MATHEMATICAL MODELING OF THE RADIATION DOSE RECEIVED FROM PHOTONS PASSING OVER AND THROUGH SHIELDING WALLS IN A PET/CT SUITE

被引:6
作者
Fog, Lotte S. [1 ]
Cormack, John [1 ]
机构
[1] Flinders Med Ctr, Bedford Pk, SA 5042, Australia
来源
HEALTH PHYSICS | 2010年 / 99卷 / 06期
关键词
computed tomography; dose; Monte Carlo; positron emission tomography;
D O I
10.1097/HP.0b013e3181e47a39
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Given that the financial cost of shielding PET/CT suites can be substantial, it has become increasingly important to be able to accurately assess the thickness of shielding required for barriers and whether it is necessary to extend such shielding all the way to the ceiling. The overall shielding requirement for a PET/CT installation must take into account both 511 keV gamma ray emissions from PET scans and lower energy x-ray scatter from CT scans. This paper deals with the overall impact of emissions from both modalities. Radiation exposure from both scatter over shielding barriers as well as transmission through these barriers is taken into account. A series of simulations of the dose received by a person positioned behind a shielding barrier in a typical PET/CT scanning suite were carried out using both Monte Carlo and analytical models. The transmission through lead barriers was found to be very dependent on the geometry of the radiation source and the resulting energy spectrum of the emitted radiation. The transmission from a patient source was found to be around half of that from a small vial and also half of that reported previously using parallel beams of mono-energetic radiation. For PET emissions, the dose from scatter over the barrier at waist height is relatively small but may have to be taken into account if the design dose limit is low. Shielding from floor to ceiling is probably not warranted in most instances for PET gamma emissions; in PET/CT installations, however, a thinner layer of shielding may need to extend to the ceiling of the imaging room to limit x-ray scatter over the wall from the CT unit. Health Phys. 99(6):769-779; 2010
引用
收藏
页码:769 / 779
页数:11
相关论文
共 11 条
[1]   DIAGNOSTIC-X-RAY SHIELDING DESIGN BASED ON AN EMPIRICAL-MODEL OF PHOTON ATTENUATION [J].
ARCHER, BR ;
THORNBY, JI ;
BUSHONG, SC .
HEALTH PHYSICS, 1983, 44 (05) :507-517
[2]  
BLAKE W, 2006, MED PHYS, V33, P3046
[3]  
FIX MK, 2007, J PHYS CONF SER, V74, pNIL34
[4]  
Kawrakow I., 2006, EGSNRC CODE SYSTEM M
[5]  
Madsen M., 2006, MED PHYS, V31
[6]   A study of scatter in diagnostic X-ray rooms [J].
McVey, G ;
Weatherburn, H .
BRITISH JOURNAL OF RADIOLOGY, 2004, 77 (913) :28-38
[7]  
*NAT COUNC RAD PRO, 2005, 147 NCRP MEAS
[8]   Fifty years of Monte Carlo simulations for medical physics [J].
Rogers, D. W. O. .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (13) :R287-R301
[9]   SHIELDING REQUIREMENTS FOR CONSTANT-POTENTIAL DIAGNOSTIC-X-RAY BEAMS DETERMINED BY A MONTE-CARLO CALCULATION [J].
SIMPKIN, DJ .
HEALTH PHYSICS, 1989, 56 (02) :151-164
[10]   TRANSMISSION DATA FOR SHIELDING DIAGNOSTIC-X-RAY FACILITIES [J].
SIMPKIN, DJ .
HEALTH PHYSICS, 1995, 68 (05) :704-709