Evaluation of the MB/BacT system compared with Middlebrook 7H11 and Lowenstein-Jensen media for detection and recovery of mycobacteria from clinical specimens

被引:17
作者
Gil-Setas, A [1 ]
Torroba, L [1 ]
Fernandez, JL [1 ]
Martinez-Artola, V [1 ]
Olite, J [1 ]
机构
[1] Hosp Virgen del Camino, Cent Lab, Pamplona 31008, Spain
关键词
Lowenstein-Jensen; MB/BacT; microcolony method; Middlebrook; 7H11; agar; mycobacteria; tuberculosis;
D O I
10.1111/j.1198-743X.2004.00733.x
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
MB/BacT vials ( an automated system), Middlebrook 7H11 agar plates with the microcolony method for growth detection, and conventional Lowenstein-Jensen egg-based medium were assessed for their rates of recovery, time to detection and ease in aiding the identification of mycobacterial isolates. Of the 2101 consecutive, non-selected specimens for tuberculosis analysed in the laboratory, 158 (7.5%) yielded growth of 159 mycobacteria on at least one of the three media, comprising 111 Mycobacterium tuberculosis and 48 non-tuberculous mycobacteria. Of the 111 specimens positive for M. tuberculosis, 100 isolates were recovered from MB/BacT vials, 99 from M7H11 agar plates, and 86 from Lowenstein-Jensen tubes. The combination of MB/BacT and M7H11 recovered 110 isolates of M. tuberculosis, 107 from M7H11 and Lowenstein-Jensen, and 105 from MB/BacT plus Lowenstein-Jensen. The average time to detection of M. tuberculosis was 11 days with M7H11 using the microcolony method, 16 days with MB/BacT, and 19.5 days with the conventional Lowenstein-Jensen method. The MB/BacT vials recovered greater numbers of non-tuberculous mycobacteria than either of the other two media, but these were considered mostly to be saprophytes. It was concluded that a combination of media was better than a single medium. None of the media combinations showed statistical differences. Use of M7H11 with the microcolony method, which shortens the detection time for mycobacteria and facilitates the detection of mixed cultures, together with the Lowenstein-Jensen tube, was a useful combination.
引用
收藏
页码:224 / 228
页数:5
相关论文
共 20 条
  • [1] Comparison of the MB/BacT and BACTEC 460 TB systems for recovery of mycobacteria from various clinical specimens
    Brunello, F
    Favari, F
    Fontana, R
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1999, 37 (04) : 1206 - 1209
  • [2] Centers for Disease Control and Prevention (CDC), 1996, MMWR Morb Mortal Wkly Rep, V45, P950
  • [3] Gindler J. S., 1995, Morbidity and Mortality Weekly Report, V44, P1
  • [4] Mycobacteriology laboratory
    Heifets, L
    [J]. CLINICS IN CHEST MEDICINE, 1997, 18 (01) : 35 - +
  • [5] RAPID DETECTION OF TUBERCULOUS AND NONTUBERCULOUS MYCOBACTERIA BY MICROSCOPIC OBSERVATION OF GROWTH ON MIDDLEBROOK 7H11 AGAR
    IDIGORAS, P
    PEREZTRALLERO, E
    ALCORTA, M
    GUTIERREZ, C
    MUNOZBAROJA, I
    [J]. EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 1995, 14 (01) : 6 - 10
  • [6] ISENBERG HD, 2003, CLIN MICROBIOLOGY PR, V1
  • [7] Kent P.T., 1985, PUBLIC HLTH MYCOBACT
  • [8] Metchock BG., 1999, Manual of clinical microbiologyx. Dans: Mycobacterium, V7th ed, P399
  • [9] Comparison of MB/BacT and BACTEC 460 TB systems for recovery of mycobacteria in a routine diagnostic laboratory
    Roggenkamp, A
    Hornef, MW
    Masch, A
    Aigner, B
    Autenrieth, IB
    Heesemann, J
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1999, 37 (11) : 3711 - 3712
  • [10] Evaluation of the MB/BacT system and comparison to the BACTEC 460 system and solid media for isolation of mycobacteria from clinical specimens
    Rohner, P
    Ninet, B
    Metral, C
    Emler, S
    Auckenthaler, R