Biplane three-dimensional augmented fluoroscopy as single navigation tool for ablation of atrial fibrillation:: Accuracy and clinical value

被引:55
作者
Ector, Joris [1 ]
De Buck, Stijn [2 ]
Huybrechts, Wim [1 ]
Nuyens, Dieter [1 ]
Dymarkowski, Steven [3 ]
Bogaert, Jan [3 ]
Maes, Frederik [2 ]
Heidbuchel, Hein [1 ]
机构
[1] Univ Hosp Gasthuisberg, Dept Cardiol, B-3000 Louvain, Belgium
[2] Univ Hosp Gasthuisberg, Dept Elect Engn, B-3000 Louvain, Belgium
[3] Univ Hosp Gasthuisberg, Dept Radiol, B-3000 Louvain, Belgium
关键词
atrial fibrillation; ablation; image integration; fluoroscopy; computed tomography;
D O I
10.1016/j.hrthm.2008.03.024
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND We developed new methods for real time biplane integration of three-dimensional (3D) Left atrial models with fluoroscopic images to assist in catheter ablation of atrial fibrillation (AF). OBJECTIVE The purpose of this study was to quantitatively assess the accuracy of 3D fluoroscopy integration and to evaluate its clinical value when used as a single navigation tool for AF ablation. METHODS Sixty patients underwent AF ablation under biplane fluoroscopic guidance after selective angiography of the four pulmonary veins. Computed tomography [CT]-based 3D models were integrated in the fluoroscopic framework using visual matching and landmark-based registration approaches. Integration accuracy was quantitatively assessed according to registration approach and different CT acquisition parameters (electrocardiogram [ECG] gating, respiratory phase). In 30 of the 60 patients (3D+ group), the integrated 3D model was used for real time 3D-augmented fluoroscopic catheter navigation, and the effects on procedural parameters and patient radiation dose were evaluated. RESULTS Landmark-based registration resulted in superior 3D fluoroscopy integration accuracy compared with the visual matching approach (P <.001 for alignment error and alignment score). The effects of ECG gating and respiratory phase during CT acquisition on integration accuracy were small and clinically irrelevant. The use of 3D-augmented fluoroscopy in the 3D+ group was gauged as extremely helpful by the operator. It resulted in a significant reduction of fluoroscopy time (61 +/- 18 minutes vs. 77 26 minutes; P = .009) and a trend toward shorter procedure duration (230 +/- 67 minutes vs. 257 +/- 58 minutes; P = .06) versus conventional procedures. The systematic use of nongated cardiac CT in the 3D+ group resulted in an important reduction in total effective patient radiation dose due to CT+ fluoroscopy (4 + 14 18 +/- 8 mSv vs. 17 + 16 = 33 +/- 13 mSv; P <.001). CONCLUSIONS Biplane 3D-augmented fluoroscopy can be used as a safe and accurate stand-atone method to guide AF ablation procedures. The use of nongated cardiac CT substantially reduces total patient radiation dose without a relevant reduction in integration accuracy.
引用
收藏
页码:957 / 964
页数:8
相关论文
共 17 条
[1]  
[Anonymous], 2000, 16262EN EUR EUR COMM
[2]   HRS/EHRA/ECAS expert consensus statement on catheter and surgical ablation of atrial fibrillation: Recommendations for personnel, policy, procedures and follow-up [J].
Calkins, Hugh ;
Brugada, Josep ;
Packer, Douglas L. ;
Cappato, Riccardo ;
Chen, Shih-Ann ;
Crijns, Harry J. G. ;
Damiano, Ralph J., Jr. ;
Davies, D. Wyn ;
Haines, David E. ;
Haissaguerre, Michel ;
Iesaka, Yoshito ;
Jackman, Warren ;
Jais, Pierre ;
Kottkamp, Hans ;
Kuck, Karl Heinz ;
Lindsay, Bruce D. ;
Marchlinski, Francis E. ;
McCarthy, Patrick M. ;
Mont, J. Lluis ;
Morady, Fred ;
Nademanee, Koontawee ;
Natale, Andrea ;
Pappone, Carlo ;
Prystowsky, Eric ;
Raviele, Antonio ;
Ruskin, Jeremy N. ;
Shemin, Richard J. .
HEART RHYTHM, 2007, 4 (06) :816-861
[3]   An augmented reality system for patient-specific guidance of cardiac catheter ablation procedures [J].
De Buck, S ;
Maes, F ;
Ector, J ;
Bogaert, J ;
Dymarkowski, S ;
Heidbüchel, H ;
Suetens, P .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2005, 24 (11) :1512-1524
[4]  
DEBUCK S, 2007, WORKSH P 10 INT C ME, P65
[5]   Impact of heart rhythm status on registration accuracy of the left atrium for catheter ablation of atrial fibrillation [J].
Dong, Jun ;
Dalal, Darshan ;
Scherr, Daniel ;
Cheema, Aamir ;
Nazarian, Saman ;
Bilchick, Kenneth ;
Almasry, Ibrahim ;
Cheng, Alan ;
Henrikson, Charles A. ;
Spragg, David ;
Marine, Joseph E. ;
Berger, Ronald D. ;
Calkins, Hugh .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2007, 18 (12) :1269-1276
[6]   Cardiac three-dimensional magnetic resonance imaging and fluoroscopy merging -: A new approach for electroanatomic mapping to assist catheter ablation [J].
Ector, J ;
De Buck, S ;
Adams, J ;
Dymarkowski, S ;
Bogaert, J ;
Maes, F ;
Heidbüchel, H .
CIRCULATION, 2005, 112 (24) :3769-3776
[7]  
ECTOR J, 2008, J CARDIOVASC EL 0326
[8]   Obesity is a major determinant of radiation dose in patients undergoing pulmonary vein isolation for atrial fibrillation [J].
Ector, Joris ;
Dragusin, Octavian ;
Adriaenssens, Bert ;
Huybrechts, Wim ;
Willems, Rik ;
Ector, Hugo ;
Heidbuechel, Hein .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 50 (03) :234-242
[9]   High-frequency jet ventilation: Utility in posterior left atrial catheter ablation [J].
Goode, JS ;
Taylor, RL ;
Buffington, CW ;
Klain, MM ;
Schwartzman, D .
HEART RHYTHM, 2006, 3 (01) :13-19
[10]   Use of only a regular diagnostic His-bundle catheter for both fast and reproducible ''Para-Hisian pacing'' and stable right ventricular pacing [J].
Heidbuchel, H ;
Ector, H ;
Adams, J ;
VandeWerf, F .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1997, 8 (10) :1121-1132