Hemodynamic Investigation of the Effectiveness of a Two Overlapping Flow Diverter Configuration for Cerebral Aneurysm Treatment

被引:12
作者
Uchiyama, Yuya [1 ,2 ]
Fujimura, Soichiro [2 ,3 ]
Takao, Hiroyuki [1 ,2 ,4 ]
Suzuki, Takashi [2 ,5 ]
Hayakawa, Motoharu [6 ]
Ishibashi, Toshihiro [4 ]
Karagiozov, Kostadin [4 ]
Fukudome, Koji [3 ]
Murayama, Yuichi [4 ]
Yamamoto, Makoto [3 ]
机构
[1] Tokyo Univ Sci, Grad Sch Mech Engn, Tokyo 1258585, Japan
[2] Jikei Univ, Dept Innovat Med Informat Technol, Sch Med, Tokyo 1058461, Japan
[3] Tokyo Univ Sci, Dept Mech Engn, Tokyo 1258585, Japan
[4] Jikei Univ, Dept Neurosurg, Sch Med, Tokyo 1058461, Japan
[5] Siemens Healthcare KK, Digital Hlth & SYNGO Dept, Tokyo 1418644, Japan
[6] Fujita Hlth Univ, Dept Neurosurg, Toyoake, Aichi 4701192, Japan
来源
BIOENGINEERING-BASEL | 2021年 / 8卷 / 10期
关键词
flow diverter; computational fluid dynamics; endovascular treatment; double-stenting technique; hemodynamics; cerebral aneurysm;
D O I
10.3390/bioengineering8100143
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Flow diverters (FDs) are widely employed as endovascular treatment devices for large or wide-neck cerebral aneurysms. Occasionally, overlapped FDs are deployed to enhance the flow diversion effect. In this study, we investigated the hemodynamics of overlapping FDs via computational fluid dynamics (CFD) simulations. We reproduced the arterial geometry of a patient who had experienced the deployment of two overlapping FDs. We utilized two stent patterns, namely the patterns for one FD and two overlapping FDs. We calculated the velocity, mass flow rate, wall shear stress, and pressure loss coefficient as well as their change rates for each pattern relative to the no-FD pattern results. The CFD simulation results indicated that the characteristics of the blood flow inside the aneurysm were minimally affected by the deployment of a single FD; in contrast, the overlapping FD pattern results revealed significant changes in the flow. Further, the velocity at an inspection plane within the aneurysm sac decreased by up to 92.2% and 31.0% in the cases of the overlapping and single FD patterns, respectively, relative to the no-FD pattern. The simulations successfully reproduced the hemodynamics, and the qualitative and quantitative investigations are meaningful with regard to the clinical outcomes of overlapped FD deployment.
引用
收藏
页数:14
相关论文
共 44 条
[1]   Hemodynamics in a giant intracranial aneurysm characterized by in vitro 4D flow MRI [J].
Amili, Omid ;
Schiavazzi, Daniele ;
Moen, Sean ;
Jagadeesan, Bharathi ;
Van de Moortele, Pierre-Francois ;
Coletti, Filippo .
PLOS ONE, 2018, 13 (01)
[2]   Intracranial Stents Being Modeled as a Porous Medium: Flow Simulation in Stented Cerebral Aneurysms [J].
Augsburger, L. ;
Reymond, P. ;
Rufenacht, D. A. ;
Stergiopulos, N. .
ANNALS OF BIOMEDICAL ENGINEERING, 2011, 39 (02) :850-863
[3]   Computational Fluid Dynamics Using a Porous Media Setting Predicts Outcome after Flow-Diverter Treatment [J].
Beppu, M. ;
Tsuji, M. ;
Ishida, F. ;
Shirakawa, M. ;
Suzuki, H. ;
Yoshimura, S. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2020, 41 (11) :2107-2113
[4]   Geometrical deployment for braided stent [J].
Bouillot, Pierre ;
Brina, Olivier ;
Ouared, Rafik ;
Yilmaz, Hasan ;
Farhat, Mohamed ;
Erceg, Gorislav ;
Lovblad, Karl-Olof ;
Vargas, Maria Isabel ;
Kulcsar, Zsolt ;
Pereira, Vitor Mendes .
MEDICAL IMAGE ANALYSIS, 2016, 30 :85-94
[5]   Computational fluid dynamics modeling of intracranial aneurysms: Qualitative comparison with cerebral angiography [J].
Cebral, Juan R. ;
Pergolizzi, Richard S., Jr. ;
Putman, Christopher M. .
ACADEMIC RADIOLOGY, 2007, 14 (07) :804-813
[6]   BLOOD VISCOSITY - INFLUENCE OF ERYTHROCYTE AGGREGATION [J].
CHIEN, S ;
USAMI, S ;
DELLENBACK, RJ ;
GREGERSEN, MI ;
NANNINGA, LB ;
GUEST, MM .
SCIENCE, 1967, 157 (3790) :829-+
[7]   Computational Hemodynamics Analysis of Intracranial Aneurysms Treated with Flow Diverters: Correlation with Clinical Outcomes [J].
Chong, W. ;
Zhang, Y. ;
Qian, Y. ;
Lai, L. ;
Parker, G. ;
Mitchell, K. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2014, 35 (01) :136-142
[8]   Compacting a Single Flow Diverter versus Overlapping Flow Diverters for Intracranial Aneurysms: A Computational Study [J].
Damiano, R. J. ;
Tutino, V. M. ;
Paliwal, N. ;
Ma, D. ;
Davies, J. M. ;
Siddiqui, A. H. ;
Meng, H. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2017, 38 (03) :603-610
[9]   RELATIONSHIP OF PLASMA FIBRINOGEN, ERYTHROCYTE FLEXIBILITY AND BLOOD-VISCOSITY [J].
DUPONT, PA ;
SIRS, JA .
THROMBOSIS AND HAEMOSTASIS, 1977, 38 (03) :660-667
[10]   Characterization of volumetric flow rate waveforms in the normal internal carotid and vertebral arteries [J].
Ford, MD ;
Alperin, N ;
Lee, SH ;
Holdsworth, DW ;
Steinman, DA .
PHYSIOLOGICAL MEASUREMENT, 2005, 26 (04) :477-488