Reducing Channel Interaction Through Cochlear Implant Programming May Improve Speech Perception: Current Focusing and Channel Deactivation

被引:59
作者
Bierer, Julie A. [1 ]
Litvak, Leonid [2 ]
机构
[1] Univ Washington, Seattle, WA 98195 USA
[2] Adv Bion Corp, Valencia, CA USA
关键词
cochlear implant; electrode configuration; channel selection; phoneme perception; speech perception; ELECTRODE-NEURON INTERFACE; SPATIAL TUNING CURVES; PARTIAL TRIPOLAR; USERS; RECOGNITION; STIMULATION; THRESHOLD; CONFIGURATION; PLACEMENT; PATTERNS;
D O I
10.1177/2331216516653389
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Speech perception among cochlear implant (CI) listeners is highly variable. High degrees of channel interaction are associated with poorer speech understanding. Two methods for reducing channel interaction, focusing electrical fields, and deactivating subsets of channels were assessed by the change in vowel and consonant identification scores with different program settings. The main hypotheses were that (a) focused stimulation will improve phoneme recognition and (b) speech perception will improve when channels with high thresholds are deactivated. To select high-threshold channels for deactivation, subjects' threshold profiles were processed to enhance the peaks and troughs, and then an exclusion or inclusion criterion based on the mean and standard deviation was used. Low-threshold channels were selected manually and matched in number and apex-to-base distribution. Nine ears in eight adult CI listeners with Advanced Bionics HiRes90k devices were tested with six experimental programs. Two, all-channel programs, (a) 14-channel partial tripolar (pTP) and (b) 14-channel monopolar (MP), and four variable-channel programs, derived from these two base programs, (c) pTP with high- and (d) low-threshold channels deactivated, and (e) MP with high-and (f) low-threshold channels deactivated, were created. Across subjects, performance was similar with pTP and MP programs. However, poorer performing subjects (scoring< 62% correct on vowel identification) tended to perform better with the all-channel pTP than with the MP program (I > 2). These same subjects showed slightly more benefit with the reduced channel MP programs (5 and 6). Subjective ratings were consistent with performance. These finding suggest that reducing channel interaction may benefit poorer performing CI listeners.
引用
收藏
页数:12
相关论文
共 33 条
[1]   Comparing spatial tuning curves, spectral ripple resolution, and speech perception in cochlear implant users [J].
Anderson, Elizabeth S. ;
Nelson, David A. ;
Kreft, Heather ;
Nelson, Peggy B. ;
Oxenham, Andrew J. .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2011, 130 (01) :364-375
[2]   Current steering and current focusing in cochlear implants: Comparison of monopolar, tripolar, and virtual channel electrode configurations [J].
Berenstein, Carlo K. ;
Mens, Lucas H. M. ;
Mulder, Jef J. S. ;
Vanpoucke, Filiep J. .
EAR AND HEARING, 2008, 29 (02) :250-260
[3]   A Fast Method for Measuring Psychophysical Thresholds Across the Cochlear Implant Array [J].
Bierer, Julie A. ;
Bierer, Steven M. ;
Kreft, Heather A. ;
Oxenham, Andrew J. .
TRENDS IN HEARING, 2015, 19
[4]   Threshold and channel interaction in cochlear implant users: Evaluation of the tripolar electrode configuration [J].
Bierer, Julie Arenberg .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2007, 121 (03) :1642-1653
[5]  
Bierer JA, 2014, EAR HEARING, V35, P641, DOI 10.1097/AUD.0000000000000058
[6]   Probing the Electrode-Neuron Interface With Focused Cochlear Implant Stimulation [J].
Bierer, Julie Arenberg .
TRENDS IN AMPLIFICATION, 2010, 14 (02) :84-95
[7]   Identifying Cochlear Implant Channels With Poor Electrode-Neuron Interfaces: Electrically Evoked Auditory Brain Stem Responses Measured With the Partial Tripolar Configuration [J].
Bierer, Julie Arenberg ;
Faulkner, Kathleen F. ;
Tremblay, Kelly L. .
EAR AND HEARING, 2011, 32 (04) :436-444
[8]   Identifying Cochlear Implant Channels with Poor Electrode-Neuron Interface: Partial Tripolar, Single-Channel Thresholds and Psychophysical Tuning Curves [J].
Bierer, Julie Arenberg ;
Faulkner, Kathleen F. .
EAR AND HEARING, 2010, 31 (02) :247-258
[9]   Effects of stimulation mode, level and location on forward-masked excitation patterns in cochlear implant patients [J].
Chatterjee, M ;
Galvin, JJ ;
Fu, QJ ;
Shannon, RV .
JARO-JOURNAL OF THE ASSOCIATION FOR RESEARCH IN OTOLARYNGOLOGY, 2006, 7 (01) :15-25
[10]   Assessing the Electrode-Neuron Interface with the Electrically Evoked Compound Action Potential, Electrode Position, and Behavioral Thresholds [J].
Devries, Lindsay ;
Scheperle, Rachel ;
Bierer, Julie Arenberg .
JARO-JOURNAL OF THE ASSOCIATION FOR RESEARCH IN OTOLARYNGOLOGY, 2016, 17 (03) :237-252