Reducing Patient Radiation Dose during CT-Guided Procedures: Demonstration in Spinal Injections for Pain

被引:37
作者
Shepherd, T. M. [1 ]
Hess, C. P. [1 ]
Chin, C. T. [1 ]
Gould, R. [1 ]
Dillon, W. P. [1 ]
机构
[1] Univ Calif San Francisco, Dept Radiol & Biomed Imaging, Neuroradiol Div, San Francisco, CA 94143 USA
基金
美国国家卫生研究院;
关键词
COMPUTED-TOMOGRAPHY; OPTIMIZATION;
D O I
10.3174/ajnr.A2634
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND AND PURPOSE: CT guidance may improve precision for diagnostic and therapeutic spinal injections, but it can increase patient radiation dose. This study examined the impact of reducing tube current on patient radiation exposure and the technical success for these procedures, by using axial acquisitions for short scan lengths and eliminating nonessential imaging. MATERIALS AND METHODS: Our institutional review board approved retrospective analysis of records from 100 consecutive outpatients undergoing spinal injections for pain before and after the CT protocol modification to reduce radiation dose. Data collected included patient age and sex, response to injection, number of sites and spinal levels treated, injection type, performing physician, CT acquisition method, number of imaging series, tube current, scan length, and DLP. RESULTS: Image contrast was reduced with the low-dose protocol, but this did not affect technical success or immediate pain relief. Mean DLP for all procedures decreased from 1458 +/- 1022 to 199 +/- 101 mGy . cm (P < .001). The range of radiologist-dependent DLP per procedure also was reduced significantly with the modified protocol. Selective nerve root blocks, lumbar injections, multiple injection sites, and the lack of prior imaging were each associated with a slightly higher DLP (<50 mGy . cm). CONCLUSIONS: Radiation to patients undergoing CT-guided spinal injections can be decreased significantly without affecting outcome by reducing tube current, using axial acquisitions for short scan lengths, and eliminating nonessential imaging guidance. These measures also decrease variability in radiation doses between different practitioners and should be useful for other CT-guided procedures in radiology.
引用
收藏
页码:1776 / 1782
页数:7
相关论文
共 26 条
  • [1] Low-dose helical computed tomography (CT) in the perioperative workup of adolescent idiopathic scoliosis
    Abul-Kasim, Kasim
    Overgaard, Angelica
    Maly, Pavel
    Ohlin, Acke
    Gunnarsson, Mikael
    Sundgren, Pia C.
    [J]. EUROPEAN RADIOLOGY, 2009, 19 (03) : 610 - 618
  • [2] Use of 100 kV versus 120 kV in cardiac dual source computed tomography: effect on radiation dose and image quality
    Blankstein, Ron
    Bolen, Michael A.
    Pale, Rodrigo
    Murphy, Meagan K.
    Shah, Amar B.
    Bezerra, Hiram G.
    Sarwar, Ammar
    Rogers, Ian S.
    Hoffmann, Udo
    Abbara, Suhny
    Cury, Ricardo C.
    Brady, Thomas J.
    [J]. INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2011, 27 (04) : 579 - 586
  • [3] Bogdanich W, 2009, NY TIMES, pA13
  • [4] Multidetector CT in patients suspected of having lumbar disk herniation: Comparison of standard-dose and simulated low-dose techniques
    Bohy, Pascale
    de Maertelaer, Viviane
    Roquigny, Aymeric
    Keyzer, Caroline
    Tack, Denis
    Gevenois, Pierre Alain
    [J]. RADIOLOGY, 2007, 244 (02) : 524 - 531
  • [5] Current concepts - Computed tomography - An increasing source of radiation exposure
    Brenner, David J.
    Hall, Eric J.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) : 2277 - 2284
  • [6] Primary care - Low back pain
    Deyo, RA
    Weinstein, JN
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2001, 344 (05) : 363 - 370
  • [7] Diagnostic Imaging Council CT Committee, 2008, 96 AAPM DIAGN IM COU
  • [8] Reducing the Radiation Dose for CT Colonography Using Adaptive Statistical Iterative Reconstruction: A Pilot Study
    Flicek, Kristina T.
    Hara, Amy K.
    Silva, Alvin C.
    Wu, Qing
    Peter, Mary B.
    Johnson, C. Daniel
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 195 (01) : 126 - 131
  • [9] CT-guided interventional procedures for pain management in the lumbosacral spine
    Gangi, A
    Dietemann, JL
    Mortazavi, R
    Pfleger, D
    Kauff, C
    Roy, C
    [J]. RADIOGRAPHICS, 1998, 18 (03) : 621 - 633
  • [10] Application of blood-pool agents in visualization of peripheral vessels
    Giovagnoni, Andrea
    Catalano, Carlo
    [J]. EUROPEAN RADIOLOGY, 2007, 17 : B18 - B23