Home sphygmomanometers: what should a nephrologist know?

被引:5
作者
Akpolat, Tekin [1 ]
机构
[1] Ondokuz Mayis Univ, Tip Fak, Nefrol Bilim Dali, Sch Med, TR-55139 Samsun, Turkey
关键词
Accuracy; Calibration; Chronic kidney disease; Home sphygmomanometers; Hypertension; Validation; BLOOD-PRESSURE-MEASUREMENT; EUROPEAN-SOCIETY; INTERNATIONAL-PROTOCOL; HYPERTENSION GUIDELINES; ARTERIAL STIFFNESS; JAPANESE SOCIETY; KIDNEY-DISEASE; VALIDATION; MONITORS; PREVALENCE;
D O I
10.5301/JN.2010.2902
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Inaccuracy of home sphygmomanometers is a common, neglected and often ignored problem. The aims of this review are to emphasize the advantages of self-measurement of blood pressure (BP) at home, to summarize basic features of automated home BP measurement devices, to point out the distinction between validation, calibration and accuracy of home sphygmomanometers and to discuss some practical points for the improvement of BP control in chronic kidney disease (CKD) patients for nephrologists. Four features of automated sphygmomanometers need to be emphasized: measurement site, cuff-size, validation status and accuracy. The validation protocols do not guarantee the accuracy of a particular device for an individual patient. Special patient groups such as those with early stages of CKD or diabetic nephropathy or patients having extraosseous/vascular calcifications may require specific validation tests. Active guidance by nephrologists in all steps of self-measurement of BP at home will improve the control of BP and will increase the quality of patient care in CKD. Health care providers must urgently reimburse the cost of accurate home sphygmomanometers for patients with CKD in countries where CKD is considered a major public health problem.
引用
收藏
页码:300 / 306
页数:7
相关论文
共 41 条
[1]   Home and ambulatory blood pressure monitoring in chronic kidney disease [J].
Agarwal, Rajiv .
CURRENT OPINION IN NEPHROLOGY AND HYPERTENSION, 2009, 18 (06) :507-512
[2]   Out-of-office blood pressure monitoring in chronic kidney disease [J].
Agarwal, Rajiv ;
Peixoto, Aldo J. ;
Santos, Sergio F. F. ;
Zoccali, Carmine .
BLOOD PRESSURE MONITORING, 2009, 14 (01) :2-11
[3]   Proposal of a method for the evaluation of inaccuracy of home sphygmomanometers [J].
Akpolat, Tekin .
BLOOD PRESSURE MONITORING, 2009, 14 (05) :208-215
[4]   Home sphygmomanometers: validation versus accuracy [J].
Akpolat, Tekin ;
Dilek, Melda ;
Aydogdu, Turkan ;
Adibelli, Zelal ;
Erdem, Dilek Gurgenyatagi ;
Erdem, Emre .
BLOOD PRESSURE MONITORING, 2009, 14 (01) :26-31
[5]  
Altunkan S, 2007, BLOOD PRESS MONIT, V12, P219
[6]   Pathogenesis and treatment of kidney disease and hypertension - Home blood pressure monitoring in CKD [J].
Andersen, MJ ;
Khawandi, W ;
Agarwal, R .
AMERICAN JOURNAL OF KIDNEY DISEASES, 2005, 45 (06) :994-1001
[7]  
*ASS ADV MAD INSTR, 1993, SP101993 ANSIAAMI
[8]   The dabl educational trust device equivalence procedure [J].
Atkins, Neil ;
O'Brien, Eoin .
BLOOD PRESSURE MONITORING, 2007, 12 (04) :245-249
[9]   The message for World Kidney Day 2009: Hypertension and kidney disease: a marriage that should be prevented [J].
Bakris, George L. ;
Ritz, Eberhard .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 2009, 24 (03) :695-697
[10]  
*BRIT HYP SOC, REC BLADD DIM