Detection of heart rate and rhythm with a smartphone-based electrocardiograph versus a reference standard electrocardiograph in dogs and cats

被引:31
作者
Kraus, Marc S. [1 ,2 ]
Gelzer, Anna R. [1 ,2 ]
Rishniw, Mark [1 ]
机构
[1] Cornell Univ, Coll Vet Med, Dept Clin Sci, Ithaca, NY 14853 USA
[2] Univ Penn, Sch Vet Med, Dept Clin Studies Philadelphia, Philadelphia, PA 19104 USA
来源
JAVMA-JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION | 2016年 / 249卷 / 02期
关键词
ATRIAL-FIBRILLATION; ECG;
D O I
10.2460/javma.249.2.189
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
OBJECTIVE To evaluate the diagnostic utility of ECGs acquired with a smartphone-based device, compared with reference 6-lead ECGs, for identification of heart rate and rhythm in dogs and cats. DESIGN Prospective study. ANIMALS 51 client-owned dogs and 27 client-owned cats. PROCEDURES Patients examined by a small animal referral cardiology service between April 2012 and January 2013 were enrolled consecutively. In each patient, a 30-second ECG was simultaneously acquired with a smartphone-based device (a bipolar, single-lead recorder coupled to a smartphone with an ECG application) and a standard 6-lead ECG machine. Recordings were evaluated by 3 board-certified cardiologists, and intra-and interobserver agreement were evaluated for both rhythm diagnosis and QRS polarity identification. RESULTS Values for instantaneous and mean heart rates for the smartphone-acquired and reference ECGs were within 1 beat of each other when mean heart rates were calculated. Intraobserver agreement for rhythm assessment was very high, with maximum disagreement for any observer for only 2 of 51 dogs and only 4 of 27 cats. There was minimal disagreement in the polarity of depolarization between the smartphone-acquired and reference ECGs in dogs but frequent disagreement in cats. Interobserver agreement for smartphone-acquired ECGs was similar to that for reference ECGs, with all 3 observers agreeing on the rhythm analysis and minimal disagreement on polarity. CONCLUSIONS AND CLINICAL RELEVANCE Results suggested that ECGs acquired with the smartphone-based device accurately identified heart rate and rhythm in dogs and cats. Thus, the device may allow veterinarians to evaluate and manage cardiac arrhythmias relatively inexpensively at the cage side and could also allow clinicians to rapidly share information via email for further consultation, potentially enhancing patient care.
引用
收藏
页码:189 / 194
页数:6
相关论文
共 8 条
[1]   QTC intervals can be assessed with the AliveCor heart monitor in patients on dofetilide for atrial fibrillation [J].
Chung, Eugene H. ;
Guise, Kimberly D. .
JOURNAL OF ELECTROCARDIOLOGY, 2015, 48 (01) :8-9
[2]  
Einthoven W., 1901, ARCH NEERL SC EXNAT, V6, P625
[3]   Wireless Smartphone ECG Enables Large-Scale Screening in Diverse Populations [J].
Haberman, Zachary C. ;
Jahn, Ryan T. ;
Bose, Rupan ;
Tun, Han ;
Shinbane, Jerold S. ;
Doshi, Rahul N. ;
Chang, Philip M. ;
Saxon, Leslie A. .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2015, 26 (05) :520-526
[4]   Smartphone ECG for evaluation of STEMI: Results of the ST LEUIS Pilot Study [J].
Muhlestein, Joseph Boone ;
Viet Le ;
Albert, David ;
Moreno, Fidela Ll. ;
Anderson, Jeffrey L. ;
Yanowitz, Frank ;
Vranian, Robert B. ;
Barsness, Gregory W. ;
Bethea, Charles F. ;
Severance, Harry W. ;
Ramo, Barry ;
Pierce, John ;
Barbagelata, Alejandro ;
Muhlestein, Joseph Brent .
JOURNAL OF ELECTROCARDIOLOGY, 2015, 48 (02) :249-259
[5]   Ubiquitous Wireless ECG Recording: A Powerful Tool Physicians Should Embrace [J].
Saxon, Leslie A. .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2013, 24 (04) :480-483
[6]   Using a novel wireless system for monitoring patients after the atrial fibrillation ablation procedure: The iTransmit study [J].
Tarakji, Khaldoun G. ;
Wazni, Oussama M. ;
Callahan, Thomas ;
Kanj, Mohamed ;
Hakim, Ali H. ;
Wolski, Kathy ;
Wilkoff, Bruce L. ;
Saliba, Walid ;
Lindsay, Bruce D. .
HEART RHYTHM, 2015, 12 (03) :554-559
[7]  
Tilley L. P., 1995, ESSENTIALS CANINE FE