Assembly force and taper angle difference influence the relative motion at the stem-neck interface of bi-modular hip prostheses

被引:31
作者
Haschke, Henning [1 ]
Jauch-Matt, Sabrina Y. [1 ,2 ]
Sellenschloh, Kay [1 ]
Huber, Gerd [1 ]
Morlock, Michael M. [1 ]
机构
[1] Hamburg Univ Technol TUHH, Inst Biomech, Denickestr 15, D-21073 Hamburg, Germany
[2] Univ Bath, Dept Mech Engn, Ctr Orthopaed Biomech, Bath BA2 7AY, Avon, England
关键词
Modular prosthesis; relative motion; assembly force; stem-neck interface; taper angle difference; ON-METAL BEARINGS; FRETTING CORROSION; FEMORAL-HEAD; MORSE TAPER; ARTHROPLASTY; WEAR; IMPLANTS; MICROMOTION; PRODUCTS; JUNCTION;
D O I
10.1177/0954411916648717
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Bi-modular hip arthroplasty prostheses allow adaptation to the individual patient anatomy and the combination of different materials but introduce an additional interface, which was related lately to current clinical issues. Relative motion at the additional taper interface might increase the overall risk of fretting, corrosion, metallic debris and early failure. The aim of this study was to investigate whether the assembly force influences the relative motion and seating behaviour at the stem-neck interface of a bi-modular hip prosthesis (Metha((R)); Aesculap AG, Tuttlingen, Germany) and whether this relation is influenced by the taper angle difference between male and female taper angles. Neck adapters made of titanium (Ti6Al4V) and CoCr (CoCr29Mo) were assembled with a titanium stem using varying assembly forces and mechanically loaded. A contactless eddy current measurement system was used to record the relative motion between prosthesis stem and neck adapter. Higher relative motion was observed for Ti neck adapters compared to the CoCr ones (p<0.001). Higher assembly forces caused increased seating distances (p<0.001) and led to significantly reduced relative motion (p=0.019). Independent of neck material type, prostheses with larger taper angle difference between male and female taper angles exhibited decreased relative motion (p<0.001). Surgeons should carefully use assembly forces above 4kN to decrease the amount of relative motion within the taper interface. Maximum assembly forces, however, should be limited to prevent periprosthetic fractures. Manufacturers should optimize taper angle differences to increase the resistance against relative motion.
引用
收藏
页码:690 / 699
页数:10
相关论文
共 54 条
[1]  
Abdullah K, 2010, MODELING SIMULATION, P121
[2]   Corrosion-Induced Fracture of a Double-Modular Hip Prosthesis A Case Report [J].
Atwood, Sara A. ;
Patten, Eli W. ;
Bozic, Kevin J. ;
Pruitt, Lisa A. ;
Ries, Michael D. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2010, 92A (06) :1522-1525
[3]  
Australian Orthopaedic Association National Joint Replacement Registry, 2013, AUSTR ORTH ASS 2013
[4]   The influence of contact conditions and micromotions on the fretting behavior of modular titanium alloy taper connections [J].
Baxmann, M. ;
Jauch, S. Y. ;
Schilling, C. ;
Bloemer, W. ;
Grupp, T. M. ;
Morlock, M. M. .
MEDICAL ENGINEERING & PHYSICS, 2013, 35 (05) :676-683
[5]   Hip contact forces and gait patterns from routine activities [J].
Bergmann, G ;
Deuretzbacher, G ;
Heller, M ;
Graichen, F ;
Rohlmann, A ;
Strauss, J ;
Duda, GN .
JOURNAL OF BIOMECHANICS, 2001, 34 (07) :859-871
[6]   Realistic loads for testing hip implants [J].
Bergmann, G. ;
Graichen, F. ;
Rohlmann, A. ;
Bender, A. ;
Heinlein, B. ;
Duda, G. N. ;
Heller, M. O. ;
Morlock, M. M. .
BIO-MEDICAL MATERIALS AND ENGINEERING, 2010, 20 (02) :65-75
[7]   Wear patterns of taper connections in retrieved large diameter metal-on-metal bearings [J].
Bishop, Nicholas ;
Witt, Florian ;
Pourzal, Robin ;
Fischer, Alfons ;
Ruetschi, Marcel ;
Michel, Markus ;
Morlock, Michael .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2013, 31 (07) :1116-1122
[8]   FRETTING CORROSION ACCELERATES CREVICE CORROSION OF MODULAR HIP TAPERS [J].
BROWN, SA ;
FLEMMING, CAC ;
KAWALEC, JS ;
PLACKO, HE ;
VASSAUX, C ;
MERRITT, K ;
PAYER, JH ;
KRAAY, MJ .
JOURNAL OF APPLIED BIOMATERIALS, 1995, 6 (01) :19-26
[9]   Friction in Total Hip Joint Prosthesis Measured In Vivo during Walking [J].
Damm, Philipp ;
Dymke, Joern ;
Ackermann, Robert ;
Bender, Alwina ;
Graichen, Friedmar ;
Halder, Andreas ;
Beier, Alexander ;
Bergmann, Georg .
PLOS ONE, 2013, 8 (11)
[10]   Fretting cracking behaviour on pre-stressed aluminium alloy specimens [J].
Degat, PR ;
Zhou, ZR ;
Vincent, L .
TRIBOLOGY INTERNATIONAL, 1997, 30 (03) :215-223