Validation of a statistical shape model-based 2D/3D reconstruction method for determination of cup orientation after THA

被引:23
作者
Zheng, G. [1 ]
von Recum, J. [2 ]
Nolte, L. -P. [1 ]
Gruetzner, P. A. [2 ]
Steppacher, S. D. [3 ]
Franke, J. [2 ]
机构
[1] Univ Bern, Inst Surg Technol & Biomech, CH-3014 Bern, Switzerland
[2] Univ Heidelberg Hosp, BG Trauma Ctr Ludwigshafen, Ludwigshafen, Germany
[3] Univ Bern, Inselspital, Dept Orthopaed Surg, CH-3010 Bern, Switzerland
基金
瑞士国家科学基金会;
关键词
2D/3D reconstruction; Statistical shape model; Total hip arthroplasty; Cup orientation; X-ray radiograph; TOTAL HIP-ARTHROPLASTY; X-RAY RADIOGRAPH; ACETABULAR CUP; REGISTRATION; REPLACEMENT; ANTEVERSION; PLACEMENT; POSITION; IMPINGEMENT; CONTAINMENT;
D O I
10.1007/s11548-011-0644-7
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim of this study was to validate the accuracy and reproducibility of a statistical shape model-based 2D/3D reconstruction method for determining cup orientation after total hip arthroplasty. With a statistical shape model, this method allows reconstructing a patient-specific 3D-model of the pelvis from a standard AP X-ray radiograph. Cup orientation (inclination and anteversion) is then calculated with respect to the anterior pelvic plane that is derived from the reconstructed model. The validation study was conducted retrospectively on datasets of 29 patients (31 hips). Among them, there were 15 men (15 hips) and 14 women (16 hips). The average age of the patients was 69.4 +/- 8.5 (49-82) years. Each dataset has one postoperative X-ray radiograph and one postoperative CT scan. The postoperative CT scan for each patient was used to establish the ground truth for the cup orientation. The cup anteversion and inclination that were calculated from the 2D/3D reconstruction method were compared to the associated ground truth. To validate reproducibility and reliability, two observers performed measurements for each dataset twice in order to measure the reproducibility and the reliability of the 2D/3D reconstruction method. Our validation study demonstrated a mean accuracy of 0.4 +/- 1.8A degrees A (-2.6A degrees to 3.3A degrees) for inclination and a mean accuracy of 0.6 +/- 1.5A degrees A (-2.0A degrees to 3.9A degrees) for anteversion. Through the Bland-Altman analysis, no systematic errors in accuracy were detected. The method showed very good consistency for both parameters. Our validation results demonstrate that the statistical shape model-based 2D/3D reconstruction-based method is an accurate, consistent, and reproducible technique to measure cup orientation from postoperative X-ray radiographs. The best results were achieved with radiographs including the bilateral anterior superior iliac spines and the cranial part of non-fractured pelvises.
引用
收藏
页码:225 / 231
页数:7
相关论文
共 32 条
[1]  
[Anonymous], J BONE JT SURG BR B
[2]   Minimal Radiation Dose Computed Tomography for Measurement of Cup Orientation in Total Hip Arthroplasty [J].
Arai, Noriyuki ;
Nakamura, Shigeru ;
Matsushita, Takashi ;
Suzuki, Shigeru .
JOURNAL OF ARTHROPLASTY, 2010, 25 (02) :263-267
[3]  
Bader RJ., 2001, Hip Int, V11, P80, DOI [10.1177/112070000101100203, DOI 10.1177/112070000101100203]
[4]   Cup positioning in THA: current status and pitfalls. A systematic evaluation of the literature [J].
Beckmann, J. ;
Luering, C. ;
Tingart, M. ;
Anders, S. ;
Grifka, J. ;
Koeck, F. X. .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2009, 129 (07) :863-872
[5]   Measurement error and correlation coefficients [J].
Bland, JM ;
Altman, DG .
BRITISH MEDICAL JOURNAL, 1996, 313 (7048) :41-42
[6]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[7]  
Blendea Sorin, 2005, Comput Aided Surg, V10, P37, DOI 10.1080/10929080500178032
[8]   Primary total hip arthroplasty with a flanged, cemented all-polyethylene acetabular component - Evaluation at a minimum of 20 years [J].
Della Valle, CJ ;
Kaplan, K ;
Jazrawi, A ;
Ahmed, S ;
Jaffe, WL .
JOURNAL OF ARTHROPLASTY, 2004, 19 (01) :23-26
[9]   Computer assisted measurement of cup placement in total hip replacement [J].
Jaramaz, B ;
DiGioia, AM ;
Blackwell, M ;
Nikou, C .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1998, (354) :70-81
[10]  
Jaramaz B, 2006, LECT NOTES COMPUT SC, V4191, P653