Portable tetracycline analyzer based on LED-excitation europium-sensitized luminescence

被引:7
作者
Chen, Guoying [1 ]
机构
[1] USDA ARS, Eastern Reg Res Ctr, Wyndmoor, PA 19038 USA
来源
SENSORS AND ACTUATORS B-CHEMICAL | 2008年 / 130卷 / 02期
关键词
portable; analyzer; LED; tetracycline; europium-sensitized luminescence; time-resolved luminescence;
D O I
10.1016/j.snb.2007.10.046
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Described in detail here is the design of a portable tetracycline (TC) analyzer based on europium-sensitized luminescence (ESL). The instrument operation and data processing are controlled by a laptop computer running a custom LabVIEW program. A 385-nm light emitting diode (LED) is used in pulse mode as excitation source. Compared to a conventional xenon flashlamp, an LED has a narrow emission bandwidth, weak residual radiation, and orders-of-magnitude-lower cost, weight, and power consumption. The time-resolved luminescence (TRL) signal is detected by a gated photomultiplier tube (PMT), and normalized by monitoring the energy of individual LED pulses with a photodiode (PD). Over three orders of magnitude (0.1-100 ppb TC), the TRL signal maintains linear dependence on TC concentration; beyond 100 ppb filter effects becomes obvious. A 0.060-ppb limit of detection (LOD) is achieved with a typical similar to 5% relative standard deviation (R.S.D.). The background noise corresponds to similar to 0.05 ppb. At 16 pounds, the analyzer is readily deployable for field analysis. Published by Elsevier B.V.
引用
收藏
页码:750 / 757
页数:8
相关论文
共 33 条