Automated flow injection method for cadmium determination with pre-concentration and on-line reagent preparation.

被引:12
作者
Di Nezio, MS [1 ]
Palomeque, ME [1 ]
Band, BSF [1 ]
机构
[1] Univ Nacl Sur, FIA Lab, Dept Chem, RA-8000 Bahia Blanca, Argentina
来源
QUIMICA NOVA | 2005年 / 28卷 / 01期
关键词
Cd (II); preconcentration; FIA;
D O I
10.1590/S0100-40422005000100025
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
AUTOMATED FLOW INJECTION METHOD FOR CADMIUM DETERMINATION WITH PRE-CONCENTRATION AND ONLINE REAGENT PREPARATION. The spectrophotometric determination of Cd(II) using a flow injection system provided with a solid-phase reactor for cadmium preconcentration and on-line reagent preparation, is described. It is based on the formation of a dithizone-Cd complex in basic medium. The calibration curve is linear between 6 and 300 pg L-1 Cd(II), with a detection limit of 5.4 mug L-1, an RSD of 3.7% (10 replicates in duplicate) and a sample frequency of 11.4 h(-1). The proposed method was satisfactorily applied to the determination of Cd(II) in surface, well and drinking waters.
引用
收藏
页码:145 / 148
页数:4
相关论文
共 13 条
[1]   Metallothionein isoforms separation and cadmium speciation by capillary electrophoresis with ultraviolet and quadrupole-inductively coupled plasma mass spectrometric detection [J].
Alvarez-Llamas, G ;
de la Campa, MRF ;
Sanz-Medel, A .
ANALYTICA CHIMICA ACTA, 2001, 448 (1-2) :105-119
[2]   On-line liquid-liquid extraction system using a new phase separator for flame atomic absorption spectrometric determination of ultra-trace cadmium in natural waters [J].
Anthemidis, AN ;
Zachariadis, GA ;
Farastelis, CG ;
Stratis, JA .
TALANTA, 2004, 62 (03) :437-443
[3]   ION EXCHANGE SEPARATION OF ZINC, CADMIUM, AND MERCURY IN AQUEOUS AND PARTIAL NONAQUEOUS MEDIA [J].
BERG, EW ;
TRUEMPER, JT .
ANALYTICAL CHEMISTRY, 1958, 30 (11) :1827-1830
[4]   Determination of barium, chromium, cadmium, manganese, lead and zinc in atmospheric particulate matter by inductively coupled plasma atomic emission spectrometry (ICP-AES) [J].
Boevski, I ;
Daskalova, N ;
Havezov, I .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2000, 55 (11) :1643-1657
[5]  
FANG Z, 1993, FLOW INJECTION SEPAR, P12
[6]   Spectrophotometric studies on the simultaneous determination of cadmium and mercury with 4-(2-pyridylazo)-resoreinol [J].
Hashem, EY .
SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2002, 58 (07) :1401-1410
[7]  
MASSART DL, 1997, HDB CHEMOMETRICS QUA, P98
[8]   Determination of cadmium by FAAS after on-line enrichment using a mini column packed with Amberlite XAD-2 loaded with TAM [J].
Melo, MHA ;
Ferreira, SLC ;
Santelli, RE .
MICROCHEMICAL JOURNAL, 2000, 65 (01) :59-65
[9]   MICELLAR COLORIMETRIC DETERMINATION OF DITHIZONE METAL-CHELATES [J].
PARADKAR, RP ;
WILLIAMS, RR .
ANALYTICAL CHEMISTRY, 1994, 66 (17) :2752-2756
[10]   Photometric screening-test for heavy metals under flow injection conditions using extractive determination with dithizone [J].
Romberg, B ;
Muller, H .
ANALYTICA CHIMICA ACTA, 1997, 353 (2-3) :165-172