Comparing keratometry readings with manual separation of lids and wire speculum in children under general anesthesia

被引:5
作者
Jethani, Jitendra [1 ]
Porwal, Kavita [3 ]
Porwal, Amit [3 ]
Dave, Paaraj [2 ]
Lalwani, Syma [1 ]
Trivedi, Mihika [1 ]
机构
[1] Baroda Children Eyecare & Squint Clin, SF 212-213,Panorama Complex,RC Dutt Rd, Vadodara 390007, Gujarat, India
[2] Dave Eye Clin, Vadodara, Gujarat, India
[3] Sanghavi Netralaya, Belgaum, Karnataka, India
关键词
Children; keratometry reading; supine position; CORNEAL POWER MEASUREMENTS;
D O I
10.4103/ijo.IJO_269_21
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: Keratometry (K) readings are crucial for intraocular lens power calculation in cataract surgery. In children who do not cooperate, the keratometry is done under general anesthesia with a handheld autokeratometer. However, there is little consensus regarding the method for the measurement of K readings. The lids can be separated either by fingers or a wire speculum may be placed to separate the lids for measurement. Methods: The children selected for the study were patients cooperative for keratometry reading. Nidek KM-500 handheld keratometer was used first in the awake period. Then under general anesthesia, readings were taken first by separating the lids manually with fingers and then after putting a wire speculum in both the eyes. Results: The average keratometry reading for participants in the OPD, anesthetized with lids manually opened and with lids separated with speculum was 44.7 +/- 1.7 D, 44.4 +/- 1.9 D, and 44.7 +/- 1.7 D, respectively. Conclusion: No significant change was observed in keratometry values in children with manual separation of eyelids or with wire speculum.
引用
收藏
页码:2050 / 2052
页数:3
相关论文
共 13 条
[1]  
Chang Minwook, 2012, Korean J Ophthalmol, V26, P10, DOI 10.3341/kjo.2012.26.1.10
[2]  
Edwards K, 2009, Optometry: Science, Techniques and Clinical Management, P257
[3]   Origins of the Keratometer and its Evolving Role in Ophthalmology [J].
Gutmark, Ron ;
Guyton, David L. .
SURVEY OF OPHTHALMOLOGY, 2010, 55 (05) :481-497
[4]  
Horner DG, 2006, Borish's Clinical Refraction, V2, P645
[5]   Comparison of corneal power, corneal astigmatism, and axis location in normal eyes obtained from an autokeratometer and a corneal topographer [J].
Kobashi, Hidenaga ;
Kamiya, Kazutaka ;
Igarashi, Akihito ;
Ishii, Rie ;
Sato, Nobuyuki ;
Wang, Guoqin ;
Shimizu, Kimiya .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (04) :648-654
[6]   Effect of instrument rotation on handheld keratometry [J].
Lam, AKC ;
Chan, RF ;
Chiu, R .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (12) :2590-2594
[7]   Effect of posture and artificial tears on corneal power measurements with a handheld automated keratometer [J].
Lam, AKC ;
Chan, R ;
Chiu, R .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (03) :645-652
[8]  
Ledford JK., 2006, The slit lamp primer, V2
[9]   Comparison of portable automated keratometry and manual keratometry for IOL calculation [J].
Manning, CA ;
Kloess, PM .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1997, 23 (08) :1213-1216
[10]  
Noonan C P, 1998, J AAPOS, V2, P186, DOI 10.1016/S1091-8531(98)90013-1