Biomechanical evaluation of the less invasive stabilization system for the internal fixation of distal femur fractures

被引:185
作者
Marti, A
Fankhauser, C
Frenk, A
Cordey, J
Gasser, B
机构
[1] Dr HC Robert MarthyS Stiftung, Belltach, Switzerland
[2] Jacques Cordey Biomech, Filisur, Switzerland
关键词
angular stability; biomechanical study; internal fixation; monocortical and bicortical fixation; supracondylar femoral fractures;
D O I
10.1097/00005131-200109000-00004
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: Comparison between a Less Invasive Stabilization System (LISS) using monocortical screws with angular stability and two conventional plate systems Condylar Buttress Plate (CBP) and Dynamic Condylar Screw (DCS) for the treatment of distal femoral fractures with respect to biomechanical properties. Design: Biomechanical study using paired cadaver femurs. In Test Configuration I (distal test), a ten-millimeter gap at the diaphysis-metaphysis junction simulates a supracondylar femoral fracture. Test Configuration 2 (proximal test) has the same configuration, but the gap was cut in the isthmic region. Proximal and distal plate ends were fixed to corresponding cortical bone fragments in both tests. Optical displacement transducers served to quantify the system's ability to withstand a stepwise increased load. Reversible (deflection) and irreversible deformation (subsidence) of the bone-plate construct was investigated. Results: In Test Configuration 1, LISS showed less irreversible deformation in 72 percent of the left-right comparisons. No correlation between bone mineral density, cross-section area of bones and the measured response of the construct under load was found between pairs. In Test Configuration 2, 83 percent of the left-right comparisons showed less permanent deformation but a higher elastic deformation for LISS. Conclusions: These results suggest an enhanced ability to withstand high loads when using the monocortical screw fixation technique with angular stability. A higher elastic deformation of LISS compared with conventional plating systems in distal femoral fractures can be explained by the lower bending stiffness caused by different design and material properties.
引用
收藏
页码:482 / 487
页数:6
相关论文
共 18 条
[1]  
Davenport S R, 1988, J Orthop Trauma, V2, P146
[2]   RESULTS OF RIGID FIXATION IN 54 SUPRACONDYLAR FRACTURES OF THE FEMUR [J].
DELLATORRE, P ;
AGLIETTI, P ;
ALTISSIMI, M .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1980, 97 (03) :177-183
[3]   MECHANICS OF RETROGRADE NAIL VERSUS PLATE FIXATION FOR SUPRACONDYLAR FEMUR FRACTURES [J].
FIROOZBAKHSH, K ;
BEHZADI, K ;
DECOSTER, TA ;
MONEIM, MS ;
NARAGHI, FF .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1995, 9 (02) :152-157
[4]  
Haas NR, 1997, OP J, V13, P340
[5]  
HANSMANN H, 1986, KONGRESS, P134
[6]  
HOPF T, 1990, UNFALLCHIRURG, V93, P100
[7]   Internal fixation of supracondylar femoral fractures: Comparative biomechanical performance of the 95-degree blade plate and two retrograde nails [J].
Ito, K ;
Grass, R ;
Zwipp, H .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1998, 12 (04) :259-266
[8]   Distal femoral fixation: A biomechanical comparison of the standard condylar buttress plate, a locked buttress plate, and the 95-degree blade plate [J].
Koval, KJ ;
Hoehl, JJ ;
Kummer, FJ ;
Simon, JA .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1997, 11 (07) :521-524
[9]   Distal femoral fixation: A laboratory comparison of the 95 degrees plate, antegrade and retrograde inserted reamed intramedullary nails [J].
Koval, KJ ;
Kummer, FJ ;
Bharam, S ;
Chen, DY ;
Halder, S .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1996, 10 (06) :378-382
[10]   Concepts of minimally invasive plate osteosynthesis - Foreword [J].
Krettek, C .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1998, 29 :1-1