Target polarization measurements with a crossed-coil NMR polarimeter

被引:0
|
作者
Caracappa, A [1 ]
Thorn, C [1 ]
机构
[1] Brookhaven Natl Lab, Upton, NY 11973 USA
来源
SPIN 2002 | 2003年 / 675卷
关键词
D O I
暂无
中图分类号
O57 [原子核物理学、高能物理学];
学科分类号
070202 ;
摘要
We have performed a complete electronic circuit analysis of the crossed coil NMR polarimeter (CC-meter), and from this analysis we have determined the optimum conditions for its operation. From this analysis, which is confirmed by NMR measurements on hydrogen, we conclude that the CC-meter can be operated to give a very high signal-to-noise ratio (SMR) for thermal equilibrium polarization while also producing a highly linear response for fully polarized targets. In general, a well-designed and properly constructed CC-meter can provide a larger SNR at similar non-linearity, compared to the more commonly used Q-meter.
引用
收藏
页码:867 / 871
页数:5