Assembly force and taper angle difference influence the relative motion at the stem-neck interface of bi-modular hip prostheses

被引:31
作者
Haschke, Henning [1 ]
Jauch-Matt, Sabrina Y. [1 ,2 ]
Sellenschloh, Kay [1 ]
Huber, Gerd [1 ]
Morlock, Michael M. [1 ]
机构
[1] Hamburg Univ Technol TUHH, Inst Biomech, Denickestr 15, D-21073 Hamburg, Germany
[2] Univ Bath, Dept Mech Engn, Ctr Orthopaed Biomech, Bath BA2 7AY, Avon, England
关键词
Modular prosthesis; relative motion; assembly force; stem-neck interface; taper angle difference; ON-METAL BEARINGS; FRETTING CORROSION; FEMORAL-HEAD; MORSE TAPER; ARTHROPLASTY; WEAR; IMPLANTS; MICROMOTION; PRODUCTS; JUNCTION;
D O I
10.1177/0954411916648717
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Bi-modular hip arthroplasty prostheses allow adaptation to the individual patient anatomy and the combination of different materials but introduce an additional interface, which was related lately to current clinical issues. Relative motion at the additional taper interface might increase the overall risk of fretting, corrosion, metallic debris and early failure. The aim of this study was to investigate whether the assembly force influences the relative motion and seating behaviour at the stem-neck interface of a bi-modular hip prosthesis (Metha((R)); Aesculap AG, Tuttlingen, Germany) and whether this relation is influenced by the taper angle difference between male and female taper angles. Neck adapters made of titanium (Ti6Al4V) and CoCr (CoCr29Mo) were assembled with a titanium stem using varying assembly forces and mechanically loaded. A contactless eddy current measurement system was used to record the relative motion between prosthesis stem and neck adapter. Higher relative motion was observed for Ti neck adapters compared to the CoCr ones (p<0.001). Higher assembly forces caused increased seating distances (p<0.001) and led to significantly reduced relative motion (p=0.019). Independent of neck material type, prostheses with larger taper angle difference between male and female taper angles exhibited decreased relative motion (p<0.001). Surgeons should carefully use assembly forces above 4kN to decrease the amount of relative motion within the taper interface. Maximum assembly forces, however, should be limited to prevent periprosthetic fractures. Manufacturers should optimize taper angle differences to increase the resistance against relative motion.
引用
收藏
页码:690 / 699
页数:10
相关论文
共 54 条
[21]  
Goldberg JR, 2002, CLIN ORTHOP RELAT R, P149
[22]   Modular titanium alloy neck adapter failures in hip replacement - failure mode analysis and influence of implant material [J].
Grupp, Thomas M. ;
Weik, Thomas ;
Bloemer, Wilhelm ;
Knaebel, Hanns-Peter .
BMC MUSCULOSKELETAL DISORDERS, 2010, 11
[23]  
Hallab NJ, 2009, BULL HOSP JT DIS, V67, P182
[24]   The relationship between the angle of version and rate of wear of retrieved metal-on-metal resurfacings A PROSPECTIVE, CT-BASED STUDY [J].
Hart, A. J. ;
Ilo, K. ;
Underwood, R. ;
Cann, P. ;
Henckel, J. ;
Lewis, A. ;
Cobb, J. ;
Skinner, J. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (03) :315-320
[25]   Impact magnitudes applied by surgeons and their importance when applying the femoral head onto the Morse taper for total hip arthroplasty [J].
Heiney, Jake P. ;
Battula, Suneel ;
Vrabec, Gregory A. ;
Parikh, Anand ;
Blice, Rebecca ;
Schoenfeld, Andrew J. ;
Njus, Glenn O. .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2009, 129 (06) :793-796
[26]   Is Increased Modularity Associated With Increased Fretting and Corrosion Damage in Metal-On-Metal Total Hip Arthroplasty Devices? A Retrieval Study [J].
Higgs, Genymphas B. ;
Hanzlik, Josa A. ;
MacDonald, Daniel W. ;
Gilbert, Jeremy L. ;
Rimnac, Clare M. ;
Kurtz, Steven M. .
JOURNAL OF ARTHROPLASTY, 2013, 28 (08) :2-6
[27]   Presence of corrosion products and hypersensitivity-associated reactions in periprosthetic tissue after aseptic loosening of total hip replacements with metal bearing surfaces [J].
Huber, Monika ;
Reinisch, Georg ;
Trettenhahn, Guenter ;
Zweymueller, Karl ;
Lintner, Felix .
ACTA BIOMATERIALIA, 2009, 5 (01) :172-180
[28]   Head-Neck Taper Corrosion in Hip Arthroplasty [J].
Hussenbocus, S. ;
Kosuge, D. ;
Solomon, L. B. ;
Howie, D. W. ;
Oskouei, R. H. .
BIOMED RESEARCH INTERNATIONAL, 2015, 2015
[29]  
JACOBS JJ, 1995, CLIN ORTHOP RELAT R, P94
[30]   Low torque levels can initiate a removal of the passivation layer and cause fretting in modular hip stems [J].
Jauch, S. Y. ;
Coles, L. G. ;
Ng, L. V. ;
Miles, A. W. ;
Gill, H. S. .
MEDICAL ENGINEERING & PHYSICS, 2014, 36 (09) :1140-1146