YAG:Ce-Phosphor Scintillators for Optical Fiber Radiation Sensors With High Temporal Resolution

被引:12
作者
Chen, Lingxia [1 ]
Chen, Shuilin [2 ,3 ]
Gillespie, Sean [4 ]
Woulfe, Peter [4 ]
Jiang, Benxue [2 ]
O'Keeffe, Sinead [1 ]
Lewis, Elfed [1 ]
机构
[1] Univ Limerick, Opt Fibre Sensors Res Ctr, Elect & Comp Engn Dept, Limerick V94 T9PX, Ireland
[2] Chinese Acad Sci, Shanghai Inst Opt & Fine Mech, Key Lab Mat High Power Laser, Shanghai 201800, Peoples R China
[3] Univ Chinese Acad Sci, Beijing 100049, Peoples R China
[4] Galway Clin, Dept Radiotherapy Phys, Galway H91 HHT0, Ireland
基金
爱尔兰科学基金会;
关键词
Scintillation detectors; optical fiber devices; fluorescence; dosimetry; radiotherapy; X-rays; yttrium compounds; PHYSICAL-CHARACTERISTICS; DOSIMETRY; RADIOTHERAPY; PHOTON; CE;
D O I
10.1109/LPT.2018.2862152
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A miniature optical fiber real-time X-ray radiation sensor based on plastic optical fiber coupled with a fast decay time trivalent cerium-activated yttrium aluminum garnet Y3Al5O12 (YAG:Ce) scintillator material is reported. The YAG material was tested in powder and ceramic form. Three separate sensors were irradiated three times on a Siemens Artiste linear accelerator (linac). The measured light output of the YAG:Ce ceramic to the detector at the distal end of the fiber is 4.52 times greater than the YAG:Ce powder, and its output intensity detected by photon counter is further increased 2.52 times using a two-layer highly reflective Teflon tape. The sensor made with YAG:Ce powder shows excellent repeatability, being a maximum of 0.36 %, but the value using YAG:Ce ceramic exhibits orientation sensitivity when located in orthogonal positions relative to the X-ray beam. To the best of our knowledge, the work described in this letter is the first instance of the direct high temporal resolution detection of the individual (circa 3.9 ms) output of X-ray light pulses of clinical linacs. The experimental results of pulsewidths are in good agreement with the manufacturer's stated value of 3.45 +/- 0.21 mu s.
引用
收藏
页码:1653 / 1656
页数:4
相关论文
共 14 条
[1]   IONIZATION-CHAMBER DOSIMETRY FOR PHOTON AND ELECTRON-BEAMS - THEORETICAL CONSIDERATIONS [J].
ALMOND, PR ;
SVENSSON, H .
ACTA RADIOLOGICA-THERAPY PHYSICS BIOLOGY, 1977, 16 (02) :177-186
[2]   On the Monte Carlo simulation of small-field micro-diamond detectors for megavoltage photon dosimetry [J].
Andreo, Pedro ;
Palmans, Hugo ;
Marteinsdottir, Maria ;
Benmakhlouf, Hamza ;
Carlsson-Tedgren, Asa .
PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (01) :L1-L10
[3]   Real-time optical-fibre luminescence dosimetry for radiotherapy:: physical characteristics and applications in photon beams [J].
Aznar, MC ;
Andersen, CE ;
Botter-Jensen, L ;
Bäck, SÅJ ;
Mattsson, S ;
Kjær-Kristoffersen, F ;
Medin, J .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (09) :1655-1669
[4]   WATER-EQUIVALENT PLASTIC SCINTILLATION DETECTORS FOR HIGH-ENERGY BEAM DOSIMETRY .1. PHYSICAL CHARACTERISTICS AND THEORETICAL CONSIDERATIONS [J].
BEDDAR, AS ;
MACKIE, TR ;
ATTIX, FH .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (10) :1883-1900
[5]   Comparison of three high-resolution digitizers for radiochromic film dosimetry [J].
Gluckman, GR ;
Reinstein, LE .
MEDICAL PHYSICS, 2002, 29 (08) :1839-1846
[6]   Quality Assurance of Radiotherapy in Cancer Treatment: Toward Improvement of Patient Safety and Quality of Care [J].
Ishikura, Satoshi .
JAPANESE JOURNAL OF CLINICAL ONCOLOGY, 2008, 38 (11) :723-729
[7]   Optical Reflectance Measurements for Commonly Used Reflectors [J].
Janecek, Martin ;
Moses, William W. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2008, 55 (04) :2432-2437
[8]   Applications of thermoluminescence dosimetry in medicine [J].
Kron, T .
RADIATION PROTECTION DOSIMETRY, 1999, 85 (1-4) :333-340
[9]  
KVAPIL J, 1981, J CRYST GROWTH, V52, P542, DOI 10.1016/0022-0248(81)90336-5
[10]   Portal imaging [J].
Langmack, KA .
BRITISH JOURNAL OF RADIOLOGY, 2001, 74 (885) :789-804