A Novel Technique Employing Tapered Fiber Bragg Grating to Solve the Axial/Transverse Forces Crosstalk in Microsurgical Instruments

被引:11
作者
Abushagur, Abdulfatah A. Ghaith [1 ]
Bakar, Ahmad Ashrif A. [1 ]
Bin Zan, Mohd Saiful Dzulkefly [1 ]
Arsad, Norhana [1 ]
机构
[1] Univ Kebangsaan Malaysia, Fac Engn & Built Environm, Dept Elect Elect & Syst Engn, Bangi 43600, Malaysia
关键词
Bio-tactile FBG force sensor; tapered FBG sensors; microsurgical instruments; fiber Bragg gratings; RETINAL MICROSURGERY; TEMPERATURE; SENSORS; WAVELENGTH;
D O I
10.1109/JSEN.2016.2602239
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper demonstrates the capability of retinal microsurgical instruments constructed with customized fiber Bragg gratings (FBGs) sensors to completely decouple the transverse and axial force components. In contrast to the constantly shifting Bragg wavelength, FBG bandwidth can only be tuned when a nonuniform strain/temperature distribution is applied. Using a tapered FBG (TFBG) at the neutral axis of the instrument makes the bandwidth tune-ability possible by applying forces both transversely and axially, while the bandwidth of the outer/lateral standard FBG (SFBG) due to the fiber symmetry remains insensitive to axial forces. Therefore, the crosstalk (transverse forces) in the TFBG can be easily filtered out when mixed forces are applied simultaneously. Tapered fiber mode analysis is carried out to calculate the local modes and is used in the analytic coupled-mode solution. Transfer-matrix method, then, is developed to analyze the reflected spectrum of the TFBG. We provide a developed mathematical model algorithm to show how force vectors can be measured. Tuning the reflected bandwidth is directly corresponding to the amount of the reflected optical power. Thus, the results show that the simulated needle sensors are able to measure transverse and axial forces with a sensitivity of 0.049 and 0.0026 dBm/mN, while the SFBG is insensitive to the axial forces. This novel method is applicable for microsurgical applications, i.e., vascular and cochlear implant surgeries and catheterization procedures.
引用
收藏
页码:7671 / 7680
页数:10
相关论文
共 33 条
  • [1] Abushagur AAG, 2014, 2014 IEEE 5TH INTERNATIONAL CONFERENCE ON PHOTONICS (ICP), P41, DOI 10.1109/ICP.2014.7002305
  • [2] Advances in Bio-Tactile Sensors for Minimally Invasive Surgery Using the Fibre Bragg Grating Force Sensor Technique: A Survey
    Abushagur, Abdulfatah A. G.
    Arsad, Norhana
    Reaz, Mamun Ibne
    Bakar, A. Ashrif A.
    [J]. SENSORS, 2014, 14 (04) : 6633 - 6665
  • [3] Abushagur Ousama M., 2005, Proceedings of the SPIE - The International Society for Optical Engineering, V5877, p58770Z, DOI 10.1117/12.614946
  • [4] Balicki M, 2010, LECT NOTES COMPUT SC, V6363, P303
  • [5] Heterodyne detection for fiber Bragg grating sensors
    Bellil, H
    Abushagur, MAG
    [J]. OPTICS AND LASER TECHNOLOGY, 2000, 32 (05) : 297 - 300
  • [6] Ultra-low-loss optical fiber nanotapers
    Brambilla, G
    Finazzi, V
    Richardson, DJ
    [J]. OPTICS EXPRESS, 2004, 12 (10): : 2258 - 2263
  • [7] A novel temperature-insensitive fiber Bragg grating sensor for displacement measurement
    Dong, XY
    Yang, XF
    Zhao, CL
    Ding, L
    Shum, P
    Ngo, NQ
    [J]. SMART MATERIALS & STRUCTURES, 2005, 14 (02) : N7 - N10
  • [8] Fiber grating spectra
    Erdogan, T
    [J]. JOURNAL OF LIGHTWAVE TECHNOLOGY, 1997, 15 (08) : 1277 - 1294
  • [9] Chirped Bragg grating fabricated in fused fibre taper for strain-temperature discrimination
    Frazao, O
    Melo, M
    Marques, PVS
    Santos, JL
    [J]. MEASUREMENT SCIENCE AND TECHNOLOGY, 2005, 16 (04) : 984 - 988
  • [10] Gupta PK, 1999, LECT NOTES COMPUT SC, V1679, P1218