摘要
This article describes the design, fabrication and characterization of a compact and deployable heavy metal analyzer. The system is autonomous and its micro-controller enables the configuration of its four micropumps and the application of different electroanalytical techniques. A range of prototyping and manufacturing techniques have been used in the fabrication of the instrument, including laser sintering, milling, and laser engraving. We present the electrochemical characterization of the system, and provide preliminary results obtained in the determination of Cu and Zn in different waters. The only sample treatment consists in its 1 : 2 dilution in pH 5.6 acetate buffer. Detection limits of 2.2 and 3.8 ppb were found for Cu and Zn, respectively.
引用
收藏
页码:929 / 937
页数:9
相关论文
共 31 条
[21]   Diagnostics of Anodic Stripping Mechanisms under Square-Wave Voltammetry Conditions Using Bismuth Film Substrates [J].
Mirceski, Valentin ;
Hocevar, Samo B. ;
Ogorevc, Bozidar ;
Gulaboski, Rubin ;
Drangov, Ivan .
ANALYTICAL CHEMISTRY, 2012, 84 (10) :4429-4436
[22]   A continuous flow system design for simultaneous determination of heavy metals in river water samples [J].
Muñoz, E ;
Palmero, S ;
García-García, MA .
TALANTA, 2002, 57 (05) :985-992
[23]   REVIEW: ELECTROCHEMICAL STRIPPING ANALYSIS OF TRACE HEAVY METALS USING SCREEN-PRINTED ELECTRODES [J].
Niu, Xiangheng ;
Lan, Minbo ;
Zhao, Hongli ;
Chen, Chen ;
Li, Yuxiu ;
Zhu, Xiang .
ANALYTICAL LETTERS, 2013, 46 (16) :2479-2502
[24]   Comparison of Different PLS Algorithms for Simultaneous Determination of Cd(II), Cu(II), Pb(II), and Zn(II) by Anodic Stripping Voltammetry at Bismuth Film Electrode [J].
Pinto, Licarion ;
Lemos, Sherlan G. .
ELECTROANALYSIS, 2014, 26 (02) :299-305
[25]   Stripping chronopotentiometry in environmental analysis [J].
Serrano, Nuria ;
Diaz-Cruz, Jose Manuel ;
Arino, Cristina ;
Esteban, Miquel .
ELECTROANALYSIS, 2007, 19 (19-20) :2039-2049
[26]   Sequential injection anodic stripping voltammetry with monosegmented flow and in-line UV digestion for determination of Zn(II), Cd(II), Pb(II) and Cu(II) in water samples [J].
Siriangkhawut, Watsaka ;
Grudpan, Kate ;
Jakmunee, Jaroon .
TALANTA, 2011, 84 (05) :1366-1373
[27]   Microstructure fabrication with a CO2 laser system [J].
Snakenborg, D ;
Klank, H ;
Kutter, JP .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2004, 14 (02) :182-189
[28]   A Decade with Bismuth-Based Electrodes in Electroanalysis [J].
Svancara, Ivan ;
Prior, Chad ;
Hocevar, Samo B. ;
Wang, Joseph .
ELECTROANALYSIS, 2010, 22 (13) :1405-1420
[29]   Simultaneous determination of copper, lead, cadmium and zinc using differential pulse anodic stripping voltammetry in a flow system [J].
van Staden, JF ;
Matoetoe, MC .
ANALYTICA CHIMICA ACTA, 2000, 411 (1-2) :201-207
[30]   FIELD-BASED HEAVY-METAL ANALYZER FOR THE SIMULTANEOUS DETERMINATION OF MULTIPLE CATIONS ON-SITE [J].
WILLIAMS, G ;
DSILVA, C .
ANALYST, 1994, 119 (11) :2337-2341