Reconsideration of EMI Phenomenon in Active Implantable Medical Devices in the Age of MR Conditional Devices

被引:0
|
作者
Toyoshima, Takeshi [1 ]
机构
[1] USCI Holdings Inc, CRDM Business Unit, Tokyo, Japan
来源
2014 INTERNATIONAL SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY, TOKYO (EMC'14/TOKYO) | 2014年
关键词
AIMD; MRI; static magnet field; magnet field gradient; rf pulses; PACEMAKERS;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Until recently, the presence of an AIMD has beencontraindication to MRI because of the potential adverse effects of MRI on AIMDs. However, all of AIMD manufacturers have released "MR Conditional" devices recently. These devices have been demonstrated to pose no known hazards in a specified MRI environment with specified conditions of use. Therefore, it is quite important to understand what kind of clinical phenomena would take place if an accidental scan of a conventional device was performed. If the MRI scanning of a patient with non MR compatible AIMD is needed from special medical requirement, intensive monitoring of heart activity during the scanning and repetitive threshold measurements after the scanning will be essential.
引用
收藏
页码:234 / 236
页数:3
相关论文
共 50 条
  • [1] Practical Aspects of MR Imaging Safety Test Methods for MR Conditional Active Implantable Medical Devices
    Al-Dayeh, Louai
    Rahman, Mizan
    MAGNETIC RESONANCE IMAGING CLINICS OF NORTH AMERICA, 2020, 28 (04) : 559 - +
  • [2] MR safety assessment of active implantable medical devices
    Aissani, Sarra
    Laistler, Elmar
    Felblinger, Jacques
    RADIOLOGE, 2019, 59 (SUPPL 1): : 40 - 45
  • [3] EMI model of automatic implantable medical devices (AIMDs)
    Mahmood, Malik Mazhar
    2008 SECOND INTERNATIONAL CONFERENCE ON ELECTRICAL ENGINEERING, 2008, : 135 - 139
  • [4] Rechargeable Active Implantable Medical Devices (AIMDs)
    Newaskar, Deepali
    Patil, B. P.
    INTERNATIONAL JOURNAL OF ONLINE AND BIOMEDICAL ENGINEERING, 2023, 19 (13) : 108 - 119
  • [5] The Implications of RoHS on Active Implantable Medical Devices
    Savage, Timothy Scott
    2011 IEEE INTERNATIONAL RELIABILITY PHYSICS SYMPOSIUM (IRPS), 2011,
  • [6] EMC ASPECTS FOR ACTIVE IMPLANTABLE MEDICAL DEVICES
    Sponberg, Curt
    Reitan, Ron
    2017 IEEE INTERNATIONAL SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY & SIGNAL/POWER INTEGRITY (EMCSI), 2017,
  • [7] RF Safety of Active Implantable Medical Devices
    Silemek, Berk
    Acikel, Volkan
    Atalar, Ergin
    EMAGRES, 2019, 8 (02): : 103 - 119
  • [8] Implantable medical devices
    Crickx, B.
    Arrault, X.
    ANNALES DE DERMATOLOGIE ET DE VENEREOLOGIE, 2008, 135 (01): : S66 - S69
  • [9] Technical Challenges of Active Implantable Medical Devices for Neurotechnology
    Morales, Jorge Mario Herrera
    Clement, Claude
    2018 IEEE CPMT SYMPOSIUM JAPAN (ICSJ), 2018, : 77 - 80
  • [10] Electronic security systems and active implantable medical devices
    Irnich, W
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (08): : 1235 - 1258