Comparison of the Xpert MTB/RIF Test with an IS6110-TaqMan Real-Time PCR Assay for Direct Detection of Mycobacterium tuberculosis in Respiratory and Nonrespiratory Specimens

被引:140
|
作者
Armand, Sylvie [1 ]
Vanhuls, Pascale [1 ]
Delcroix, Guy [1 ]
Courcol, Rene [1 ,2 ,3 ,4 ,5 ]
Lemaitre, Nadine [1 ,2 ,3 ,4 ,5 ]
机构
[1] CHU Lille, F-59000 Lille, France
[2] Univ Lille Nord France, F-59000 Lille, France
[3] UDSL, F-59000 Lille, France
[4] INSERM, U1019, F-59021 Lille, France
[5] Ctr Natl Rech Sci, UMR8204, F-59021 Lille, France
关键词
ACID AMPLIFICATION TESTS; DIAGNOSTIC-ACCURACY; PULMONARY; PERFORMANCE; COMPLEX; DNA;
D O I
10.1128/JCM.02157-10
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
The sensitivities of the Xpert MTB/RIF test and an in-house IS6110-based real-time PCR using TaqMan probes (IS6110-TaqMan assay) for the detection of Mycobacterium tuberculosis complex (MTBC) DNA were compared by use of 117 clinical specimens (97 culture positive and 20 culture negative for MTBC) that were frozen in sediment. The 97 clinical specimens included 60 respiratory and 37 nonrespiratory specimens distributed into 36 smear-positive and 61 smear-negative specimens. Among the 97 culture-positive specimens, 4 had rifampin-resistant isolates. Both methods were highly specific and exhibited excellent sensitivity (100%) with smear-positive specimens. The sensitivity of the Xpert MTB/RIF test with the whole smear-negative specimens was more reduced than that of the IS6110-TaqMan assay (48 versus 69%, P = 0.005). Both methods exhibited similar sensitivities with smear-negative respiratory specimens, but the Xpert MTB/RIF test had lower sensitivity with smear-negative nonrespiratory specimens than the IS6110-TaqMan assay (37 versus 71%, P = 0.013). Finally, the sensitivities of the Xpert MTB/RIF test and the IS6110-TaqMan assay were 79% and 84%, respectively, with respiratory specimens and 53% and 78%, respectively (P = 0.013), with nonrespiratory specimens. The Xpert MTB/RIF test correctly detected the rifampin resistance in smear-positive specimens but not in the one smear-negative specimen. The Xpert MTB/RIF test is a simple rapid method well adapted to a routine laboratory that appeared to be as sensitive as the IS6110-TaqMan assay with respiratory specimens but less sensitive with paucibacillary specimens, such as smear-negative nonrespiratory specimens.
引用
收藏
页码:1772 / 1776
页数:5
相关论文
共 32 条
  • [21] Detection of Mycobacteria, Mycobacterium avium Subspecies, and Mycobacterium tuberculosis Complex by a Novel Tetraplex Real-Time PCR Assay
    Sevilla, Iker A.
    Molina, Elena
    Elguezabal, Natalia
    Perez, Valentin
    Garrido, Joseba M.
    Justea, Ramn A.
    JOURNAL OF CLINICAL MICROBIOLOGY, 2015, 53 (03) : 930 - 940
  • [22] Multi-Fluorescence Real-Time PCR Assay for Detection of RIF and INH Resistance of M-tuberculosis
    Peng, Jingfu
    Yu, Xiaoli
    Cui, Zhenling
    Xue, Wenfei
    Luo, Ziyi
    Wen, Zilu
    Liu, Minghua
    Jiang, Danqing
    Zheng, Heping
    Wu, Hai
    Zhang, Shulin
    Li, Yao
    FRONTIERS IN MICROBIOLOGY, 2016, 7
  • [23] Audit and improve! Evaluation of a real-time probe-based PCR assay with internal control for the direct detection of Mycobacterium tuberculosis complex
    Inoue, M.
    Tang, W. Y.
    Wee, S. Y.
    Barkham, T.
    EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2011, 30 (01) : 131 - 135
  • [24] Performance evaluation of cobas HBV real-time PCR assay on Roche cobas 4800 System in comparison with COBAS AmpliPrep/COBAS TaqMan HBV Test
    Kim, Hanah
    Hur, Mina
    Bae, Eunsin
    Lee, Kyung-A
    Lee, Woo-In
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2018, 56 (07) : 1133 - 1139
  • [25] Paradigm for diagnosing mycobacterial disease: direct detection and differentiation of Mycobacterium tuberculosis complex and non-tuberculous mycobacteria in clinical specimens using multiplex real-time PCR
    Kim, Jeong-Uk
    Ryu, Dae-Shick
    Cha, Choong-Hwan
    Park, Seon-Hee
    JOURNAL OF CLINICAL PATHOLOGY, 2018, 71 (09) : 774 - 780
  • [26] Sensitivity and specificity of Cobas TaqMan MTB real-time polymerase chain reaction for culture-proven Mycobacterium tuberculosis: meta-analysis of 26999 specimens from 17 Studies
    Horita, Nobuyuki
    Yamamoto, Masaki
    Sato, Takashi
    Tsukahara, Toshinori
    Nagakura, Hideyuki
    Tashiro, Ken
    Shibata, Yuji
    Watanabe, Hiroki
    Nagai, Kenjiro
    Nakashima, Kentaro
    Ushio, Ryota
    Ikeda, Misako
    Sakamaki, Kentaro
    Yoshiyama, Takashi
    Kaneko, Takeshi
    SCIENTIFIC REPORTS, 2015, 5
  • [27] Detection of Mycobacterium avium subsp paratuberculosis in bovine faecal samples by Quantitative Real-Time Polymerase Chain Reaction (qRT-PCR) TaqMan Assay
    Rasool, S.
    Narang, D.
    Chandra, M.
    Sharma, N. S.
    Singh, Tejinder S.
    TROPICAL BIOMEDICINE, 2016, 33 (02) : 327 - 334
  • [28] Comparison of Single-Copy and Multicopy Real-Time PCR Targets for Detection of Mycobacterium tuberculosis in Paraffin-Embedded Tissue
    Luo, Robert F.
    Scahill, Michael D.
    Banaei, Niaz
    JOURNAL OF CLINICAL MICROBIOLOGY, 2010, 48 (07) : 2569 - 2570
  • [29] Validation of a Multiplex Real-Time PCR Assay for Detection of Mycobacterium spp., Mycobacterium tuberculosis Complex, and Mycobacterium avium Complex Directly from Clinical Samples by Use of the BD Max Open System
    Rocchetti, Talita T.
    Silbert, Suzane
    Gostnell, Alicia
    Kubasek, Carly
    Widen, Raymond
    JOURNAL OF CLINICAL MICROBIOLOGY, 2016, 54 (06) : 1644 - 1647
  • [30] Validation of the MycAssay Pneumocystis Kit for Detection of Pneumocystis jirovecii in Bronchoalveolar Lavage Specimens by Comparison to a Laboratory Standard of Direct Immunofluorescence Microscopy, Real-Time PCR, or Conventional PCR
    McTaggart, Lisa R.
    Wengenack, Nancy L.
    Richardson, Susan E.
    JOURNAL OF CLINICAL MICROBIOLOGY, 2012, 50 (06) : 1856 - 1859