Effects of washer on the stress distribution of mini-implant A finite element analysis

被引:12
作者
Jang, Hyoung-Jun [2 ]
Kwon, Soon-Yong [3 ]
Kim, Seong-Hun [1 ]
Park, Young-Guk [1 ]
Kim, Su-Jung [1 ]
机构
[1] Kyung Hee Univ, Dept Orthodont, Seoul, South Korea
[2] Kyung Hee Univ, Grad Sch Dent, Seoul, South Korea
[3] Cent Orthodont Clin, Seoul, South Korea
关键词
Finite element analysis; Mini-implant; Stability; Washer; ORTHODONTIC ANCHORAGE; MINISCREW IMPLANTS; BONE; STABILITY; FAILURE;
D O I
10.2319/021411-107.1
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: To investigate the biomechanical effects of 'washer' designed for improving mini-implant stability. Materials and Methods: Four three-dimensional finite element models of the mini-implant and surrounding bone were constructed with washers in different spike lengths (1.5 mm, 2.0 mm, and 2.5 mm). The force was applied in two directions (45 degrees and 90 degrees). The stress distribution on surrounding bone and the displacement of the mini-implant were analyzed. Plots of tensile stress, compression stress, and displacement were calculated, and maximum values in each category were analyzed. Results: The stress distribution was different between the models with washer and without washer. However, no remarkable differences in stress distribution were observed among the models with washer, regardless of spike length. A significantly greater displacement value was observed in the model without washer compared to the models with washer, but no notable difference in displacement value was found among the models with washer. The plots of the displacement distribution of the models with washer presented notable pattern differences as compared with that of the model without washer. Conclusion: With the use of the washer, a more homogeneous distribution of bone stress and less displacement of the mini-implant can be achieved. (Angle Orthod. 2012;82:137-144.)
引用
收藏
页码:137 / 144
页数:8
相关论文
共 50 条
  • [41] Comparative Evaluation of Three Abutment-Implant Interfaces on Stress Distribution in and Around Different Implant Systems: A Finite Element Analysis
    Kharsan, Vishwas
    Bandgar, Vaibhav
    Mirza, Aftab
    Jagtiani, Kiran
    Dhariwal, Neha
    Kore, Rahul
    CONTEMPORARY CLINICAL DENTISTRY, 2019, 10 (04) : 590 - 594
  • [42] EFFECTS OF LINGUALIZED AND LINEAR OCCLUSION SCHEMES ON THE STRESS DISTRIBUTION OF AN IMPLANT RETAINED OVERDENTURE USING FINITE ELEMENT ANALYSIS
    Abu Hasan, Md
    PROCEEDINGS OF THE ASME INTERNATIONAL MECHANICAL ENGINEERING CONGRESS AND EXPOSITION, 2015, VOL 3, 2016,
  • [43] Effect of Implant Design on Stress Distribution: A Finite Element Study
    Houshmand B.
    Ardakani M.T.
    Moscowchi A.
    Oskoui I.Z.
    Journal of Long-Term Effects of Medical Implants, 2022, 32 (04) : 39 - 45
  • [44] Influence of insertion depth on stress distribution in orthodontic miniscrew and the surrounding bone by finite element analysis
    Hirai, Yuzo
    Watanabe, Keiichiro
    Deguchi, Toru
    Ueda, Kanji
    Hamada, Kenichi
    Tanaka, Eiji
    DENTAL MATERIALS JOURNAL, 2021, 40 (05) : 1270 - 1276
  • [45] Predisposing Factors for Orthodontic Mini-Implant Failure Defined by Bone Strains in Patient-Specific Finite Element Models
    Mhd Hassan Albogha
    Toru Kitahara
    Mitsugu Todo
    Hiroto Hyakutake
    Ichiro Takahashi
    Annals of Biomedical Engineering, 2016, 44 : 2948 - 2956
  • [46] Three-Dimensional Finite Element Analysis on Stress Distribution of Internal Implant-Abutment Engagement Features
    Cho, Sung-Yong
    Huh, Yun-Hyuk
    Park, Chan-Jin
    Cho, Lee-Ra
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2018, 33 (02) : 319 - 327
  • [47] Influence of the implant prosthetic restorations components on the stress values and distribution in surrounding bone- a finite element analysis
    Tanasescu, L. A.
    Farcasiu, A. T.
    Farcasiu, C.
    Imre, M.
    Tierean, M. H.
    Munteanu, D.
    Andrei, O. C.
    2015 E-HEALTH AND BIOENGINEERING CONFERENCE (EHB), 2015,
  • [48] Single implant retained overdentures: Evaluation of effect of implant length and diameter on stress distribution by finite element analysis
    Eraslan, Ravza
    Yagci, Filiz
    Esim, Emir
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2024, 33 (04): : 348 - 357
  • [49] Orthodontic Mini-Implant Stability under Continuous or Intermittent Loading: A Histomorphometric and Biomechanical Analysis
    Wu, Yeke
    Xu, Zhenrui
    Tan, Luyuan
    Tan, Lijun
    Zhao, Zhihe
    Yang, Pu
    Li, Yu
    Tang, Tian
    Zhao, Lixing
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2015, 17 (01) : 163 - 172
  • [50] Effects of Implant Diameter, Insertion Depth, and Loading Angle on Stress/Strain Fields in Implant/Jawbone Systems: Finite Element Analysis
    Qian, Lihe
    Todo, Mitsugu
    Matsushita, Yasuyuki
    Koyano, Kiyoshi
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2009, 24 (05) : 877 - 886