PET/CT with intravenous contrast can be used for PET attenuation correction in cancer patients

被引:121
作者
Berthelsen, AK
Holm, S
Loft, A
Klausen, TL
Andersen, F
Hojgaard, L [1 ]
机构
[1] Copenhagen Univ Hosp, Dept Clin Physiol Nucl Med & PET, Copenhagen, Denmark
[2] Copenhagen Univ Hosp, Rigshosp, Ctr Diagnost Invest, Dept Radiol, Copenhagen, Denmark
关键词
positron emission tomography; artefact; computed tomography; contrast media; neoplasms;
D O I
10.1007/s00259-005-1784-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: If the CT scan of a combined PET/CT study is performed as a full diagnostic quality CT scan including intravenous (IV) contrast agent, the quality of the joint PET/CT procedure is improved and a separate diagnostic CT scan can be avoided. CT with IV contrast can be used for PET attenuation correction, but this may result in a bias in the attenuation factors. The clinical significance of this bias has not been established. Our aim was to perform a prospective clinical study where each patient had CT performed with and without IV contrast agent to establish whether PET/CT with IV contrast can be used for PET attenuation without reducing the clinical value of the PET scan. Methods: A uniform phantom study was used to document that the PET acquisition itself is not significantly influenced by the presence of IV contrast medium. Then, 19 patients referred to PET/CT with IV contrast underwent CT scans without, and then with contrast agent, followed by an F-18-fluorodeoxyglucose whole-body PET scan. The CT examinations were performed with identical parameters on a GE Discovery LS scanner. The PET data were reconstructed with attenuation correction based on the two CT data sets. A global comparison of standard uptake value (SUV) was performed, and SUVs in tumour, in non-tumour tissue and in the subclavian vein were calculated. Clinical evaluation of the number and location of lesions on all PET/CT scans was performed twice, blinded and in a different random order, by two independent nuclear medicine specialists. Results: In all patients, the measured global SUV of PET images based on CT with IV contrast agent was higher than the global activity using non-contrast correction. The overall increase in the mean SUV (for two different conversion tables tested) was 4.5 +/- 2.3% and 1.6 +/- 0.5%, respectively. In 11/19 patients, focal uptake was identified corresponding to malignant tumours. Eight out of 11 tumours showed an increased SUVmax (2.9 +/- 3.1%) on the PET images reconstructed using IV contrast. The clinical evaluation performed by the two specialists comparing contrast and non-contrast CT attenuated PET images showed weighted kappa values of 0.92 (doctor A) and 0.82 (doctor B). No contrast-introduced artefacts were found. Conclusion: This study demonstrates that CT scans with IV contrast agent can be used for attenuation correction of the PET data in combined modality PET/CT scanning, without changing the clinical diagnostic interpretation.
引用
收藏
页码:1167 / 1175
页数:9
相关论文
共 23 条
[11]   Estimation of the noise-contributions from blank, transmission and emission scans in PET [J].
Holm, S ;
Toft, P ;
Jensen, M .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (04) :2285-2291
[12]   Review of photon interaction cross section data in the medical and biological context [J].
Hubbell, JH .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (01) :R1-R22
[13]  
JESSEN K, 2000, 16262 EUR OFF OFF P
[14]   CT vs 68Ge attenuation correction in a combined PET/CT system:: evaluation of the effect of lowering the CT tube current [J].
Kamel, E ;
Hany, TF ;
Burger, C ;
Treyer, V ;
Lonn, AHR ;
von Schulthess, GK ;
Buck, A .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2002, 29 (03) :346-350
[15]  
KEYES JW, 1995, J NUCL MED, V36, P1836
[16]   Attenuation correction for a combined 3D PET/CT scanner [J].
Kinahan, PE ;
Townsend, DW ;
Beyer, T ;
Sashin, D .
MEDICAL PHYSICS, 1998, 25 (10) :2046-2053
[17]   Staging of non-small-cell lung cancer with integrated positron-emission tomography and computed tomography [J].
Lardinois, D ;
Weder, W ;
Hany, TF ;
Kamel, EM ;
Korom, S ;
Seifert, B ;
von Schulthess, GK ;
Steinert, HC .
NEW ENGLAND JOURNAL OF MEDICINE, 2003, 348 (25) :2500-2507
[18]  
Lonn AHR, 2004, IEEE NUCL SCI CONF R, P2220
[19]   Effects of nonionic intravenous contrast agents at PET/CT imaging: Phantom and canine studies [J].
Nakamoto, Y ;
Chin, BB ;
Kraitchman, DL ;
Lawler, LP ;
Marshall, LT ;
Wahl, RL .
RADIOLOGY, 2003, 227 (03) :817-824
[20]  
Valk P.E., 2003, POSITRON EMISSION TO