Durability of implanted electrodes and leads in an upper-limb neuroprosthesis

被引:49
作者
Kilgore, KL
Peckham, PH
Keith, MW
Montague, FW
Hart, RL
Gazdik, MM
Bryden, AM
Snyder, SA
Stage, TG
机构
[1] Metrohlth Med Ctr, Dept Orthopaed, Cleveland, OH 44109 USA
[2] Case Western Reserve Univ, Dept Biomed Engn, Cleveland, OH 44106 USA
[3] Dept Vet Affairs, Rehabil Res & Dev Serv, Cleveland, OH USA
关键词
D O I
10.1682/JRRD.2003.11.0457
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
Implanted neuroprosthetic systems have been successfully used to provide upper-limb function for over 16 years. A critical aspect of these implanted systems is the safety, stability, and reliability of the stimulating electrodes and leads. These components are (1) the stimulating electrode itself, (2) the electrode lead, and (3) the lead-to-device connector. A failure in any of these components causes the direct loss of the capability to activate a muscle consistently, usually resulting in a decrement in the function provided by the neuroprosthesis. Our results indicate that the electrode, lead, and connector system are extremely durable. We analyzed 238 electrodes that have been implanted as part of an upper-limb neuroprosthesis. Each electrode had been implanted at least 3 years, with a maximum implantation time of over 16 years. Only three electrode-lead failures and one electrode infection occurred, for a survival rate of almost 99 percent. Electrode threshold measurements indicate that the electrode response is stable over time, with no evidence of electrode migration or continual encapsulation in any of the electrodes studied. These results have an impact on the design of implantable neuroprosthetic systems. The electrode-lead component of these systems should no longer be considered a weak technological link.
引用
收藏
页码:457 / 468
页数:12
相关论文
共 29 条
[1]  
AKERS JM, 1996, IEEE T REHABIL ENG, V5, P207
[2]   Micromodular implants to provide electrical stimulation of paralyzed muscles and limbs [J].
Cameron, T ;
Loeb, GE ;
Peck, RA ;
Schulman, JH ;
Strojnik, P ;
Troyk, PR .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1997, 44 (09) :781-790
[3]   COMPLICATIONS OF IMPLANTABLE CARDIOVERTER DEFIBRILLATOR THERAPY - FOLLOW-UP OF 241 PATIENTS [J].
GRIMM, W ;
FLORES, BF ;
MARCHLINSKI, FE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (01) :218-222
[4]   LONG-TERM PERFORMANCE OF EPIMYOCARDIAL PACING LEADS IN ADULTS - COMPARISON WITH ENDOCARDIAL LEADS [J].
HELGUERA, ME ;
MALONEY, JD ;
WOSCOBOINIK, JR ;
TROHMAN, RG ;
MCCARTHY, PM ;
MORANT, VA ;
WILKOFF, BL ;
CASTLE, LW ;
PINSKI, SL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (03) :412-417
[5]  
JEAGER RJ, 1994, STIMULATING STEPS
[6]  
KAZAMA S, 1993, JPN HEART J, V34, P193
[7]   IMPLANTABLE FUNCTIONAL NEUROMUSCULAR STIMULATION IN THE TETRAPLEGIC HAND [J].
KEITH, MW ;
PECKHAM, PH ;
THROPE, GB ;
STROH, KC ;
SMITH, B ;
BUCKETT, JR ;
KILGORE, KL ;
JATICH, JW .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1989, 14A (03) :524-530
[8]   ELECTRODE CHARACTERIZATION FOR FUNCTIONAL APPLICATION TO UPPER EXTREMITY FNS [J].
KILGORE, KL ;
PECKHAM, PH ;
KEITH, MW ;
THROPE, GB .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1990, 37 (01) :12-21
[9]   An implanted upper-extremity neuroprosthesis - Follow-up of five patients [J].
Kilgore, KL ;
Peckham, PH ;
Keith, MW ;
Thrope, GB ;
Wuolle, KS ;
Bryden, AM ;
Hart, RL .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (04) :533-541
[10]   SYNTHESIS OF HAND GRASP USING FUNCTIONAL NEUROMUSCULAR STIMULATION [J].
KILGORE, KL ;
PECKHAM, PH ;
THROPE, GB ;
KEITH, MW ;
GALLAHERSTONE, KA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1989, 36 (07) :761-770