High-Quality Tissue Imaging Using a Catheter-Based Swept-Source Optical Coherence Tomography Systems With an Integrated Semiconductor Optical Amplifier

被引:25
作者
Mao, Youxin [1 ]
Flueraru, Costel [1 ]
Chang, Shoude [1 ]
Popescu, Dan Paul [1 ]
Sowa, Michael G. [1 ]
机构
[1] Natl Res Council Canada, Ottawa, ON K1A 0R6, Canada
关键词
Biomedical imaging; optical coherence tomography (OCT); optical fiber probe; quadrature interferometer; semiconductor amplifier; INTERFEROMETER; DOMAIN;
D O I
10.1109/TIM.2011.2126950
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We present the analysis and characteristics of an optical fiber catheter-based swept-source optical coherence tomography (SS-OCT) system for tissue imaging. A quadrature Mach-Zehnder interferometer based on multiport fiber couplers and a semiconductor optical amplifier were employed in our SS-OCT system in order to increase the signal-to-noise ratio, therefore improving image quality deeper within the tissue. We demonstrate OCT images of human skin and arterial tissue acquired with our OCT engine. The results show that this OCT system is an effective imaging modality for real-time diagnosis and in vivo biological applications.
引用
收藏
页码:3376 / 3383
页数:8
相关论文
共 23 条
[1]  
Andersen P. E., 2005, US Patent, Patent No. [6900943 B2, 6900943]
[2]   Current and future developments in intracoronary optical coherence tomography imaging [J].
Barlis, Peter ;
Schmitt, Joseph M. .
EUROINTERVENTION, 2009, 4 (04) :529-533
[3]   Instantaneous quadrature low-coherence interferometry with 3 x 3 fiber-optic couplers [J].
Choma, MA ;
Yang, CH ;
Izatt, JA .
OPTICS LETTERS, 2003, 28 (22) :2162-2164
[4]   OPTICAL COHERENCE TOMOGRAPHY [J].
HUANG, D ;
SWANSON, EA ;
LIN, CP ;
SCHUMAN, JS ;
STINSON, WG ;
CHANG, W ;
HEE, MR ;
FLOTTE, T ;
GREGORY, K ;
PULIAFITO, CA ;
FUJIMOTO, JG .
SCIENCE, 1991, 254 (5035) :1178-1181
[5]   Visualization of coronary atherosclerotic plaques in patients using optical coherence tomography: Comparison with intravascular ultrasound [J].
Jang, IK ;
Bouma, BE ;
Kang, DH ;
Park, SJ ;
Park, SW ;
Seung, KB ;
Choi, KB ;
Shishkov, M ;
Schlendorf, K ;
Pomerantsev, E ;
Houser, SL ;
Aretz, HT ;
Tearney, GJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (04) :604-609
[6]   Improved sensitivity of optical frequency domain reflectometry-optical coherence tomography using a semiconductor optical amplifier [J].
Jayavel, Pachamuthu ;
Amano, Takuji ;
Choi, DongHak ;
Furukawa, Hiroyuki ;
Hiro-Oka, Hideaki ;
Asakai, Kota ;
Ohbayashi, Kohji .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 2-LETTERS & EXPRESS LETTERS, 2006, 45 (46-50) :L1317-L1319
[7]  
KORDE VR, 2005, FRONT OPT FIO TUCS A
[8]   Imaging needle for optical coherence tomography [J].
Li, XD ;
Chudoba, C ;
Ko, T ;
Pitris, C ;
Fujimoto, JG .
OPTICS LETTERS, 2000, 25 (20) :1520-1522
[9]  
MAO Y, 2008, P SOC PHOTO-OPT INS, V6826
[10]   3x3 Mach-Zehnder interferometer with unbalanced differential detection for full-range swept-source optical coherence tomography [J].
Mao, Youxin ;
Sherif, Sherif ;
Flueraru, Costel ;
Chang, Shoude .
APPLIED OPTICS, 2008, 47 (12) :2004-2010