Characteristics of x-ray beams in two commercial multidetector computed tomography simulators: Monte Carlo simulations

被引:11
作者
Kim, Sangroh [1 ]
Song, Haijun [2 ]
Movsas, Benjamin [1 ]
Chetty, Indrin J. [1 ]
机构
[1] Henry Ford Hlth Syst, Dept Radiat Oncol, Detroit, MI 48202 USA
[2] Duke Univ, Dept Radiat Oncol, Med Ctr, Durham, NC 27710 USA
关键词
MDCT simulator; Monte Carlo; BEAMNRC/EGSNRC; MDCT beam; radiochromic film; ORBITAL ELECTRONS; ENERGY-SPECTRA; DOSE INDEX; CT; DOSIMETRY; VALIDATION; DETECTOR; NUCLEI; MODELS; SYSTEM;
D O I
10.1118/1.3670377
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: As multidetector computed tomography (MDCT) scanning is routinely performed for treatment planning in radiation oncology, understanding the characteristics of the MDCT x-ray beam is essential to accurately estimate patient dose. The purpose of this study is to characterize the x-ray beams of two commercial MDCT simulators widely used in radiation oncology by Monte Carlo (MC) simulations. Methods: X-ray tube systems of two wide bore MDCT scanners (GE LightSpeed RT 4 and Philips Brilliance Big Bore) were modeled in the BEAMNRC/EGSNRC MC system. All the tube components were modeled from targets to bowtie filters. To validate our MC models, the authors measured half-value layers (HVL) using aluminum sheets and multifunctional radiation detectors and compared them to those obtained from MC simulations for 120 kVp beams. The authors also compared x-ray spectra obtained from MC simulation to the data provided by manufacturers. Additionally, lateral/axial beam profiles were measured in-air using radiochromic films and compared to the MC results. To understand the scatter effect, the authors also derived the scatter-to-primary energy fluence ratio (SPR) profiles and calculated the total SPR for each CT system with the CT dose index (CTDI) head and body phantoms using the BEAMNRC system. Results: The authors found that the HVL, x-ray spectrum and beam profiles of the MC simulations agreed well with the manufacturer-specified data within 1%-10% on average for both scanners. The total SPR were ranged from 7.8 to 13.7% for the head phantom and from 10.7 to 18.9% for the body phantom. Conclusions: The authors demonstrate the full MC simulations of two commercial MDCT simulators to characterize their x-ray beams. This study may be useful to establish a patient-specific dosimetry for the MDCT systems. (C) 2012 American Association of Physicists in Medicine. [DOI: 10.1118/1.3670377]
引用
收藏
页码:320 / 329
页数:10
相关论文
共 38 条
[1]  
[Anonymous], 2006, PIRS701 NRCC
[2]  
[Anonymous], PIRS794 NRCC
[3]  
[Anonymous], 1987, NBSIR873597 NIST
[4]   Monte Carlo simulation of x-ray spectra in diagnostic radiology and mammography using MCNP4C [J].
Ay, MR ;
Shahriari, M ;
Sarkar, S ;
Adib, M ;
Zaidi, H .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (21) :4897-4917
[5]   Development and validation of MCNP4C-based Monte Carlo simulator for fan- and cone-beam x-ray CT [J].
Ay, MR ;
Zaidi, H .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (20) :4863-4885
[6]   Off-axis x-ray spectra: A comparison of Monte Carlo simulated and computed x-ray spectra with measured spectra [J].
Bhat, M ;
Pattison, J ;
Bibbo, G ;
Caon, M .
MEDICAL PHYSICS, 1999, 26 (02) :303-309
[7]   COMPUTATION OF BREMSSTRAHLUNG X-RAY-SPECTRA AND COMPARISON WITH SPECTRA MEASURED WITH A GE(LI) DETECTOR [J].
BIRCH, R ;
MARSHALL, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1979, 24 (03) :505-517
[8]  
Birch R., 1979, CATALOGUE SPECTRAL D
[9]   Accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV [J].
Boone, JM ;
Seibert, JA .
MEDICAL PHYSICS, 1997, 24 (11) :1661-1670
[10]   Dose spread functions in computed tomography: A Monte Carlo study [J].
Boone, John M. .
MEDICAL PHYSICS, 2009, 36 (10) :4547-4554