Potential Impact of SPECT Resolution on Quantification of Left Ventricular Volumes and Ejection Fraction: A Phantom Study

被引:0
作者
Liu, Hui [1 ,2 ,3 ]
Aslan, Mehmet [1 ,4 ]
Sandoval, Veronica [5 ]
Liu, Yi-Hwa [1 ,5 ,6 ,7 ]
机构
[1] Yale Univ, Dept Internal Med Cardiol, New Haven, CT 06520 USA
[2] Tsinghua Univ, Dept Engn Phys, Beijing 100084, Peoples R China
[3] Tsinghua Univ, Key Lab Particle & Radiat Imaging, Minist Educ, Beijing 100084, Peoples R China
[4] Univ Texas Southwestern Med Ctr Dallas, Dept Internal Med, Div Hematol & Oncol, Dallas, TX 75390 USA
[5] Yale New Haven Med Ctr, Nucl Cardiol Heart & Vasc Ctr, New Haven, CT 06520 USA
[6] Natl Yang Ming Chiao Tung Univ, Dept Biomed Imaging & Radiol Sci, Taipei 11221, Taiwan
[7] Chung Yuan Christian Univ, Dept Biomed Engn, Taoyuan 320, Taiwan
关键词
LV Volume; LVEF; Partial Volume Effect; SPECT resolution; INCREMENTAL PROGNOSTIC VALUE; GATED SPECT; SPATIAL-RESOLUTION; PERFUSION IMAGES; RECONSTRUCTION; TOMOGRAPHY; ACCURACY; QUANTITATION; METHODOLOGY; VALIDATION;
D O I
10.1007/s40846-022-00747-y
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose Accuracy of quantitative assessment of the left ventricular (LV) function derived from ECG-gated SPECT may be affected by image resolution due to the partial volume effect (PVE). This study was to investigate the potential impact of SPECT resolution on quantification of the LV volumes and ejection fraction (EF). Methods Two hybrid SPECT/CT systems of different image resolutions were adopted in this study. One was a conventional dual-head SPECT scanner with low resolution (LR), and the other was a dedicated cardiac SPECT camera with high resolution (HR). Six custom-made cardiac phantoms each with a different LV volume were scanned using both scanners, and images were reconstructed with and without CT-based attenuation correction. A total of 11-bin gated SPECT images and 30 LVEFs were generated from the six phantom volumes to mimic a sequence of cardiac gated SPECT. Four commercial quantification software: Cedars-Sinai-QGS (QGS), Emory-Cardiac Toolbox (ECTB), Invia-Corridor4DM (4DM), and Yale Wackers- Liu CQ (WLCQ), were used to quantify the phantom LV volumes and EFs for comparisons. Results Both SPECT-quantified phantom volumes and EFs are correlated excellently with the true phantom values using Yale-WLCQ software, in which SPECT image resolution and CT-based attenuation show little effect on the correlations. However, the correlations of true phantom values vs. SPECT-quantified LV phantom volumes and EFs calculated from HR-SPECT with attenuation correction are either worse using ECTB, QGS, WLCQ or similar using 4DM as compared to those calculated from LR-SPECT. Quantification errors of the LV phantom EFs quantified from HR-SPECT are significantly greater than those from LR-SPECT, regardless which quantitative software was used (- 21 +/- 11% vs. - 9 +/- 8% with QGS, - 10 +/- 16% vs. - 1 +/- 11% with ECTB, - 16 +/- 10% vs. - 8 +/- 11% with 4DM, - 13 +/- 7% vs. - 1 +/- 5% with WLCQ; all p < 0.02). Conclusion Quantitative assessments of the LVEF from HR-SPECT mainly derived from the PVE results in a significant underestimation of true LVEFs as compared to that from LR-SPECT. Accordingly, current software tools for quantification of the LVEF may need to be remedied for HR-SPECT.
引用
收藏
页码:734 / 743
页数:10
相关论文
共 44 条
  • [1] Gated SPECT in assessment of regional and global left ventricular function: An update
    Abidov, Aiden
    Germano, Guido
    Hachamovitch, Rory
    Slomka, Piotr
    Berman, Daniel S.
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 2013, 20 (06) : 1118 - 1143
  • [2] Phantom evaluation of a cardiac SPECT/VCT system that uses a common set of solid-state detectors for both emission and transmission scans
    Bai, Chuanyong
    Conwell, Richard
    Kindem, Joel
    Babla, Hetal
    Gurley, Mike
    De Los Santos, Romer, II
    Old, Rex
    Weatherhead, Randy
    Arram, Samia
    Maddahi, Jamshid
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 2010, 17 (03) : 459 - 469
  • [3] Comparison of techniques for the measurement of left ventricular function following cardiac transplantation
    Bellenger, NG
    Marcus, NJ
    Rajappan, K
    Yacoub, M
    Banner, NR
    Pennell, DJ
    [J]. JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2002, 4 (02) : 255 - 263
  • [4] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [5] A fast cardiac gamma camera with dynamic SPECT capabilities: design, system validation and future potential
    Bocher, Moshe
    Blevis, Ira M.
    Tsukerman, Leonid
    Shrem, Yigal
    Kovalski, Gil
    Volokh, Lana
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 (10) : 1887 - 1902
  • [6] Chugh A, 2001, J AM COLL CARDIOL, V37, p394A
  • [7] COOKE CD, 1994, J NUCL MED, V35, P1185
  • [8] Dey J, 2012, IEEE T NUCL SCI, V59, P334, DOI [10.1109/TNS.2011.2182660, 10.1109/tns.2011.2182660]
  • [9] A review of partial volume correction techniques for emission tomography and their applications in neurology, cardiology and oncology
    Erlandsson, Kjell
    Buvat, Irene
    Pretorius, P. Hendrik
    Thomas, Benjamin A.
    Hutton, Brian F.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (21) : R119 - R159
  • [10] Performance evaluation of D-SPECT: a novel SPECT system for nuclear cardiology
    Erlandsson, Kjell
    Kacperski, Krzysztof
    van Gramberg, Dean
    Hutton, Brian F.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09) : 2635 - 2649