Realistic loads for testing hip implants

被引:196
作者
Bergmann, G. [1 ]
Graichen, F. [1 ]
Rohlmann, A. [1 ]
Bender, A. [1 ,2 ]
Heinlein, B. [1 ]
Duda, G. N. [1 ]
Heller, M. O. [1 ]
Morlock, M. M. [3 ]
机构
[1] Charite, Julius Wolff Inst, D-13353 Berlin, Germany
[2] Berlin Brandenburg Ctr Regenerat Therapies, Berlin, Germany
[3] Tech Univ Hamburg, Biomech Sect, Hamburg, Germany
关键词
Hip implant; THR; load; force; telemetry; CEMENTLESS FEMORAL STEMS; STRAIN DISTRIBUTION; MUSCLE FORCES; CONTACT FORCES; PROXIMAL FEMUR; WALKING; JOINT; STABILITY; SIMULATION; COMPONENT;
D O I
10.3233/BME-2010-0616
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim here was to define realistic load conditions for hip implants, based on in vivo contact force measurements, and to see whether current ISO standards indeed simulate real loads. The load scenarios obtained are based on in vivo hip contact forces measured in 4 patients during different activities and on activity records from 31 patients. The load scenarios can be adapted to various test purposes by applying average or high peak loads, high-impact activities or additional low-impact activities, and by simulating normal or very active patients. The most strenuous activities are walking (average peak forces 1800 N, high peak forces 3900 N), going up stairs (average peak forces 1900 N, high peak forces 4200 N) and stumbling (high peak forces 11,000 N). Torsional moments are 50% higher for going up stairs than for walking. Ten million loading cycles simulate an implantation time of 3.9 years in active patients. The in vitro fatigue properties of cementless implant fixations are exceeded during stumbling. At least for heavyweight and very active subjects, the real load conditions are more critical than those defined by the ISO standards for fatigue tests.
引用
收藏
页码:65 / 75
页数:11
相关论文
共 28 条
[1]  
[Anonymous], 72064 ISO
[2]  
[Anonymous], 1995, 72068 ISO
[3]   HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1993, 26 (08) :969-990
[4]   Hip contact forces and gait patterns from routine activities [J].
Bergmann, G ;
Deuretzbacher, G ;
Heller, M ;
Graichen, F ;
Rohlmann, A ;
Strauss, J ;
Duda, GN .
JOURNAL OF BIOMECHANICS, 2001, 34 (07) :859-871
[5]   Hip joint contact forces during stumbling [J].
Bergmann, G ;
Graichen, F ;
Rohlmann, A .
LANGENBECKS ARCHIVES OF SURGERY, 2004, 389 (01) :53-59
[6]   Averaging of strongly varying signals [J].
Bergmann, G ;
Gandini, R ;
Ruder, H .
BIOMEDIZINISCHE TECHNIK, 2001, 46 (06) :168-171
[7]   IS STAIRCASE WALKING A RISK FOR THE FIXATION OF HIP IMPLANTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1995, 28 (05) :535-553
[8]  
Bergmann G., 2001, HIP98 LOADING HIP JO
[9]   The effect of muscle loading on the simulation of bone remodelling in the proximal femur [J].
Bitsakos, C ;
Kerner, J ;
Fisher, I ;
Amis, AA .
JOURNAL OF BIOMECHANICS, 2005, 38 (01) :133-139
[10]   Mechanical simulation of muscle loading on the proximal femur: analysis of cemented femoral component migration with and without muscle loading [J].
Britton, JR ;
Walsh, LA ;
Prendergast, PJ .
CLINICAL BIOMECHANICS, 2003, 18 (07) :637-646