Correlation of pre-operative MRI and intra-operative 3D ultrasound to measure brain tissue shift

被引:6
作者
Gobbi, DG [1 ]
Lee, BKH [1 ]
Peters, TM [1 ]
机构
[1] John P Robarts Res Inst, London, ON N6A 4K8, Canada
来源
MEDICAL IMAGING 2001: VISUALIZATION, DISPLAY, AND IMAGE-GUIDED PROCEDURES | 2001年 / 4319卷
关键词
3D ultrasound; brain shift; intraoperative ultrasound; image guided surgery; neurosurgery;
D O I
10.1117/12.428064
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
B-Mode ultrasound is often used during neurosurgery to provide intra-operative images of the brain through a craniotomy, but the use of 3D ultrasound during surgery is still in its infancy. We have developed a system that provides real-time freehand 3D ultrasound reconstruction at a reduced resolution. The reconstruction proceeds incrementally and the 3D image is overlayed, via a computer, on a pre-operative 3D MRI scan. This provides the operator with the necessary feedback to maintain a constant freehand sweep-rate, and also ensures that the sweep covers the desired anatomical volume. All of the ultrasound video frames are buffered, and a full-resolution, compounded reconstruction proceeds once the manual sweep is complete. We have also developed tools for manual tagging of homologous landmarks in the 3D MRI and 3D ultrasound volumes that use a piecewise cubic approximation of thin-plate spline interpolation to achieve interactive nonlinear registration and warping of the MRI volume to the ultrasound volume: Each time a homologous point-pair is identified by the user, the image of the warped MRI is updated on the computer screen after less than 0.5 s.
引用
收藏
页码:264 / 271
页数:8
相关论文
共 17 条
  • [1] BOOKSTEIN FL, 1999, BRAIN WARPING
  • [2] Bucholz RD, 1997, LECT NOTES COMPUT SC, V1205, P459, DOI 10.1007/BFb0029268
  • [3] Polyvinyl alcohol cryogel: An ideal phantom material for MR studies of arterial flow and elasticity
    Chu, KC
    Rutt, BK
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (02) : 314 - 319
  • [4] Intraoperative ultrasound for guidance and tissue shift correction in image-guided neurosurgery
    Comeau, RM
    Sadikot, AF
    Fenster, A
    Peters, TM
    [J]. MEDICAL PHYSICS, 2000, 27 (04) : 787 - 800
  • [5] COMEAU RM, SPIE, V3338, P474
  • [6] Preliminary clinical experience with intraoperative stereotactic ultrasound imaging
    Giorgi, C
    Casolino, DS
    [J]. STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1997, 68 (1-4) : 54 - 58
  • [7] Gobbi DG, 2000, LECT NOTES COMPUT SC, V1935, P106
  • [8] Gobbi DG, 1999, LECT NOTES COMPUT SC, V1679, P920
  • [9] SonoWand, an ultrasound-based neuronavigation system
    Gronningsaeter, A
    Kleven, A
    Ommedal, S
    Aarseth, TE
    Lie, T
    Lindseth, F
    Lango, T
    Unsgard, G
    [J]. NEUROSURGERY, 2000, 47 (06) : 1373 - 1379
  • [10] Development of a frameless and armless stereotactic neuronavigation system with ultrasonographic registration
    Hata, N
    Dohi, T
    Iseki, H
    Takakura, K
    [J]. NEUROSURGERY, 1997, 41 (03) : 608 - 613