Flux or speed? Examining speckle contrast imaging of vascular flows

被引:76
作者
Kazmi, S. M. Shams [1 ]
Faraji, Ehssan [1 ]
Davis, Mitchell A. [1 ]
Huang, Yu-Yen [1 ]
Zhang, Xiaojing J. [1 ]
Dunn, Andrew K. [1 ]
机构
[1] Univ Texas Austin, Dept Biomed Engn, Austin, TX 78712 USA
基金
美国国家卫生研究院;
关键词
CEREBRAL-BLOOD-FLOW; DIFFUSE CORRELATION SPECTROSCOPY; MULTI-EXPOSURE; MICROCIRCULATION; FLOWMETRY; BRAIN; MICROSCOPY; SCATTERING; VESSELS; SYSTEM;
D O I
10.1364/BOE.6.002588
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Speckle contrast imaging enables rapid mapping of relative blood flow distributions using camera detection of back-scattered laser light. However, speckle derived flow measures deviate from direct measurements of erythrocyte speeds by 47 +/- 15% (n = 13 mice) in vessels of various calibers. Alternatively, deviations with estimates of volumetric flux are on average 91 +/- 43%. We highlight and attempt to alleviate this discrepancy by accounting for the effects of multiple dynamic scattering with speckle imaging of microfluidic channels of varying sizes and then with red blood cell (RBC) tracking correlated speckle imaging of vascular flows in the cerebral cortex. By revisiting the governing dynamic light scattering models, we test the ability to predict the degree of multiple dynamic scattering across vessels in order to correct for the observed discrepancies between relative RBC speeds and multi-exposure speckle imaging estimates of inverse correlation times. The analysis reveals that traditional speckle contrast imagery of vascular flows is neither a measure of volumetric flux nor particle speed, but rather the product of speed and vessel diameter. The corrected speckle estimates of the relative RBC speeds have an average 10 +/- 3% deviation in vivo with those obtained from RBC tracking. (C)2015 Optical Society of America
引用
收藏
页码:2588 / 2608
页数:21
相关论文
共 49 条
[1]   Laser speckle flowmetry for the study of cerebrovascular physiology in normal and ischemic mouse cortex [J].
Ayata, C ;
Dunn, AK ;
Gursoy-Özdemir, Y ;
Huang, ZH ;
Boas, DA ;
Moskowitz, MA .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 2004, 24 (07) :744-755
[2]   Speckle-visibility spectroscopy: A tool to study time-varying dynamics [J].
Bandyopadhyay, R ;
Gittings, AS ;
Suh, SS ;
Dixon, PK ;
Durian, DJ .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2005, 76 (09)
[3]   SCATTERING AND IMAGING WITH DIFFUSING TEMPORAL FIELD CORRELATIONS [J].
BOAS, DA ;
CAMPBELL, LE ;
YODH, AG .
PHYSICAL REVIEW LETTERS, 1995, 75 (09) :1855-1858
[4]   Spatially varying dynamical properties of turbid media probed with diffusing temporal light correlation [J].
Boas, DA ;
Yodh, AG .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1997, 14 (01) :192-215
[5]   Laser speckle contrast imaging in biomedical optics [J].
Boas, David A. ;
Dunn, Andrew K. .
JOURNAL OF BIOMEDICAL OPTICS, 2010, 15 (01)
[6]   MODEL FOR LASER DOPPLER MEASUREMENTS OF BLOOD-FLOW IN TISSUE [J].
BONNER, R ;
NOSSAL, R .
APPLIED OPTICS, 1981, 20 (12) :2097-2107
[7]   Laser speckle contrast imaging: theoretical and practical limitations [J].
Briers, David ;
Duncan, Donald D. ;
Hirst, Evan ;
Kirkpatrick, Sean J. ;
Larsson, Marcus ;
Steenbergen, Wiendelt ;
Stromberg, Tomas ;
Thompson, Oliver B. .
JOURNAL OF BIOMEDICAL OPTICS, 2013, 18 (06)
[8]   Due to intravascular multiple sequential scattering, Diffuse Correlation Spectroscopy of tissue primarily measures relative red blood cell motion within vessels [J].
Carp, Stefan A. ;
Roche-Labarbe, Nadege ;
Franceschini, Maria-Angela ;
Srinivasan, Vivek J. ;
Sakadzic, Sava ;
Boas, David A. .
BIOMEDICAL OPTICS EXPRESS, 2011, 2 (07) :2047-2054
[9]   Simplified laser-speckle-imaging analysis method and its application to retinal blood flow imaging [J].
Cheng, Haiying ;
Duong, Timothy Q. .
OPTICS LETTERS, 2007, 32 (15) :2188-2190
[10]   Laser speckle imaging for monitoring blood flow dynamics in the in vivo rodent dorsal skin fold model [J].
Choi, B ;
Kang, NM ;
Nelson, JS .
MICROVASCULAR RESEARCH, 2004, 68 (02) :143-146