Reliability of left ventricular ejection fraction calculated with gated myocardial perfusion single photon emission computed tomography in patients with extensive perfusion defect

被引:10
作者
Harisankar, Chidambaram Natrajan Balasubramanian [1 ]
Mittal, Bhagwant Rai [1 ]
Kamaleshwaran, Koramadai Karuppuswamy [1 ]
Parmar, Madan [1 ]
Bhattacharya, Anish [1 ]
Singh, Baljinder [1 ]
Mahajan, Rajiv [2 ]
机构
[1] Postgrad Inst Med Educ & Res, Dept Nucl Med, Chandigarh 160012, India
[2] Postgrad Inst Med Educ & Res, Dept Cardiol, Chandigarh 160012, India
关键词
echocardiography; ejection fraction; gated single photon emission computed tomography; radionuclide ventriculography; BLOOD-POOL TOMOGRAPHY; AUTOMATIC QUANTIFICATION; RADIONUCLIDE ANGIOGRAPHY; NORMAL LIMITS; SPECT; TL-201; VOLUMES; ACCURACY; ECHOCARDIOGRAPHY; IMAGES;
D O I
10.1097/MNM.0b013e328342b97e
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction Gated myocardial perfusion scintigraphy (g-MPS) provides functional information on the left ventricle (LV) apart from the perfusion status. Computer algorithm-based LV ejection fraction (EF) calculated from resting g-MPS has been found to be reliable in patients with normal perfusion. However, its role in patients with extensive perfusion defects is not clear. Aim To find the reliability of LVEF calculated from resting g-MPS in patients with extensive perfusion defects (>25% of LV myocardium) and to correlate this with echocardiography and multigated radionuclide ventriculography (RNV). Materials and methods Thirty patients with fixed perfusion defect of size greater than or equal to 25% of LV myocardium on rest g-M PS study were included. EF was calculated using three software packages: Emory Cardiac Toolbox, Myometrix, and quantitative gated single photon emission computed tomography (SPECT)/quantitative perfusion SPECT. The patients underwent RNV (gold standard) and echocardiography within a week of the g-M PS. Agreement among the EF values obtained by different methods was determined using Bland-Altman analysis. Correlation among the EF values was measured using Spearman's rank correlation. Results Thirty patients (23 male; seven female; mean age 51 years, range 32-70 years) were included prospectively. The average size of perfusion defect was 38% of the LV myocardium (range 25-56% of LV). The average ejection fraction values were 33% for Emory Cardiac Toolbox (range 11-50%), 310/0 for Myometrix (range 18-46%), and 33% for quantitative gated SPECT/quantitative perfusion SPECT (range 17-49%). The mean EF on echocardiography was 37% (range 22-60%), whereas that of RNV was 33% (range 10-50%). The data were normally distributed. There was statistically significant positive agreement between algorithm-based EF measurements to the gold standard RNV. Conclusion This study suggests that EF calculations, from g-MPS SPECT data using different software, have high agreement with the gold standard RNV even in patients with extensive perfusion defects. Nucl Med Commun 32:503-507 (C) 2011 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:503 / 507
页数:5
相关论文
共 23 条
[1]   Normal limits for left ventricular ejection fraction and volumes estimated with gated myocardial perfusion imaging in patients with normal exercise test results: Influence of tracer, gender, and acquisition camera [J].
Ababneh, AA ;
Sciacca, RR ;
Kim, B ;
Bergmann, SR .
JOURNAL OF NUCLEAR CARDIOLOGY, 2000, 7 (06) :661-668
[2]   An investigation of the estimation of ejection fractions and cardiac volumes by a quantitative gated SPECT software package in simulated gated SPECT images [J].
Achtert, AD ;
King, MA ;
Dahlberg, ST ;
Pretorius, PH ;
LaCroix, KJ ;
Tsui, BMW .
JOURNAL OF NUCLEAR CARDIOLOGY, 1998, 5 (02) :144-152
[3]   Thallium-201 gated single-photon emission tomography for the assessment of left ventricular ejection fraction and regional wall motion abnormalities in comparison with two-dimensional echocardiography [J].
Bacher-Stier, C ;
Müller, S ;
Pachinger, O ;
Strolz, S ;
Erler, H ;
Moncayo, R ;
Wenger, M ;
Donnemiller, E ;
Riccabona, G .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1999, 26 (12) :1533-1540
[4]   QUANTITATIVE GATED BLOOD POOL TOMOGRAPHIC ASSESSMENT OF REGIONAL EJECTION FRACTION - DEFINITION OF NORMAL LIMITS [J].
CERQUEIRA, MD ;
HARP, GD ;
RITCHIE, JL .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 20 (04) :934-941
[5]  
Cwajg E, 1999, J NUCL MED, V40, P1857
[6]   Comparative value of ECG-gated blood pool SPET and ECG-gated myocardial perfusion SPET in the assessment of global systolic left ventricular function [J].
Daou, D ;
Vilain, D ;
Colin, P ;
Lebtahi, R ;
Fourme, T ;
Coaguila, C ;
Benada, A ;
Idy-Peretti, I ;
Slama, M ;
Le Guludec, D .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (06) :859-+
[7]  
Daou D, 2001, J NUCL MED, V42, P1043
[8]  
Germano G, 1997, J NUCL MED, V38, P749
[9]  
GERMANO G, 1995, J NUCL MED, V36, P2138
[10]  
He ZX, 1999, J NUCL CARDIOL, V6, P412