Signal detection and location-dependent noise in cone-beam computed tomography using the spatial definition of the Hotelling SNR

被引:20
作者
Brunner, Claudia C. [1 ,2 ]
Abboud, Samir F. [1 ,3 ,4 ]
Hoeschen, Christoph [2 ]
Kyprianou, Iacovos S. [1 ]
机构
[1] US FDA, Ctr Devices & Radiol Hlth, Silver Spring, MD 20993 USA
[2] Helmholtz Zentrum Munchen, Res Unit Med Radiat Phys & Diagnost, D-85764 Neuherberg, Germany
[3] Univ Maryland, Sch Med, Baltimore, MD 21201 USA
[4] Univ Maryland, Dept Bioengn, College Pk, MD 20742 USA
关键词
cone-beam CT; image quality; discrete object transfer matrix; covariance matrix; MTF; NPS; SNR; detectability; CT SCANNER; SYSTEM; POWER; RECONSTRUCTION; OBSERVER;
D O I
10.1118/1.4718572
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Quality assurance in computed tomography (CT) is commonly performed with the Fourier-based modulation transfer function (MTF) and the noise variance, while more recently the noise power spectrum (NPS) has increased in popularity. The Fourier-based methods make assumptions such as shift-invariance and cyclostationarity. These assumptions are violated in real clinical systems and consequently are expected to result in systematic errors. A spatial approach, based on the object transfer matrix (T) and the covariance matrix (K) theory, does not require these assumptions and can provide a more general description of the imaging system. In this paper, the authors present an experimental methodology and data treatment for quality assessment of a lab cone-beam CT system by comparing the spatial with the Fourier approach in 2D reconstructed slices. Methods: In order to have control over all experimental parameters and image reconstruction, a bench-top flat-panel-based cone-beam CT scanner and a cylindrical water-filled poly(methyl methacrylate) (PMMA) phantom were used for the noise measurements. An aluminum foil inserted in the water phantom enabled the estimation of the line response function (LRF) with a limited number of assumptions. The authors evaluated the spatial blur, the noise and the signal-to-noise ratio (SNR) using the spatial approach as well as the Fourier-based approach. In order to evaluate the degree of noise nonstationarity of their cone-beam CT system, the authors evaluated both the local and global CT noise properties and compared them using both approaches. Results: For the laboratory cone-beam CT, the location-dependent noise evaluation showed that in addition to the noise variance, the NPS and covariance eigenvector symmetry depend on the location in the image. The estimated signal transfer was similar for both approaches. Unlike the Fourier approach which uses the same exponential wave function basis for both MTF and NPS, the eigenvectors of T and K were significantly different. Conclusions: By using the eigenvectors of the noise and object transfer to characterize the system, the spatial approach provides additional information to the Fourier approach and is therefore an important tool for a thorough understanding of a CT system. [http://dx.doi.org/10.1118/1.4718572]
引用
收藏
页码:3214 / 3228
页数:15
相关论文
共 34 条
[21]  
Kyprianou I S, 2009, P SOC PHOTO-OPT INS, V7258
[22]  
Kyprianou I. S., 2007, P SOC PHOTO-OPT INS, V6510
[23]   Singular value description of a digital radiographic detector: Theory and measurements [J].
Kyprianou, Iacovos S. ;
Badano, Aldo ;
Gallas, Brandon D. ;
Myers, Kyle J. .
MEDICAL PHYSICS, 2008, 35 (10) :4744-4756
[24]  
Kyprianou IS, 2006, I S BIOMED IMAGING, P1248
[25]   Study of the generalized MTF and DQE for a new microangiographic system. [J].
Kyprianou, IS ;
Rudin, S ;
Bednarek, DR ;
Hoffmann, KR .
MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 :349-360
[26]   A practical method for measuring the H matrix of digital x-ray and cone beam CT imaging systems [J].
Kyprianou, Lacovos S. ;
Badano, Aldo ;
D Gallas, Brandon ;
Park, Subok ;
Myers, Kyle J. .
MEDICAL IMAGING 2006: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2006, 6142
[27]   Generalizing the MTF and DQE to include x-ray scatter and focal spot unsharpness: Application to a new microangiographic system [J].
Kyprianou, LS ;
Rudin, S ;
Bednarek, DR ;
Hoffmann, KR .
MEDICAL PHYSICS, 2005, 32 (02) :613-626
[28]   Analysis of image noise in 3D cone-beam CT: Spatial and Fourier domain approaches under conditions of varying stationarity [J].
Pineda, Angel R. ;
Siewerdsen, Jeffrey H. ;
Tward, Daniel J. .
MEDICAL IMAGING 2008: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2008, 6913
[29]   Beyond noise power in 3D computed tomography: The local NPS and off-diagonal elements of the Fourier domain covariance matrix [J].
Pineda, Angel R. ;
Tward, Daniel J. ;
Gonzalez, Antonio ;
Siewerdsen, Jeffrey H. .
MEDICAL PHYSICS, 2012, 39 (06) :3240-3252
[30]   Figures of merit for detectors in digital radiography. I. Flat background and deterministic blurring [J].
Pineda, AR ;
Barrett, HH .
MEDICAL PHYSICS, 2004, 31 (02) :348-358