Miniature fixed-point cells for self-calibration of type C high-temperature thermocouples

被引:3
作者
Sun, J. P. [1 ]
Pan, J. [2 ]
Lu, X. F. [1 ]
Manoi, A. [3 ]
Zhang, T. T. [4 ]
机构
[1] Natl Inst Metrol, Beijing, Peoples R China
[2] China Jiliang Univ, Hangzhou, Peoples R China
[3] Natl Inst Metrol Thailand, Pathum Thani, Thailand
[4] Xian Polytech Univ, Xian, Peoples R China
基金
中国国家自然科学基金;
关键词
thermocouple; thermal drift; self-calibration; miniature eutectic fixed-point cell; COBALT-CARBON;
D O I
10.1088/1361-6501/ab3d58
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A high-temperature tungsten-rhenium (W-Re) thermocouple is commonly used at temperatures up to 2000 degrees C for the high melting temperature of the thermoelement materials. When exposed to high temperatures, a thermocouple can show significant thermoelectric drift, which increases measurement uncertainty. The uncontrolled drift may come from the change in the crystallographic structure, oxidization of the thermoelement materials, interaction between the thermoelement materials and insulator materials and some other unknown reasons. Therefore, periodic recalibration of the thermocouple should be performed; however, sometimes it is not possible to remove the sensor out of the process, especially in some special fields, such as a nuclear power plant. Self-validation methods for thermocouples provide a solution to avoid this drawback. In this paper, miniature eutectic fixed-point cells are presented for self-validation of W-Re thermocouples. To prevent the breakage of the graphite crucible, these cylindrical miniature fixed-point cells contain several small independent crucibles and one thermocouple well, which increase the robustness of the miniature cell. The melting temperature was assigned by a radiation thermometer traced to the primary radiation standard at NIM. The effect of the temperature offset and temperature ramping rate on the melting temperature were checked. The performance of the miniature fixed-point cells and type C thermocouple, including characterization of the stability and repeatability, is presented.
引用
收藏
页数:7
相关论文
共 16 条
[1]  
[Anonymous], E988 ASTM INT
[2]   Investigation of the furnace effect in cobalt-carbon high-temperature fixed-point cells [J].
Dong, W. ;
Lowe, D. ;
Machin, G. ;
Bloembergen, P. ;
Wang, T. ;
Lu, X. .
MEASUREMENT, 2017, 106 :88-94
[3]   High temperature exposure of in-situ thermocouple fixed-point cells: stability with up to three months of continuous use [J].
Elliott, C. J. ;
Greenen, A. ;
Lowe, D. ;
Pearce, J. V. ;
Machin, G. .
METROLOGIA, 2015, 52 (02) :267-271
[4]   Self-testable industrial platinum resistance thermometers integrated with miniature mercury and indium fixed-point cells [J].
Kang, Kee Hoon ;
Kim, Yong-Gyoo ;
Gam, Kee Sool ;
Yang, Inseok .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2007, 18 (09) :3005-3008
[5]  
Morice R., 2008, Measure, V3, P44
[6]   Evaluation of cobalt-carbon and palladium-carbon eutectic point cells for thermocouple calibration [J].
Ogura, H. ;
Izuchi, M. ;
Arai, M. .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2008, 29 (01) :210-221
[7]   Stability of Tungsten-Rhenium Thermocouples in the Range from 0 °C to 1500 °C [J].
Ogura, H. ;
Izuchi, M. ;
Tamba, J. .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2011, 32 (11-12) :2420-2435
[8]   Multi-Mini-Eutectic Fixed-Point Cell for Type C Thermocouple Self-Calibration [J].
Ongrai, O. ;
Pearce, J. V. ;
Machin, G. ;
Norranim, U. .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2015, 36 (2-3) :423-432
[9]  
Ongrai O, 2011, C INT METR
[10]   Self-Validating Type C Thermocouples to 2300 °C Using High Temperature Fixed Points [J].
Pearce, J. V. ;
Elliott, C. J. ;
Machin, G. ;
Ongrai, O. .
TEMPERATURE: ITS MEASUREMENT AND CONTROL IN SCIENCE AND INDUSTRY, VOL 8, 2013, 1552 :595-600