Detecting penicillin G in milk with impedimetric label-free immunosensor

被引:79
|
作者
Thavarungkul, Panote [1 ,2 ]
Dawan, Supapom [2 ,3 ]
Kanatharana, Proespichaya [2 ,3 ]
Asawatreratanakul, Punnee [2 ,4 ]
机构
[1] Prince Songkla Univ, Fac Sci, Dept Phys, Hat Yai 90112, Songkhla, Thailand
[2] Prince Songkla Univ, Biophys Res Unit Biosensors & Biocurrents, Hat Yai 90112, Songkhla, Thailand
[3] Prince Songkla Univ, Fac Sci, Dept Chem, Hat Yai 90112, Songkhla, Thailand
[4] Prince Songkla Univ, Fac Sci, Dept Biochem, Hat Yai 90112, Songkhla, Thailand
关键词
impedimetric; immunosensor; penicillin G; benzylpenicillin; label-free; flow injection;
D O I
10.1016/j.bios.2007.08.003
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A label-free impedimetric flow injection immunosensor for the direct detection of penicillin G has been developed. Anti-penicillin G was immobilized on a gold working electrode modified with a self-assembled monolayer of thioctic acid. Real time monitoring of impedance was carried out at the optimum frequency of 160 Hz. Under optimum operating conditions the system provided a wide linear range between 1.0 x 10(-13) and 1.0 x 10(-8) M with a very low detection limit of 3.0 x 10(-15) M, much lower than the MRL of penicillin G in milk (1.2 x 10(-8) M). The immobilized anti-penicillin G on self-assembled thioctic acid monolayer gold electrode was very stable and provided good reproducible signal after regeneration up to 45 times with a relative standard deviation (R.S.D.) lower than 4%. Good recoveries and precisions were obtained when spiked raw milk samples were analyzed. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:688 / 694
页数:7
相关论文
empty
未找到相关数据