Cracked cartridges during foldable intraocular lens implantation

被引:14
作者
Habib, NE
Singh, J
Adams, AD
Bartholomew, RS
机构
[1] Princess Alexandra Eye Pavilion, Edinburgh EH3 9HA, Chalmers Street
关键词
D O I
10.1016/S0886-3350(96)80023-X
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
We describe five routine phacoemulsification procedures in which a longitudinal split in the injector cartridge occurred during foldable intraocular lens (IOL) implantation. In some cases, this was associated with a broken lens. All lenses (STAAR AA-4203VF) were folded, and implantation was attempted through a 3.5 mm corneal or scleral incision. Visual outcome was not adversely affected in any case. The systems for introducing foldable IOLs have not been perfected, and further refinements in materials and techniques are required.