Geant4 Simulation for Commissioning of Proton Therapy Centre

被引:1
|
作者
Tan, Hong Qi [1 ]
Phua, Jun Hao [2 ]
Tan, Lloyd [2 ]
Ang, Khong Wei [2 ]
Lee, James [2 ]
Bettiol, Andrew Anthony [1 ]
机构
[1] Natl Univ Singapore, Dept Phys, Ctr Ion Beam Applicat, Singapore, Singapore
[2] Natl Canc Ctr Singapore, Singapore, Singapore
来源
WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2018, VOL 1 | 2019年 / 68卷 / 01期
关键词
Proton therapy; Monte Carlo; Beamline simulation;
D O I
10.1007/978-981-10-9035-6_108
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In anticipation of the new Proton Therapy Centre in Singapore at 2020, the medical physicists and physicists in NCCS and NUS have come together to do several preparation/preliminary works. Of which, one of the most important programme is to develop the Monte Carlo simulation capability using GEANT4 toolkit. We simulated the entire beam delivery and nozzle based on specifications given by Hitachi. The results of the Energy-Range curves in water phantom together with spot size in air agrees well with measurement data provided by Hitachi. We will show the complete methodology of simulation leading up to the comparison with measurement data. There were 3 main aspects we focused on-(1) The choice of physics models, (2) The determination of initial proton phase space based on Twiss parameters from beam profile measurement and (3) The choice of cut-off energy and step size. This simulation is important for the commissioning of the Proton Therapy System and the methodology presented will be helpful for other upcoming centers who are interested in doing the same.
引用
收藏
页码:583 / 587
页数:5
相关论文
共 50 条
  • [1] Verification of the dose distributions with GEANT4 simulation for proton therapy
    Aso, T
    Kimura, A
    Tanaka, S
    Yoshida, H
    Kanematsu, N
    Sasaki, T
    Akagi, T
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (04) : 896 - 901
  • [2] Verification of the dose distributions with GEANT4 simulation for proton therapy
    Aso, T
    Kimura, A
    Tanaka, S
    Yoshida, H
    Kanematsu, N
    Sasaki, T
    Akagi, T
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 2138 - 2142
  • [3] Physics settings for using the Geant4 toolkit in proton therapy
    Jarlskog, Christina Zacharatou
    Paganetti, Harald
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2008, 55 (03) : 1018 - 1025
  • [4] Significance of the proton energy loss mechanism to gold nanoparticles in proton therapy: a Geant4 simulation
    Tabbakh, Farshid
    SCIENTIFIC REPORTS, 2024, 14 (01):
  • [5] Pencil Beam Algorithm on the GEANT4 Based Simulation Framework and Verification of the Dose Distributions for Proton Therapy
    Aso, Tsukasa
    Yamashita, Tomohiro
    Akagi, Takashi
    Sasaki, Takashi
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 2166 - +
  • [6] Dosimetric Comparison of Radiation Therapy and Proton Therapy in Prostate Cancer Using the Monte Carlo Simulation Platform GEANT4
    Ismail Ghazi
    Zineb Sobhy
    Mustapha Krim
    Othmane Kaanouch
    Meriem Tantaoui
    Abdelkrim Kartouni
    Jamal Inchaouh
    Hamid Chakir
    Said Ouaskit
    Moscow University Physics Bulletin, 2021, 76 : 326 - 332
  • [7] Dosimetric Comparison of Radiation Therapy and Proton Therapy in Prostate Cancer Using the Monte Carlo Simulation Platform GEANT4
    Ghazi, Ismail
    Sobhy, Zineb
    Krim, Mustapha
    Kaanouch, Othmane
    Tantaoui, Meriem
    Kartouni, Abdelkrim
    Inchaouh, Jamal
    Chakir, Hamid
    Ouaskit, Said
    MOSCOW UNIVERSITY PHYSICS BULLETIN, 2021, 76 (05) : 326 - 332
  • [8] PIXE Simulation With Geant4
    Pia, Maria Grazia
    Weidenspointner, Georg
    Augelli, Mauro
    Quintieri, Lina
    Saracco, Paolo
    Sudhakar, Manju
    Zoglauer, Andreas
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (06) : 3614 - 3649
  • [9] Validation of Geant4 Simulation of Proton Energy Straggling: First Results
    Han, Min Cheol
    Pia, Maria Grazia
    Saracco, Paolo
    2018 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE PROCEEDINGS (NSS/MIC), 2018,
  • [10] Validation of Geant4 Physics Models for the Simulation of the Proton Bragg Peak
    Cirrone, G. A. P.
    Cuttone, G.
    Di Rosa, F.
    Guatelli, S.
    Mascialino, B.
    Pia, M. G.
    Russo, G.
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 788 - 792