Correlations Between Controlled Endotracheal Tube Cuff Pressure and Postprocedural Complications: A Multicenter Study

被引:163
作者
Liu, Jianhui [1 ]
Zhang, Xiaoqing [1 ]
Gong, Wei [2 ]
Li, Shitong [2 ]
Wang, Fen [3 ]
Fu, Shukun [3 ]
Zhang, Mazhong [4 ]
Hang, Yannan [4 ]
机构
[1] Tongji Univ, Sch Med, Tongji Hosp, Dept Anesthesiol, Shanghai 200065, Peoples R China
[2] Jiao Tong Univ, Sch Med, Peoples Hosp 1, Dept Anesthesiol, Shanghai 200030, Peoples R China
[3] Tongji Univ, Sch Med, Shanghai Peoples Hosp 10, Dept Anesthesiol, Shanghai 200065, Peoples R China
[4] Jiao Tong Univ, Sch Med, Renji Hosp, Dept Anesthesiol, Shanghai 200030, Peoples R China
关键词
INTUBATION;
D O I
10.1213/ANE.0b013e3181f2ecc7
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
BACKGROUND: Postoperative respiratory complications related to endotracheal intubation usually present as cough, sore throat, hoarseness, and blood-streaked expectorant. In this study, we investigated the short-term (hours) impact of measuring and controlling endotracheal tube cuff (ETTc) pressure on postprocedural complications. METHODS: Five hundred nine patients from 4 tertiary care university hospitals in Shanghai, China scheduled for elective surgery under general anesthesia were assigned to a control group without measuring ETTc pressure, and a study group with ETTc pressure measured and adjusted. The duration of the procedure and duration of endotracheal intubation were recorded. Twenty patients whose duration of endotracheal intubation was between 120 and 180 minutes were selected from each group and examined by fiberoptic bronchoscopy immediately after removing the endotracheal tube. Endotracheal intubation related complications including cough, sore throat, hoarseness, and blood-streaked expectorant were recorded at 24 hours postextubation. RESULTS: There was no significant difference in sex, age, height, weight, procedure duration, and duration of endotracheal intubation between the 2 groups. The mean ETTc pressure measured after estimation by palpation of the pilot balloon of the study group was 43 +/- 23.3 mm Hg before adjustment (the highest was 210 mm Hg), and 20 +/- 3.1 mm Hg after adjustment (P < 0.001). The incidence of postprocedural sore throat, hoarseness, and blood-streaked expectoration in the control group was significantly higher than in the study group. As the duration of endotracheal intubation increased, the incidence of sore throat and blood-streaked expectoration in the control group increased. The incidence of sore throat in the study group also increased with increasing duration of endotracheal intubation. Fiberoptic bronchoscopy in the 20 patients showed that the tracheal mucosa was injured in varying degrees in both groups, but the injury was more severe in the control group than in the study group. CONCLUSIONS: ETTc pressure estimated by palpation with personal experience is often much higher than measured or what may be optimal. Proper control of ETTc pressure by a manometer helped reduce ETT-related postprocedural respiratory complications such as cough, sore throat, hoarseness, and blood-streaked expectoration even in procedures of short duration (1-3 hours). (Anesth Analg 2010;111:1133-7)
引用
收藏
页码:1133 / 1137
页数:5
相关论文
共 15 条
[1]   Intracuff Pressure and Tracheal Morbidity - Influence of Filling Cuff with Saline during Nitrous Oxide Anesthesia [J].
Combes, X ;
Schauvliege, F ;
Peyrouset, O ;
Motamed, C ;
Kirov, K ;
Dhonneur, G ;
Duvaldestin, P .
ANESTHESIOLOGY, 2001, 95 (05) :1120-1124
[2]  
Curiel Garcia J A, 2001, Gac Med Mex, V137, P179
[3]   Obstructive fibrinous tracheal pseudomembrane -: A potentially fatal complication of tracheal intubation [J].
Deslée, G ;
Brichet, A ;
Lebuffe, G ;
Copin, MC ;
Ramon, P ;
Marquette, CH .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2000, 162 (03) :1169-1171
[4]   Tracheal rupture complicating emergent endotracheal intubation [J].
Fan, CM ;
Ko, PCI ;
Tsai, KC ;
Chiang, WC ;
Chang, MC ;
Chen, WJ ;
Yuan, A .
AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2004, 22 (04) :289-293
[5]   ENDOTRACHEAL-TUBE CUFF PRESSURE ASSESSMENT - PITFALLS OF FINGER ESTIMATION AND NEED FOR OBJECTIVE MEASUREMENT [J].
FERNANDEZ, R ;
BLANCH, L ;
MANCEBO, J ;
BONSOMS, N ;
ARTIGAS, A .
CRITICAL CARE MEDICINE, 1990, 18 (12) :1423-1426
[6]  
Ganner C, 2001, Br J Nurs, V10, P1127
[7]  
Granja C, 2002, Rev Esp Anestesiol Reanim, V49, P137
[8]   Influence of airway pressure on minimum occlusive endotracheal tube cuff pressure [J].
Guyton, DC ;
Barlow, MR ;
Besselievre, TR .
CRITICAL CARE MEDICINE, 1997, 25 (01) :91-94
[9]   Acute tracheal rupture related to endotracheal intubation: Case report [J].
Harris, R ;
Joseph, A .
JOURNAL OF EMERGENCY MEDICINE, 2000, 18 (01) :35-39
[10]   Experienced emergency medicine physicians cannot safety inflate or estimate endotracheal tube cuff pressure using standard techniques [J].
Hoffman, RJ ;
Parwani, V ;
Hahn, IH .
AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2006, 24 (02) :139-143