applanation Tonometry [J].
Cook, Jonathan Alistair ;
Botello, Adriana Paola ;
Elders, Andrew ;
Ali, Alia Fathi ;
Azuara-Blanco, Augusto ;
McCormack, Kirsty ;
Burr, Jennifer Margaret .
OPHTHALMOLOGY, 2012, 119 (08) :1552-1557
[3]   Intraocular pressure and corneal thickness - A comparison between non-contact tonometry and applanation tonometry [J].
Domke, N. ;
Hager, A. ;
Wiegand, W. .
OPHTHALMOLOGE, 2006, 103 (07) :583-587
[4]   Latanoprost for open-angle glaucoma (UKGTS): a randomised, multicentre, placebo-controlled trial [J].
Garway-Heath, David F. ;
Crabb, David P. ;
Bunce, Catey ;
Lascaratos, Gerassimos ;
Amalfitano, Francesca ;
Anand, Nitin ;
Azuara-Blanco, Augusto ;
Bourne, Rupert R. ;
Broadway, David C. ;
Cunliffe, Ian A. ;
Diamond, Jeremy P. ;
Fraser, Scott G. ;
Ho, Tuan A. ;
Martin, Keith R. ;
McNaught, Andrew I. ;
Negi, Anil ;
Patel, Krishna ;
Russell, Richard A. ;
Shah, Ameet ;
Spry, Paul G. ;
Suzuki, Katsuyoshi ;
White, Edward T. ;
Wormald, Richard P. ;
Xing, Wen ;
Zeyen, Thierry G. .
LANCET, 2015, 385 (9975) :1295-1304
[5]  
GOLDMANN H, 1957, Ophthalmologica, V134, P221
[6]  
Gordon MO, 2002, ARCH OPHTHALMOL-CHIC, V120, P714
[7]   Comparison of the iCare tonometer with the Goldmann tonometer in Malawi [J].
Hohmann, J. ;
Schulze-Schwering, M. ;
Nyaka, T. Chirambo ;
Moyo, V. ;
Kayange, P. C. ;
Doycheva, D. ;
Batumba, N. H. ;
Spitzer, M. S. .
OPHTHALMOLOGE, 2012, 109 (11) :1098-1102
[8]   A New Tonometer-The Corvis ST Tonometer: Clinical Comparison with Noncontact and Goldmann Applanation Tonometers [J].
Hong, Jiaxu ;
Xu, Jianjiang ;
Wei, Anji ;
Deng, Sophie X. ;
Cui, Xinhan ;
Yu, Xiaobo ;
Sun, Xinghuai .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (01) :659-665
[9]   Repeatability of intra-ocular pressure and central corneal thickness measurements provided by a non-contact method of tonometry and pachymetry [J].
Lomoriello, Domenico Schiano ;
Lombardo, Marco ;
Tranchina, Laura ;
Oddone, Francesco ;
Serrao, Sebastiano ;
Ducoli, Pietro .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2011, 249 (03) :429-434
[10]   More than 4 years' experience with electronic intraocular needle tonometry [J].
Marx, W ;
Madjlessi, F ;
Reinhard, T ;
Althaus, C ;
Sundmacher, R .
OPHTHALMOLOGE, 1999, 96 (08) :498-502